Calibration and conditions that affect accuracy
1 min read
Calibration compares an instrument with a standard of known, higher accuracy that is traceable through an unbroken chain to the national standards kept by NIST.
Comparing a micrometer with another shop micrometer, a good production part or the nominal size on the print proves nothing, because none of them carries a traceable value. Each calibration check and its result are documented.
When a gauge is found out of calibration, tag it and pull it from service, and also identify and re-inspect every part it measured since its last good calibration. Re-zeroing the tool or shortening the interval does nothing for parts already measured.
Example: a caliper last passed on the 1st and fails on the 20th, so everything it checked from the 1st onward is suspect, not just the parts from the 20th, and any of that product still in the plant is held until it is re-checked with a known-good tool.
- Heat makes metal grow: a part warmed in a bare hand measures slightly larger than at the 20 degrees C (68 degrees F) reference temperature, so let parts and tools normalize and handle them with a cloth or by the insulated pads.
- A part found outside the print limits during in-process checks is tagged, segregated and reported to the supervisor before anyone adjusts the machine.
Check yourself
1. A micrometer used on production all week is found to be out of calibration. Besides tagging it and removing it from service, what must happen?
2. Calibration of a measuring instrument means comparing that instrument against what?
Facts checked 2026-11-15 · Sources: MSSC: Certified Production Technician (CPT 4.0), NIST: Metrological traceability