Chapter 3 · Lesson 4 of 5
Release decisions: requalification, implants and immediate use
1 min read
Some situations raise the bar before anything may go to a patient. After installation, relocation, a malfunction, a major repair, a process failure or a change in utilities, the sterilizer is requalified.
That means consecutive challenge cycles monitored with biological indicators, plus an air-removal test on a dynamic-air-removal unit. It stays out of service until those cycles pass. A vendor's service report does not release it.
Every load containing an implant is monitored with a BI and quarantined until the result is read. In a documented emergency only, an implant may be released early on a passing integrating indicator and the physical record. Even then the BI keeps incubating, the result is documented, and the early release goes on an exception form with follow-up.
Immediate-use steam sterilization (IUSS) is for an item needed right away when no sterile replacement exists. Its record must trace the load contents and cycle data to the individual patient and the operator.
Example: a department owns 1 set of an instrument used in 3 cases a day and flashes it between cases. That is 3 IUSS cycles a day covering an inventory shortage, which is misuse; the fix is buying more sets.
Check yourself
1. A steam sterilizer has just undergone a major repair to its vacuum system. What must happen before it processes items for patient use again?
2. An implant was processed in a load monitored with a process challenge device holding both a biological indicator and an integrating indicator. The biological indicator is still incubating and the implant is needed for a documented emergency. What does the standard allow?