1 min read
The exam separates two situations. If a biological indicator is positive and the cause is not found right away, the department must assume the sterilizer has been failing since it was last proven effective.
Every load processed since the last negative biological indicator is recalled and reprocessed, and the sterilizer stays out of service until the cause is corrected. The recall does not wait for a laboratory to identify the growth.
If the cause is found immediately and is limited to one load, such as the operator choosing the wrong cycle, the response is narrower: correct the cause, reprocess that load only, and document both the failure and the corrective action. Releasing the load anyway is never correct, because the indicator shows that load was not proven sterile.
Example: the last negative biological indicator was in Monday's first load. Afterward the sterilizer ran 6 more loads on Monday, 7 on Tuesday and 3 on Wednesday, and the third Wednesday load has a positive indicator with no cause found.
Loads to recall: 6 + 7 + 3 = 16. The written failure report must include the date and time, the sterilizer, the load control numbers and the monitoring results, because without load numbers the affected packages cannot be found.
Check yourself
1. Today's biological indicator in the process challenge device is positive and no cause can be found. What happens to the loads processed since the last negative biological indicator?
2. A biological indicator is positive, and the investigation immediately shows the operator selected the wrong cycle for that load. What is the correct response?