Quibank app: now on the App Store — free, works offline. Get the app ↗

Chapters0/41 ▾
  1. 1.Department Operations & Safety0/5
    1. The work environment: pressure, humidity, cleaning and workflow
    2. Attire, jewelry and traffic control
    3. Workplace safety: spills, chemicals and safe lifting
    4. Regulators, standards and the order of authority
    5. Training records, documentation and quality improvement tools
  2. 2.Decontamination & Cleaning0/5
    1. Point-of-use treatment and safe transport
    2. The decontamination room and personal protective equipment
    3. Manual cleaning: water, brushes and detergent
    4. Ultrasonic cleaners and washer-disinfectors
    5. Flexible endoscopes and automated reprocessors
  3. 3.Sterilization Methods0/5
    1. How steam kills and how air is removed
    2. The daily air-removal (Bowie-Dick) test
    3. Physical, chemical and biological monitoring
    4. Release decisions: requalification, implants and immediate use
    5. Loading, wet packs, cooling and low-temperature methods
  4. 4.Packaging & Sterile Storage0/4
    1. Inspecting and assembling instruments
    2. Wraps and wrapping techniques
    3. Peel pouches, rigid containers and labels
    4. Cooling, wet packs and sterile storage
  5. 5.Quality Assurance & Monitoring0/5
    1. Sterilizer records, lot control and traceability
    2. Releasing a load and handling implants
    3. Positive biological indicators, recalls and failure reports
    4. Shelf life, stock rotation and package inspection
    5. Instructions for use, loaners, competency and quality benchmarks
  6. 6.Infection Prevention0/5
    1. The chain of infection, Standard Precautions and hand hygiene
    2. Decontamination design, airflow and safe transport
    3. Sharps, PPE and exposure incidents
    4. The Spaulding classification and high-level disinfection
    5. Microbial resistance, soil, biofilm and prions
  7. 7.Storage, Inventory & Distribution0/4
    1. Replenishment systems and ordering
    2. Receiving in the break-out area
    3. Sterile storage, rotation and transport
    4. Tracking, recalls and inventory problems
  8. 8.Patient Care Equipment0/4
    1. Why patient care equipment is noncritical
    2. Cleaning, contact time and a clean work area
    3. Inspection and removal from service
    4. Preparing, storing, distributing and tracking
  9. 9.Professional Skills & Terminology0/4
    1. Procedure suffixes: what the surgeon will do
    2. Body-part roots and putting words together
    3. Protecting patient information under HIPAA
    4. Professional communication and teamwork
Saved in this browser. Move it to another device →
← Patient Care Equipment

Chapter 8 · Lesson 3 of 4

Inspection and removal from service

1 min read

Every unit is inspected during processing, and some findings mean it cannot go out. OSHA's electrical safe-work rule says cord-and-plug equipment must be checked for external defects and for signs of possible internal damage, such as a pinched or crushed outer jacket.

Equipment with a defect that could injure someone is removed from service and not used until it is repaired and tested.

A passed startup self-test checks the device's functions, not the cord's insulation, so it does not clear a crushed cord.

The same logic applies to preventive maintenance: a pump that powers on normally but carries an expired PM label is out of compliance, because powering on does not show that the safety and performance checks are current. It goes to Biomedical Engineering, not to the unit.

FindingCorrect action
Frayed or crushed cordRemove from service and tag for Biomedical Engineering
Missing or expired PM labelHold the unit for Biomedical Engineering
Three-prong plug, two-slot outletUse a grounded outlet and report the old one; never an adapter that breaks the ground
Part that cannot be decontaminated before shippingAttach a biohazard label stating which part is contaminated and tell the servicer

Check yourself

1. While cleaning a hypothermia unit, a technician finds that the outer jacket of its power cord is frayed and the inner wires show. What should happen to the unit?

2. A portable suction unit's power cord has a flattened, crushed section where a bed wheel rolled over it. The unit powers on and passes its startup self-test. How should it be handled?

Facts checked 2026-11-15 · Sources: HSPA: CRCST certification and exam content outline, CDC/HICPAC: Guideline for Disinfection and Sterilization in Healthcare Facilities, OSHA 29 CFR 1910.1030: Bloodborne Pathogens

Sterile Processing (CRCST) cheat sheet: every key fact on one page →