Inspecting and assembling instruments
1 min read
Every instrument is inspected for cleanliness and function before it goes into a set. If dried debris turns up in the serrations of a single clamp, the whole tray is considered contaminated and returns to decontamination, because an item that is not clean cannot be sterilized. Wiping or brushing one clamp at the prep table is never the answer.
| Instrument | Test | Failure means |
|---|---|---|
| Scissors | Cut test material suited to their size | Must cut cleanly to the tip; otherwise repair |
| Clamp (hemostat) | Close and hold to the light | Light between jaws = misaligned; remove from service |
| Needle holder | Close on a needle at the second ratchet and try to turn it | Needle turns easily = worn jaws; repair |
| Insulated laparoscopic | Look for nicks and pull back on the insulation | Sliding insulation or a gap = re-insulate |
Technicians do not sharpen, bend or patch instruments; those are repair jobs. Stiff hinges get a water-soluble lubricant per the IFU, never silicone or mineral oil, which can interfere with sterilization.
Flaking color tape comes off completely. Ratcheted instruments are unlatched, heavy items go on the bottom, and flat or concave items such as malleable retractors stand on edge so water does not pool.
Example: a set's tray and wrap weigh 7.5 lb and its instruments 14 lb, and the surgeons add two 2.25-lb retractors.
The new total is 7.5 + 14 + 2 x 2.25 = 26 lb, over the 25-lb limit, so the changed set is re-weighed and reconfigured, for example in a tray that spreads the weight and metal mass evenly, until it is within the limit.
Check yourself
1. A technician closes a hemostatic clamp and holds it up to the light. Light is visible between the jaws. What does this mean, and what comes next?
2. During inspection in the assembly area, a technician finds dried debris in the serrations of one clamp. The other instruments in that tray look clean. What is the correct action?