Quibank app: now on the App Store — free, works offline. Get the app ↗

42 The Machine & Water Treatment Practice Questions & Answers

Every The Machine & Water Treatment practice question from the Dialysis Technician Practice Test, with the correct answer and a short explanation.

Start practice test →

Practice tests

  1. 1. A technician is preparing the water treatment system before the first patients of the day. Where should the sample for chlorine testing be drawn, and why?

    • A.From the port between the two carbon tanks, because it reveals breakthrough of the first tank while the second still guards.✓ Answer
    • B.From the port after the softener, because chlorine is exchanged onto the resin there along with calcium and magnesium.
    • C.From the port after the reverse osmosis unit, because that membrane is the last barrier standing before water reaches the machines.
    • D.From the port on the city feed line, because the utility sets the disinfectant level that the whole system must handle.

    Carbon beds are installed in series precisely so that exhaustion of the first (worker) tank can be detected while the second (polisher) tank still protects patients. Sampling only at the very end of the train would hide that early warning, and neither the softener nor the reverse osmosis membrane removes chlorine or chloramine.

    Ref QB-2VSCH7Source: ANSI/AAMI standard for hemodialysis water treatment equipment, carbon adsorption bed requirements; CMS ESRD Conditions for Coverage, 42 CFR 494.40Report a problem with this question

  2. 2. The chlorine sample drawn between the two carbon tanks exceeds the facility's limit before the first shift. What should the technician do FIRST?

    • A.Rebed both carbon tanks at once and restart the system, since a reading above the limit condemns the whole carbon bank.
    • B.Start the treatments and repeat the test on the same port later in the shift, since the second tank has not been checked yet.
    • C.Switch the loop to the bypass line and run the treatments on softened water that has not yet passed the carbon tanks.
    • D.Draw and test a sample from the port after the second carbon tank to learn whether treated water is still within limit.✓ Answer

    A result above the limit at the intermediate port means the worker tank is exhausted, not that the treated water is unsafe. The required next step is to test immediately after the polisher tank, because that reading decides whether treatment may proceed while the first tank is serviced.

    Ref QB-KCBV99Source: ANSI/AAMI standard for hemodialysis water treatment equipment, chlorine and chloramine monitoring; CMS ESRD Conditions for Coverage, 42 CFR 494.40Report a problem with this question

  3. 3. The sample taken after the second (polisher) carbon tank also exceeds the limit, and patients are waiting to be put on. What is the correct action?

    • A.Reduce the blood flow rate on every machine and begin, because slower flow lowers the chloramine dose each patient receives.
    • B.Install a fresh carbon tank and start the shift meanwhile, because the polisher still removes most of the chloramine present.
    • C.Hold all treatments, notify the nurse in charge and the biomedical staff, and keep the water system out of use.✓ Answer
    • D.Add an extra ultrafilter to each machine and proceed, because those filters trap the oxidant before it reaches the dialyzer.

    When the sample after the polisher tank is also over the limit, the entire carbon bank has failed and the product water may not be used for dialysis; treatments already running must be stopped and none may be started. Ultrafilters remove bacteria and endotoxin but not dissolved oxidants, and lowering blood flow does not make contaminated water safe.

    Ref QB-3V5QP8Source: ANSI/AAMI standard for hemodialysis water quality, chloramine action limits; CMS ESRD Conditions for Coverage, 42 CFR 494.40Report a problem with this question

  4. 4. Over two weeks several patients on the same shift show falling hemoglobin and rising erythropoietin requirements, and one reports chest and back pain during the run. Which water problem best explains this pattern?

    • A.Endotoxin passing a torn ultrafilter, which triggers chills and fever during the run and drives the same anemia pattern.
    • B.Aluminum passing a failed reverse osmosis membrane, which binds oxygen on the red cell and causes this same oxidant injury.
    • C.Calcium and magnesium passing a spent softener, which stiffen red cell membranes and shorten their survival in the loop.
    • D.Chloramine passing an exhausted carbon bed, which oxidizes red cells and causes hemolysis that no infection explains.✓ Answer

    Chloramine is an oxidant that attacks the red cell directly, so carbon breakthrough shows up as unexplained hemolytic anemia, resistance to erythropoietin, and chest or back pain in patients sharing the same water. Endotoxin produces fever and chills during the treatment rather than progressive anemia, and aluminum injury presents as bone and neurologic disease.

    Ref QB-MCP7FASource: ANSI/AAMI standard for hemodialysis water quality, contaminant effects; Core Curriculum for the Dialysis Technician, water treatmentReport a problem with this question

  5. 5. A new technician asks why the carbon tanks are plumbed ahead of the reverse osmosis unit rather than after it. What is the best answer?

    • A.Carbon needs the higher line pressure found upstream, because adsorption of dissolved gases only works under that pressure.
    • B.Carbon removes the sand and rust that would score the RO membrane, a job the membrane cannot do to protect itself.
    • C.Carbon adsorbs the chlorine and chloramine that would otherwise damage the RO membrane, which cannot remove them itself.✓ Answer
    • D.Carbon lowers water hardness ahead of the membrane, since scale from calcium is what destroys thin-film membranes fastest.

