Chapters0/29 ▾
Chapter 3 · Lesson 1 of 5
Choosing and reading micrometers and calipers
1 min read
Start by matching the tool to the job. The rule of ten says the instrument should be about ten times finer than the tolerance, so a feature with a total tolerance of 0.020 in calls for a tool that resolves 0.002 in or better.
A standard inch micrometer resolves 0.001 in; a vernier or digital micrometer resolves 0.0001 in. A dial caliper whose dial makes one turn per 0.100 in has 100 graduations on the face, so each one is 0.001 in, not 0.01 in.
Example: the sleeve shows the 0.300 mark plus two more 0.025 in lines, and the thimble reads 17. The reading is 0.300 + 0.050 + 0.017 = 0.367 in. On a metric micrometer, 50 thimble divisions of 0.01 mm make one 0.50 mm turn, so a sleeve showing 7.50 mm with the thimble at 23 reads 7.73 mm. A vernier caliper reads the fraction from the one vernier line that lines up exactly with a main-scale line, so nothing is guessed by eye.
Always close a micrometer with the ratchet or friction thimble, which slips at a set force so every operator gets the same reading. For unit changes, divide millimetres by 25.4: 19.05 mm is 0.750 in.
Check yourself
1. On an inch outside micrometer the sleeve shows the 0.200 numbered mark plus three more 0.025 graduations fully exposed, and the thimble reads 8. What is the measurement?
2. On a vernier caliper, how is the fractional part of a main-scale division determined?
Facts checked 2026-11-15 · Sources: MSSC: Certified Production Technician (CPT 4.0), NIST: Metrological traceability