Chapter 5 · Lesson 1 of 5
Sterilizer records, lot control and traceability
1 min read
Two kinds of documentation work together. Each package carries a lot (load) control identifier that links it to the sterilizer, the cycle and the date. Separately, the sterilizer record for each cycle lets anyone reconstruct that load after the fact.
Ideally the identifier is also recorded at the point of use, so a device can be traced to the patient it was used on. The label proves only where and when an item was processed. It does not prove the item is sterile.
| Record | What it holds |
|---|---|
| Lot control label on each package | Sterilizer, cycle number and date |
| Sterilizer record for each cycle | Load number, contents, exposure time and temperature, operator, and the biological, Bowie-Dick and chemical indicator results |
Example: a tray is about to be issued, but the cycle record has no operator initials and no process challenge device result. The tray is held and the supervisor is told, because the record cannot show the release criteria were met.
Filling in entries from memory is never acceptable, and neither is having the operator initial it later in the shift. Retention is another common question: no single national period exists. Each facility sets it in policy based on state, federal and accrediting requirements, because these are legal records.
Check yourself
1. Which set of information must be documented in the sterilizer record for each sterilization cycle?
2. A technician about to issue a tray finds that the load record for that cycle has no operator initials and no entry for the process challenge device result. What is correct?
Facts checked 2026-11-15 · Sources: HSPA: Certified Registered Central Service Technician (CRCST), CDC: Guideline for Disinfection and Sterilization in Healthcare Facilities (2008)