    Reverse osmosis rejects dissolved ions, bacteria and endotoxin but does not remove chlorine or chloramine, and those oxidants degrade the membrane itself. Placing carbon adsorption upstream both protects the membrane and is the only step in the train that takes the oxidant out of the water.

    Ref QB-4VG2V9Source: ANSI/AAMI standard for hemodialysis water treatment equipment, pretreatment sequenceReport a problem with this question

  6. 6. The hardness test on water leaving the softener is suddenly high and the brine tank is empty. What is the consequence if the system keeps running this way?

    • A.Calcium and magnesium are no longer exchanged for sodium, so they scale the RO membrane and burden every later stage.✓ Answer
    • B.Sediment is no longer trapped in the resin bed, so particles reach the carbon and shorten the contact time it provides.
    • C.Bacteria are no longer held on the resin bed, so endotoxin in the product water climbs above the action level.
    • D.Chlorine is no longer exchanged for sodium, so the carbon tanks take the full oxidant load and break through much earlier.

    The softener works by ion exchange, trading calcium and magnesium for sodium, and it needs salt in the brine tank to regenerate that resin. Without regeneration, hardness passes downstream and scales the reverse osmosis membrane, which is the component the softener exists to protect. Sediment is removed by the multimedia or cartridge filters, bacteria and endotoxin are controlled by the RO and disinfection rather than by the resin, and chlorine and chloramine are removed by the carbon beds, not exchanged by the softener, so an empty brine tank does not change those stages.

    Ref QB-NC2JMBSource: ANSI/AAMI standard for hemodialysis water treatment equipment, softener and pretreatment requirementsReport a problem with this question

  7. 7. A facility uses a deionizer to polish the water after reverse osmosis. Why must an ultrafilter be installed downstream of that deionizer?

    • A.The resin bed raises the water temperature as it works, and the filter after it cools the water before the storage tank.
    • B.The resin bed removes ions but grows bacteria and releases endotoxin, so a filter after it protects the product water.✓ Answer
    • C.The resin bed releases carbon fines as it works, and the filter after it keeps those particles out of the storage tank.
    • D.The resin bed removes bacteria but not ions, so the filter after it is what finally lowers the conductivity of the water.

    Deionization removes charged particles only; the warm, wet resin bed is an excellent site for bacterial growth and the endotoxin those organisms release passes straight downstream. An ultrafilter after the deionizer, together with continuous resistivity monitoring, is therefore required to keep the product water microbiologically acceptable.

    Ref QB-5VUC2BSource: ANSI/AAMI standard for hemodialysis water treatment equipment, deionization requirementsReport a problem with this question

  8. 8. The machine display shows conductivity and temperature inside the set range and no alarms are active. What must the technician still do before the patient is connected?

    • A.Ask a second technician to read the same machine display and initial the treatment record beside the machine's value.
    • B.Watch the machine display through a full rinse cycle and write down the lowest reading seen during that observation.
    • C.Measure the dialysate with a separate calibrated meter and record the reading beside the value the machine displays.✓ Answer
    • D.Confirm in the manufacturer's service log that the machine's internal cell was calibrated during the last service visit.

    The machine's own cell can drift or fail, so its display cannot verify itself; an independent, calibrated meter is required to confirm the dialysate before the patient is connected, and both values are documented. A second person reading the same suspect display adds nothing, because it is the instrument and not the observer that is being checked.

    Ref QB-PCDVSCSource: ANSI/AAMI standard for hemodialysis systems, independent verification of dialysate; CMS ESRD Conditions for Coverage, 42 CFR 494.40Report a problem with this question

  9. 9. The independent meter reading for conductivity differs from the machine's display by more than the manufacturer's allowed tolerance, although both values fall inside the facility's acceptable range. What should the technician do?

    • A.Take the machine out of service, tag it, notify biomedical staff, and move the patient to a machine that verifies correctly.✓ Answer
    • B.Reduce the dialysate flow rate until the two readings match, then begin and re-check the pair midway through the run.
    • C.Record both readings on the treatment sheet and start, since each value sits inside the acceptable range for dialysate.
    • D.Re-zero the independent meter and use the machine reading, since the machine's cell is calibrated more often than the meter.

    The point of independent verification is that the machine's display must be validated by an outside measurement; when the two disagree, the machine cannot be trusted to control proportioning even if the numbers look acceptable. The safe response is to remove and tag the machine, report it to biomedical staff, and dialyze the patient on equipment that verifies.

    Ref QB-6V7Q7CSource: ANSI/AAMI standard for hemodialysis systems, independent verification of dialysate; CMS ESRD Conditions for Coverage, 42 CFR 494.40Report a problem with this question

  10. 10. A conductivity alarm shows a value well below the set range. Compared with correctly proportioned dialysate, what would this fluid do to the patient's blood?

