Study guide · Chapter 2 of 5

The Machine & Water Treatment Study Guide — Dialysis Technician

42 free practice questions

This section covers the water treatment system, the dialysate, the pressure monitors and the safety alarms on the machine. Exact limits and setpoints come from the manufacturer and your facility, so this lesson focuses on what each component protects against and what to do first when a reading or alarm says something is wrong.

Key takeaways

  • Carbon tanks run in series ahead of the reverse osmosis unit, and carbon is the stage that removes chlorine and chloramine.
  • A high chlorine result between the tanks means test after the second tank; high there too means no treatments.
  • Dialysate is checked with a separate calibrated meter, and a machine that differs by more than the manufacturer's tolerance is taken out of service.
  • Rising venous pressure at a steady pump speed means a narrowing path, and a sudden drop means a possible disconnection.
  • Safety stops from the air detector or blood leak detector are respected until the cause is found, never overridden.

Lessons in this chapter

5 short lessons · 5 min read

  1. 1The water treatment train1 min
  2. 2Testing for chlorine and chloramine1 min
  3. 3Dialysate: concentrates, conductivity and temperature1 min
  4. 4Reading the circuit pressures1 min
  5. 5Priming, safety monitors and reused dialyzers1 min
View all questions & answers →

Facts checked 2026-11-15 · Sources: NNCC: Certified Clinical Hemodialysis Technician (CCHT), 42 CFR Part 494: Conditions for Coverage for ESRD Facilities

Dialysis Technician cheat sheet: every key fact on one page →Next chapter: Vascular Access →

Practice tests

Practice questions based on the NNCC CCHT content outline, CDC recommendations for hemodialysis settings, and the OSHA Bloodborne Pathogens Standard. Not affiliated with or endorsed by NNCC, and not medical advice. Machine settings, dialysate composition, flow rates, alarm thresholds, water-quality limits and laboratory targets are deliberately never the answer to a question here — those are set by your machine's manufacturer, your facility's protocol and the physician's prescription. Always follow your facility's policies and your patient's prescribed treatment. About the CCHT exam (NNCC) →