Chapter 3 · Lesson 3 of 5
Physical, chemical and biological monitoring
1 min read
Every load is monitored in layers, and the exam asks what each layer can prove. Physical monitors, such as the printout or electronic record, report what the chamber did: time, temperature and pressure.
Chemical indicators show that certain conditions were reached at the spot where the indicator sat. Only a biological indicator (BI), which carries a known population of highly resistant spores, directly measures whether the cycle was lethal.
- An integrating (Type 5) indicator responds to all critical variables but contains no living spores, so it cannot replace routine biological monitoring.
- An emulating (Type 6) indicator is labeled for one named cycle; in any other cycle its reading means nothing. Type numbers describe what an indicator responds to, not a quality ranking.
- A process challenge device (PCD) holding the BI goes in a fully loaded chamber where steam arrives last: low, toward the front, over the drain.
| Processed BI | Unprocessed control BI | Meaning |
|---|---|---|
| No growth | Growth | Pass |
| Growth | Growth | Fail: sterilization failure |
| No growth | No growth | Invalid: spores or incubator cannot be trusted |
Example: the day's processed BI and an unprocessed control from the same lot are incubated together. After incubation, neither vial shows growth. It is tempting to call that a pass, but the control proves the spores were alive and the incubator worked. With no growth in the control, the run is invalid and must be repeated.
Check yourself
1. A technician incubates the day's processed biological indicator together with an unprocessed control from the same lot. The processed indicator shows no growth, and the control also shows no growth. How is this interpreted?
2. Physical, chemical and biological monitors are all applied to a steam load. Which statement describes what they can establish?