    • A.It is hypotonic, so sodium moves rapidly into the blood and produces cramping with a rise in the patient's blood pressure.
    • B.It is hypotonic, so water moves into the red cells and they swell and burst, which can be fatal within minutes.✓ Answer
    • C.It is hypertonic, so water leaves the red cells and they shrink, producing thirst, headache and a rising blood pressure.
    • D.It is hypertonic, so potassium is driven into the cells and the patient develops weakness and a slowed heart rhythm.

    Conductivity is a surrogate for the total ion content of the dialysate, so a low value means the fluid is too dilute, that is hypotonic to blood. Water then moves osmotically into the red cells until they lyse, which is why low conductivity is the more rapidly lethal direction and why the machine must stay in bypass until the cause is found.

    Ref QB-6RRCERSource: Kallenbach, Review of Hemodialysis for Nurses and Dialysis Personnel, dialysate composition and delivery systemsReport a problem with this question

  11. 11. When a conductivity or temperature alarm sounds, the machine goes into bypass. What does bypass physically do?

    • A.It routes the dialysate once more through the heater, so the machine can correct the reading without interrupting the run.
    • B.It stops the blood pump and clamps the venous line, so no more blood leaves the patient until the fault is corrected.
    • C.It closes the ultrafiltration controller only, so fluid removal stops while dialysate keeps flowing through the dialyzer.
    • D.It diverts dialysate to the drain instead of through the dialyzer, so the unsafe fluid never meets the patient's blood.✓ Answer

    Bypass is a protective dialysate-side action: the fluid that failed its conductivity or temperature check is routed to drain instead of through the dialyzer, so it never contacts blood across the membrane. Dialysis effectively stops during bypass, which is why the cause must be corrected rather than the bypass defeated.

    Ref QB-P8XHZRSource: Kallenbach, Review of Hemodialysis for Nurses and Dialysis Personnel, delivery system safety monitorsReport a problem with this question

  12. 12. The dialysate temperature reading is climbing above the upper limit and the machine has alarmed. What is the danger if dialysate that is too hot reaches the dialyzer?

    • A.Heat raises the conductivity of the dialysate, so the patient receives a sodium load and becomes thirsty and hypertensive.
    • B.Heat thickens the blood inside the fibers and the dialyzer clots, so the treatment must be restarted with a new circuit.
    • C.Heat drives dissolved gas out of the dialysate and that air crosses the membrane into the blood on the patient's side.
    • D.Heat damages the red cells and they hemolyze, releasing potassium and leaving the venous blood a translucent red.✓ Answer

    Red cells are heat labile, and dialysate warmer than the safe window destroys them across the membrane, causing acute hemolysis with potassium release, translucent cherry-red or port-wine venous blood, back and chest pain and possible cardiac arrest. That is why the high temperature limit forces the machine into bypass before the fluid can reach the dialyzer. Temperature does not add sodium to the bath (the conductivity monitor is temperature compensated), heat does not thicken blood or clot the dialyzer, and air in the dialysate is caught by the air separator rather than crossing an intact membrane, so the distinctive danger of overheating is hemolysis.

    Ref QB-7R4QKSSource: Kallenbach, Review of Hemodialysis for Nurses and Dialysis Personnel, dialysate temperature controlReport a problem with this question

  13. 13. A new technician asks why the acid concentrate and the bicarbonate concentrate are never combined in one container before they reach the machine. What is the reason?

    • A.Undiluted, they neutralize each other and the mixture loses all of the potassium the prescription calls for that day.
    • B.Undiluted, they raise the sodium of the mixture so far that no proportioning ratio can bring it back into the range.
    • C.Undiluted, they react and precipitate calcium and magnesium carbonate while releasing carbon dioxide from the solution.✓ Answer
    • D.Undiluted, they foam so heavily that the proportioning pumps draw air and the machine cannot hold its conductivity.

    The acid concentrate carries the calcium and magnesium; brought together with bicarbonate at full strength, those cations form insoluble carbonates and carbon dioxide comes out of solution. The machine avoids this by proportioning each concentrate separately into a large stream of treated water, where the dilution keeps the salts in solution.

    Ref QB-Q8AV6TSource: ANSI/AAMI standard for concentrates for hemodialysis, bicarbonate and acid concentrate handlingReport a problem with this question

  14. 14. Before a treatment the extracorporeal circuit is primed and rinsed with normal saline. What is the main purpose of this step?

    • A.To displace all air from the dialyzer and the lines and rinse the dialyzer as the manufacturer's instructions direct.✓ Answer
    • B.To coat the fibers with saline protein so that heparin can be reduced for patients at risk of bleeding that day.
    • C.To warm the dialyzer fibers so that blood entering them does not cool and clot during the first minutes of the run.
    • D.To test the venous pressure alarm limits, since saline flowing through the circuit reproduces the pressures of blood.

