Chapter 8 · Lesson 2 of 4
Cleaning, contact time and a clean work area
1 min read
Blood and other organic matter react with germicides and shield microorganisms, so visible soil comes off first and the disinfectant goes on second. A one-pass wipe on a bloody keypad is the wrong answer even when the product claims to clean and disinfect. Next comes contact time.
CDC states that label instructions on EPA-registered products must be followed by law, and a user who changes the exposure conditions takes on liability for off-label use.
Example: the label says 3 minutes wet, which is 180 seconds. Towel-drying at 30 seconds leaves the surface 150 seconds short, and a second short wipe after drying does not add up to one continuous wet period.
- Choose cleaning agents from the device's operator manual or IFU, which lists what its plastics and display can tolerate; a new model may differ from the old one.
- OSHA requires the employer to set a written cleaning schedule and method for each work area, based on location, surface type, soil present and tasks done.
- Replace a protective plastic cover as soon as feasible once it is overtly contaminated, or at the end of the shift if it may have been contaminated.
- When cleaning could spatter, wear gloves plus a mask with goggles or a chin-length face shield.
Example: two hours into a shift, blood splashes the plastic cover on a return-area touchscreen. The cover comes off and is replaced as soon as feasible; waiting for the end of the shift is only for covers that may have been contaminated, not ones visibly splashed.
Check yourself
1. A disinfectant label says the surface must stay visibly wet for 3 minutes. A technician wipes a sequential compression device (SCD) pump and, after 30 seconds, towels it dry so it can be bagged. What is the correct practice?
2. An infusion pump comes back from a nursing unit with dried blood streaked across its housing and keypad. The department uses an EPA-registered hospital disinfectant wipe. What should the technician do?