Ultrasonic cleaners and washer-disinfectors
1 min read
An ultrasonic cleaner works by cavitation. High-frequency sound waves create tiny bubbles that collapse violently against the instrument and pull fine soil out of box locks and serrations that a brush cannot reach.
It cleans only; it has no antimicrobial claim. Gross soil comes off first, items are fully submerged with hinges open, the lid stays closed to contain aerosols, and items are rinsed afterward.
Fresh solution must be degassed, because dissolved air cushions the bubbles and weakens cavitation. Stainless steel is not mixed with chrome-plated, brass, aluminum or copper items, because dissimilar metals in an electrolyte set up galvanic action that pits the metal.
A washer-disinfector cleans by spray impingement, so it cleans only where water strikes the surface. Load it with hinges open, concave items on edge, heavy items on the bottom, lumens attached to manifolds, and no stacking or overloading. Both machines need low-foaming detergent, because foam absorbs the spray and cavitation energy before it reaches the instrument.
Example: a washer is used Monday through Friday and gets a major repair on Wednesday afternoon.
It is tested with a cleaning verification device each of those 5 days, plus once more after the repair before loads go back in. If any test fails, the machine comes out of service, affected loads are held or reprocessed, and the cause is investigated and documented.
Check yourself
1. Freshly prepared solution in an ultrasonic cleaner must be degassed before instruments are processed. Why?
2. How does an ultrasonic cleaner remove soil, and what does the process actually accomplish?