    Priming fills the circuit with saline so that no air remains to be pumped toward the patient, and the rinse removes manufacturing residues and any storage or germicide solution from the dialyzer. Saline contains no protein, so it cannot coat fibers or substitute for anticoagulation.

    Ref QB-8RF2STSource: Core Curriculum for the Dialysis Technician, extracorporeal circuit setup and primingReport a problem with this question

  15. 15. Ten minutes into treatment the pre-pump arterial pressure becomes steadily more negative and the line pulls flat with each pump stroke. What should the technician do FIRST?

    • A.Raise the blood pump speed to pull past the obstruction and then watch whether the arterial pressure moves back toward zero.
    • B.Move the arterial pressure limits wider so the alarm clears, then finish the run and report the reading at the end.
    • C.Lower the blood pump speed and inspect the arterial needle and line for kinking or a needle lying against the wall.✓ Answer
    • D.Give a saline bolus into the venous line and increase the ultrafiltration rate to compensate for the volume given.

    An increasingly negative pre-pump pressure means the pump is demanding more blood than the inflow can deliver, from a kink, a clamp, a needle against the vessel wall, or an access that cannot supply the set rate. Reducing pump speed relieves the suction while the inflow is inspected; raising the speed increases the negative pressure and can damage the access and the red cells.

    Ref QB-R8N7CUSource: Core Curriculum for the Dialysis Technician, extracorporeal pressure monitoring and troubleshootingReport a problem with this question

  16. 16. Two hours into a treatment the venous pressure has risen steadily although the blood pump speed has not been changed, and the venous drip chamber looks dark. What is the most likely cause?

    • A.A separation at the venous needle connection, which lets blood escape and lowers the resistance the pump must meet.
    • B.Clot forming in the venous drip chamber and in the dialyzer, which narrows the path the blood takes back to the patient.✓ Answer
    • C.Air entering ahead of the pump, which compresses in the chamber and cushions the pressure the monitor can sense.
    • D.Hypotension in the patient, which reduces the volume the access can deliver and lowers pressure across the circuit.

    Venous pressure reflects resistance between the dialyzer and the patient, so a steady rise at an unchanged pump speed means that path is narrowing, most often from clot in the drip chamber, the dialyzer or the venous needle. Line separation and hypotension both reduce resistance and would move the venous pressure down instead of up.

    Ref QB-9RTCXUSource: Core Curriculum for the Dialysis Technician, venous pressure monitoring and circuit clottingReport a problem with this question

  17. 17. The venous pressure falls abruptly toward zero and the alarm sounds while the blood pump keeps turning. What should the technician do FIRST?

    • A.Add saline to the venous drip chamber until the level rises, then reset the alarm and continue the treatment as ordered.
    • B.Lower the venous pressure alarm floor so the machine will run, then examine the drip chamber level at the next check.
    • C.Raise the blood pump speed so the venous pressure climbs back inside the limits, then look for the cause of the drop.
    • D.Stop the blood pump, clamp the lines, and check the venous needle and every connection for separation from the patient.✓ Answer

    A sudden loss of venous pressure means the circuit has lost resistance, and the most dangerous explanation is a disconnected line or a dislodged venous needle pumping blood onto the floor or the bed. The protective action comes first: stop the pump and clamp, then inspect the access and connections and call the nurse before restarting anything.

    Ref QB-S8ZHHVSource: Core Curriculum for the Dialysis Technician, venous pressure alarms and needle dislodgementReport a problem with this question

  18. 18. The air detector alarms, the blood pump stops and the venous line clamp closes. What is the correct response?

    • A.Restart the pump at the slowest speed so any bubbles present will dissolve before the blood returns to the patient.
    • B.Override the detector so the pump restarts, then watch the drip chamber closely for the rest of the treatment time.
    • C.Keep the pump stopped and the line clamped, find where air entered, clear the circuit and have the nurse assess the patient.✓ Answer
    • D.Tap the sensor housing until the alarm clears, since a low chamber level rather than real air usually triggers it.

    The detector and clamp are the last barrier between an air embolus and the patient's circulation, so the machine's stop must be respected until the source is identified and the air removed. Overriding the monitor, tapping the sensor or restarting slowly all allow air to be pumped into the venous return, and any suspicion that air reached the patient requires immediate nursing assessment.

    Ref QB-AR6Q4WSource: Core Curriculum for the Dialysis Technician, air detection and safety monitorsReport a problem with this question

  19. 19. A blood leak alarm sounds and a test of the dialysate outflow is positive for blood. What is the correct action?

    • A.End the treatment and discard the blood in the circuit rather than returning it through the damaged dialyzer.✓ Answer
    • B.Return the blood to the patient with saline, then replace the dialyzer and finish the remaining time on the run.
    • C.Silence the alarm and lower the transmembrane pressure, since a small leak usually seals as the fibers rewet.
    • D.Rinse the dialysate side clear with fresh dialysate and continue, since the detector reads bubbles as false blood.

