Spaulding decides the minimum processing level by what a device will touch on that use, not by what the device is called. The device is always cleaned first.
| Category | Contact | Minimum processing |
|---|---|---|
| Critical | Sterile tissue, joint spaces or the bloodstream | Sterilization |
| Semicritical | Mucous membranes or non-intact skin | High-level disinfection |
| Noncritical | Intact skin only | Cleaning, then low-level disinfection at the labeled contact time |
Example: one ultrasound transducer is used on intact skin in the morning and enters a sterile joint space in the afternoon. For the skin scan it is noncritical; for the joint it is critical and must be sterilized, and a probe cover does not lower that.
A rigid arthroscope in the knee is critical too, even though it is an endoscope. High-level disinfection destroys all microorganisms except large numbers of bacterial spores. A disinfectant basin is discarded when either its test strip fails or its labeled reuse period ends, whichever comes first.
Check yourself
1. The same ultrasound transducer is used on intact skin for a bedside scan and, in a later procedure, is introduced into a sterile joint space. Under the Spaulding classification, what does this mean?
2. A transesophageal echocardiography probe contacts the mucous membrane of the esophagus and does not enter sterile tissue. What does the Spaulding classification require for it?
Facts checked 2026-11-15 · Sources: CDC: Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), OSHA 29 CFR 1910.1030: Bloodborne Pathogens, HSPA: Certified Registered Central Service Technician (CRCST)