Flexible endoscopes and automated reprocessors
1 min read
Flexible endoscopes are long, narrow and full of internal channels, which makes them one of the hardest devices to clean. The exam focuses on two points: the order of the steps, and the fact that automation does not replace hand cleaning.
- Leak test first, before the scope is immersed. Fluid entering a breached scope destroys internal parts and creates a contaminated space that cannot be cleaned.
- Manual cleaning with channel brushing comes next, following the scope's own IFU for brush sizes and channel adapters.
- Only then does the automated endoscope reprocessor (AER) run its cycle. Soil and biofilm left in a channel would inactivate the high-level disinfectant and shield organisms from it.
A borescope can be used to look inside channels that cannot be seen directly. Example: a scope arrives and the AER is free right away.
The technician still leak tests it, brushes every channel by hand, and only then loads it, even though the AER has its own cleaning phase. Skipping the leak test to save time, or doing it after disinfection, are both wrong answers.
Check yourself
1. A flexible endoscope is being reprocessed in a department equipped with an automated endoscope reprocessor (AER). Which statement describes correct practice?
2. How often should mechanical cleaning equipment such as washer-disinfectors and ultrasonic cleaners be challenged with cleaning verification test devices, and what must happen when a test fails?
Facts checked 2026-11-15 · Sources: HSPA: Certified Registered Central Service Technician (CRCST) exam content outline, CDC/HICPAC: Guideline for Disinfection and Sterilization in Healthcare Facilities, OSHA 29 CFR 1910.1030: Bloodborne Pathogens
Sterile Processing (CRCST) cheat sheet: every key fact on one page →