    A confirmed blood leak means the membrane is ruptured and the blood in the circuit has been exposed to non-sterile dialysate, so it must not be rinsed back into the patient. The treatment is terminated, the circuit discarded, the nurse and biomedical staff notified and the event documented; only a negative confirmatory test points to a false alarm from bubbles or a soiled sensor.

    Ref QB-T8CVPWSource: Core Curriculum for the Dialysis Technician, blood leak detection and responseReport a problem with this question

  20. 20. Midway through a treatment the transmembrane pressure climbs steadily although the ultrafiltration goal and the pump speed are unchanged, and streaks of the dialyzer are darkening. What does this indicate?

    • A.Fibers are clotting, so the surface left for fluid removal shrinks and the machine needs more pressure to meet the same goal.✓ Answer
    • B.The heparin dose is too high, so blood thins and crosses the membrane more easily, lifting the reading.
    • C.Dialysate flow has stopped, so the pressure on that side falls away and the reading drifts up without any clotting.
    • D.Fibers are wetting fully at last, so the membrane passes fluid more freely and the reading rises with the better flow.

    Transmembrane pressure is the driving pressure the machine must generate to pull the ordered volume of fluid across the membrane. As fibers clot, effective membrane surface falls, so the same ultrafiltration goal demands progressively higher pressure; the darkening streaks confirm the clotting, and the nurse should be notified so anticoagulation and the circuit can be evaluated.

    Ref QB-HZF8ZGSource: Kallenbach, Review of Hemodialysis for Nurses and Dialysis Personnel, transmembrane pressure and dialyzer clottingReport a problem with this question

  21. 21. A facility that reprocesses dialyzers hands the technician one for the next patient. What must be confirmed before it is connected?

    • A.That the label carries this patient's own name and number and that the rinsed dialyzer tests free of residual germicide.✓ Answer
    • B.That the fiber bundle volume is unchanged from new and that it has been used fewer times than the patient's last one.
    • C.That the housing carries no cracks and that the dialyzer has been stored in the clinic for at least one full day.
    • D.That the ports were capped with fresh caps and that the dialyzer was rinsed with the same water used to mix concentrate.

    A reprocessed dialyzer belongs to one patient only, so identity must be verified against the label before it is used, and the germicide used for storage must be rinsed out and shown absent by test, because residual disinfectant entering the blood causes hemolysis and can be fatal. Fiber bundle volume is checked during reprocessing but does not stay identical to new, and a dialyzer that fails any of these checks is discarded.

    Ref QB-ZGA3EGSource: ANSI/AAMI standard for reprocessing of hemodialyzers, residual germicide testing and patient identificationReport a problem with this question

  22. 22. During reprocessing, a dialyzer's total cell volume (TCV) measures 76% of the value recorded when it was new. What should happen to this dialyzer?

    • A.Discard it, because a reused dialyzer must keep at least 80% of its original TCV.✓ Answer
    • B.Use it again, because a reused dialyzer must keep at least 70% of its original TCV.
    • C.Keep it in use, but add treatment time to make up for the clearance that was lost.
    • D.Use it again, because the UF controller makes up for any drop in fiber volume.

    Under the AAMI reuse standard adopted by 42 CFR 494.50(b)(1), TCV is measured after each use and must be at least 80% of the original value. A 20% loss of fiber volume corresponds to roughly a 10% loss of clearance, so a dialyzer below 80% fails and must not be returned to the patient; the technician does not change the treatment time or rely on the UF controller to make up for it.

    Ref QB-JZTJ6JSource: CMS ESRD Interpretive Guidance V336: ANSI/AAMI RD47:2002/A1:2003 11.3.1 (TCV at least 80% of original), adopted by 42 CFR 494.50(b)(1)Report a problem with this question

  23. 23. A reprocessed dialyzer is taken from storage. Its germicide's maker requires a presence test, and the test shows no germicide in the blood compartment. What should the technician do?

    • A.Start rinsing it, since a negative result shows the germicide has already cleared.
    • B.Keep it from use, since it may never have been properly disinfected in storage.✓ Answer
    • C.Start rinsing it, since the residual test after priming will catch any problem.
    • D.Let it sit one more hour, since germicide often seeps back into the blood side.

    The presence test is done before rinsing to prove the dialyzer was filled with germicide at an effective strength during storage. A negative presence test means the dialyzer may not have been disinfected, so it must be held out of use and reported per the reuse procedure; a later residual (absence) test only shows the germicide is gone, not that disinfection ever happened.

    Ref QB-2HNDKHSource: CMS ESRD Interpretive Guidance V350: ANSI/AAMI RD47:2002/A1:2003 12.3.1 (germicide presence test before rinsing), adopted by 42 CFR 494.50(b)(1)Report a problem with this question

  24. 24. A reprocessed dialyzer was rinsed, primed and tested free of residual germicide. The patient then arrives 90 minutes late, and flow through the circuit was stopped while it waited. What should be done before starting?

