Chapter 5 · Lesson 5 of 5
Instructions for use, loaners, competency and quality benchmarks
1 min read
The manufacturer's validated written instructions for use (IFU) govern how a device is cleaned, assembled, packaged and sterilized, because only the manufacturer has tested that the method works for that design. If a loaner set arrives without instructions, the technician obtains them from the manufacturer or vendor and holds the set until they arrive.
Using the longest cycle or copying a similar set is guessing. If a device's IFU and a rigid container's validated cycles do not match, both manufacturers are contacted and the set waits for a method validated by both. A longer cycle is not a validated cycle.
Loaner sets are tracked like any other inventory: receipt and count on arrival, decontamination and inspection, the load identification they were sterilized under, and their condition when returned. A set that arrives already wrapped is still decontaminated and sterilized by the facility.
Competency is verified and documented, and continuing education is required, because devices and IFU keep changing. Defects found at the point of use are reported in writing and trended with the user department, not handled as a quiet word to one technician.
Standard quality benchmarks are tray audit results, the immediate-use steam sterilization rate, set turnaround time, case cart accuracy and inventory fill rates. Example: 12 immediate-use cycles out of 400 loads in a month is 12 ÷ 400 = 3%, a rate the department compares with its benchmark over time.
Check yourself
1. A loaner set arrives for tomorrow's case with no manufacturer's instructions for use in the container. What does the technician do?
2. What is the sterile processing technician's documented role in handling a loaner instrument set?
Facts checked 2026-11-15 · Sources: HSPA: Certified Registered Central Service Technician (CRCST), CDC: Guideline for Disinfection and Sterilization in Healthcare Facilities (2008)