    • A.Start right away, since the earlier negative residual test still covers this run.
    • B.Start right away, but give extra saline to dilute any trace of the germicide.
    • C.Rinse the circuit again and repeat the residual test, since germicide can rebound.✓ Answer
    • D.Replace just the bloodlines, since germicide collects in the tubing, not the fibers.

    Germicide disperses into the solid parts of the dialyzer and can leach back out after rinsing, especially when circulation stops, which is called rebound. AAMI RD47 and the CMS guidance require a maximum time between rinsing and starting dialysis, and if the circuit was prepared ahead, the residual test must be repeated just before treatment, with more rinsing until the level is below the maximum allowed.

    Ref QB-KZ6WBKSource: CMS ESRD Interpretive Guidance V353: ANSI/AAMI RD47:2002/A1:2003 12.4.1 (residual germicide testing; rebound after delay), adopted by 42 CFR 494.50(b)(1)Report a problem with this question

  25. 25. During a treatment shift, the resistivity monitor after the final deionizer (DI) bed alarms because the product water has fallen below 1 megohm-cm. What must happen?

    • A.Keep treating, because the RO ahead of it has already removed most dissolved ions.
    • B.Raise the alarm set point so it clears, then log the reading for the next tank swap.
    • C.Regenerate the softener, because hardness breakthrough is what lowers resistivity.
    • D.Stop that water from reaching any point of use and tell the nurse; the bed is spent.✓ Answer

    Falling resistivity means the DI resin is exhausted and ions are passing, or even being released, into the water; exhausted deionizers have caused fatal fluoride exposure. AAMI requires a continuous resistivity monitor with audible and visual alarms in the patient area, product water below 1.0 megohm-cm must be prevented from reaching any point of use (for example by diverting to drain), and patients must not be dialyzed on it.

    Ref QB-3HZQRJSource: CMS ESRD Interpretive Guidance V202-V203: ANSI/AAMI RD52:2004 5.2.8 and RD62:2001 4.3.6 (DI resistivity at least 1.0 megohm-cm, alarm, divert to drain), adopted by 42 CFR 494.40(a)Report a problem with this question

  26. 26. The purified-water distribution loop in a unit includes an ultraviolet (UV) irradiator. A new technician asks what job it does there. What is the best answer?

    • A.It breaks down any chloramine that has slipped past both carbon tanks.
    • B.It removes the endotoxin released by bacteria that have already died.
    • C.It controls bacterial growth in the stored and circulating purified water.✓ Answer
    • D.It captures dissolved metals such as aluminum that pass through the RO.

    Per AAMI RD52, a UV irradiator in storage and distribution uses a low-pressure mercury lamp at 254 nm to control bacterial proliferation. It does not remove endotoxin; killing bacteria can actually release it, and the ultrafilter is the component that removes endotoxin. Chloramine removal is the carbon tanks' job, and dissolved metals are removed by RO or DI.

    Ref QB-MZH8HMSource: CMS ESRD Interpretive Guidance V214: ANSI/AAMI RD52:2004 5.3.4.1 (UV irradiators, 254 nm, bacterial control in storage and distribution); NNCC CCHT Certification Preparation Guide, sample item 26 rationale (ultrafilter removes endotoxins)Report a problem with this question

  27. 27. A single-pass machine was chemically disinfected after the last shift yesterday. Before setting it up for today's first patient, what must the technician confirm and document?

    • A.A test after rinsing shows no residual disinfectant in the fluid path.✓ Answer
    • B.A presence test after rinsing shows disinfectant still in the fluid path.
    • C.The conductivity reads in range, proving the disinfectant has rinsed out.
    • D.The rinse cycle ran its full length, so no separate test is required.

    CMS guidance expects each facility's machine disinfection procedure to include testing for residual disinfectant. A residual (absence) test after the rinse proves the chemical will not reach the patient's blood, where it can cause hemolysis and other harm; conductivity is not a specific test for the disinfectant, and a completed rinse cycle is not proof by itself. The presence test belongs before the rinse, not after it.

    Ref QB-4HC3XKSource: CMS ESRD Interpretive Guidance V122, 42 CFR 494.30(a)(1) (machine disinfection procedures must include testing for residual disinfectant)Report a problem with this question

  28. 28. Midway through a treatment, the machine sounds a low conductivity alarm and goes into bypass. What should the technician check FIRST?

    • A.Whether the heater has let the dialysate temperature drift too low.
    • B.Whether the ultrafiltration goal was entered higher than ordered.
    • C.Whether the RO is sending water with too few dissolved minerals.
    • D.Whether the acid or bicarbonate jug is empty or its wand came out.✓ Answer

    Conductivity measures the total electrolyte content that the acid and bicarbonate concentrates add to treated water, so when one concentrate runs out or its wand or connector pulls loose, conductivity falls. Running out of concentrate is the most common cause of a low conductivity alarm; purified water is supposed to be low in minerals, and temperature or UF settings do not lower conductivity.

    Ref QB-DZ2WACSource: NNCC CCHT Certification Preparation Guide, sample item 14 rationale (most common cause of low conductivity alarm is running out of acid or bicarbonate concentrate; Core Curriculum p. 115)Report a problem with this question

  29. 29. A patient who often cramps and drops blood pressure late in treatment gets a new order for an ultrafiltration (UF) profile. How does a typical profile change fluid removal?

    • A.It removes less fluid early in the run and more toward the end.
    • B.It removes more fluid early in the run and less toward the end.✓ Answer
    • C.It removes the same fluid each hour but lowers the total goal.
    • D.It removes fluid only during the final hour of the treatment.

    UF profiling removes fluid in a planned pattern, typically more in the first part of treatment while the patient is most fluid-overloaded and less later, to reduce hypotension and cramping while still reaching the same UF goal. The goal itself is part of the prescription and is not lowered by the profile.

    Ref QB-VGVQQBSource: NNCC CCHT Certification Preparation Guide, sample item 27 rationale (UF profiling removes more fluid early to reduce hypotension; Review of Hemodialysis for Nurses and Dialysis Personnel pp. 87-88)Report a problem with this question

  30. 30. During treatment, blood backs up the venous pressure monitor line and wets the external transducer protector. What should the technician do?

    • A.Leave it in place, since the protector still blocks blood once it dries.
    • B.Wipe it dry with alcohol, then reconnect it to restore the pressure reading.
    • C.Clamp that line and finish the run without any venous pressure reading.
    • D.Replace it at once and check for fluid on the side facing the machine.✓ Answer

    A wetted transducer protector no longer reliably blocks blood from the machine's internal pressure monitor and can falsify the pressure reading. CDC guidance adopted by CMS says to replace it immediately and inspect it; if fluid has reached the machine side, qualified staff must open the machine after the treatment to check for blood contamination, and a contaminated machine is taken out of service and disinfected.

    Ref QB-EXF3GDSource: CMS ESRD Interpretive Guidance V120: CDC RR-5 adopted by 42 CFR 494.30(a)(1)(i) (external transducer protectors; replace when wet, check machine side)Report a problem with this question

  31. 31. A machine passed its alarm and pressure holding tests before yesterday's last treatment and has not been moved since. A coworker suggests skipping the tests this morning to save time. What is correct?

    • A.Skip them, because a passed test stays valid until the machine is moved
    • B.Skip them, because the tests are only required after a repair or service
    • C.Run the alarm and pressure holding tests again before this treatment✓ Answer
    • D.Run only the conductivity check, because it also confirms the alarms work

    CMS guidance says that, as the manufacturer requires, safety features such as alarms and pressure holding tests are tested before each dialysis treatment, not once a day or only after service. A conductivity check verifies the dialysate, not the alarm circuits, so it cannot replace these tests.

    Ref QB-DX4RACSource: CMS ESRD Interpretive Guidance V403, 42 CFR 494.60(b) (testing of alarms, pressure holding tests and independent pH/conductivity verification prior to each treatment)Report a problem with this question

  32. 32. While setting up, a technician places a spare saline-filled drip chamber in the machine's air detector so the machine will stop alarming before the patient's lines are ready. Why is this practice prohibited in the treatment area?

    • A.It wastes saline that should be saved for the patient's prime
    • B.If left in place, air could reach the patient with no alarm✓ Answer
    • C.It lets the blood leak detector miss a ruptured dialyzer fiber
    • D.It makes the venous pressure read falsely high during treatment

    A 'dummy' drip chamber is a fluid-filled chamber used to bypass the air detector. If it is not removed when dialysis starts, the real venous line is not being watched and air can be infused undetected, so CMS treats having these chambers in the patient treatment area as an immediate and serious threat; they are allowed only for maintenance outside that area.

    Ref QB-WEAWVCSource: CMS ESRD Interpretive Guidance V403, 42 CFR 494.60(b) ("dummy" drip chambers bypass air detectors; presence in treatment area is an immediate and serious threat)Report a problem with this question

  33. 33. A machine repeatedly fails its self-test during setup. The patient is moved to another station and starts treatment safely. What should be done with the failed machine?

    • A.Leave it at the station and retry the self-test next shift
    • B.Power it off and on until it passes, then note the restart
    • C.Use it only for short treatments until biomed can visit
    • D.Remove it from service, tag it, report it and document it✓ Answer

    CMS guidance requires that malfunctioning machines awaiting repair be removed from service and labeled or tagged so no one uses them by mistake. Reporting and documenting the malfunction lets the equipment be repaired and lets the facility track the event, while restarting until it passes hides the fault.

    Ref QB-CXRD5BSource: CMS ESRD Interpretive Guidance V403, 42 CFR 494.60(b) (malfunctioning machines awaiting repair must be removed from service and labeled or tagged)Report a problem with this question

  34. 34. A facility's policy cites these limits for product water: total viable count under 200 CFU/mL with an action level of 50 CFU/mL. This month's culture returns 80 CFU/mL. What does this result call for?

    • A.Prompt disinfection of the system and a repeat culture✓ Answer
    • B.No action, because the result is still under the maximum
    • C.Stopping all treatments until the next culture is clean
    • D.Replacing the RO membrane before the next patient shift

    Action levels exist because culture results take about 48 hours and bacteria multiply quickly, so corrective action must start before the maximum is reached. AAMI rationale adopted by CMS says dialysis may continue between the action level and the maximum, but the system should be disinfected promptly; stopping dialysis is a medical director decision, not an automatic step.

    Ref QB-VEXJQBSource: CMS ESRD Interpretive Guidance V178: ANSI/AAMI RD52:2004 4.1.2 and rationale A.4.1.2 (action levels; prompt disinfection when exceeded), adopted by 42 CFR 494.40(a)Report a problem with this question

  35. 35. Several patients on one shift develop fever and shaking chills within an hour of starting, and their blood cultures later come back negative. Which contaminant in the water or dialysate best explains this?

    • A.Chloramine that broke through the carbon adsorption beds
    • B.Calcium and magnesium left behind by an exhausted softener
    • C.Endotoxin from the cell walls of gram-negative bacteria✓ Answer
    • D.Aluminum carried over from the facility's municipal feed

    Pyrogenic reactions are caused by lipopolysaccharides (endotoxins) from gram-negative bacteria, which multiply readily in purified water, and endotoxin fragments can cross both low-flux and high-flux membranes. Because the patients receive bacterial products rather than live bacteria, they can have fever and chills while their blood cultures stay negative.

    Ref QB-BXD3Y9Source: CMS ESRD Interpretive Guidance V178: AAMI rationale A.4.1.2 (pyrogenic reactions caused by endotoxins of gram-negative bacteria; endotoxins cross low- and high-flux membranes)Report a problem with this question

  36. 36. During the daily water room check, the RO feed water conductivity reads 400 µS/cm and the product (permeate) conductivity reads 20 µS/cm. What is the percent rejection?

    • A.5.0%
    • B.95.0%✓ Answer
    • C.20.0%
    • D.80.0%

    Percent rejection equals (feed conductivity minus permeate conductivity) divided by feed conductivity, times 100: (400 - 20) / 400 x 100 = 95%. The 5.0% figure is the fraction that passed through the membrane, not the fraction it rejected; a falling rejection trend signals membrane fouling, scaling or damage.

    Ref QB-UEK8JASource: CMS ESRD Interpretive Guidance V200: ANSI/AAMI RD52:2004 6.2.7 (Rejection (%) = (feed water conductivity - permeate conductivity) / feed water conductivity x 100)Report a problem with this question

  37. 37. With the system running, the gauge before the sediment filter reads 70 and the gauge after it reads 50. Facility policy sets the maximum pressure drop across this filter at 10. What should be done?

    • A.Nothing yet, since the outlet reading of 50 is above the limit of 10
    • B.Raise the feed pressure until the outlet gauge also reads near 70
    • C.Bypass the filter and rely on the carbon tanks to catch the sediment
    • D.Backwash or replace the filter, since the drop of 20 exceeds the limit✓ Answer

    The pressure drop (delta P) is the inlet reading minus the outlet reading, here 70 - 50 = 20, and it is an index of resistance to flow through the filter. A delta P above the facility limit means the filter is loaded with sediment and needs backwashing or replacement; comparing a single gauge to the limit, or raising pressure, misses the point of the measurement.

    Ref QB-JXVD6JSource: CMS ESRD Interpretive Guidance V188: ANSI/AAMI RD52:2004 Table 4 (sediment filter pressure drop monitored daily; delta P above facility limit means backwash or replace; worked example 70 - 50 = 20)Report a problem with this question

  38. 38. A facility's carbon beds are plumbed in parallel: two separate pairs of tanks, each treating about half of the water. Before the shift, how should total chlorine be tested?

    • A.Sample each set at its own port, as one set tells nothing of the other✓ Answer
    • B.Sample only the busier set, since it will exhaust well before the other
    • C.Sample after the RO only, since the membrane blends both streams' water
    • D.Sample the city feed only, since both sets receive the identical water

    In a parallel layout each set of beds treats its own share of the water, so one set can be exhausted while the other is fine. CMS guidance therefore requires sample ports for each set, because testing one set is irrelevant to the function of the other, and a blended or feed sample would hide a breakthrough in one branch.

    Ref QB-3F3KRJSource: CMS ESRD Interpretive Guidance V193: ANSI/AAMI RD52:2004 5.2.5 (parallel carbon beds need sample ports for each set; testing one set is irrelevant to the other), adopted by 42 CFR 494.40(a)Report a problem with this question

  39. 39. A technician mixing bicarbonate concentrate from powder leaves the mixer running long past the set mixing time. What problem can this cause?

    • A.Extra dissolution that raises the final sodium too high
    • B.CO2 loss, a higher pH and calcium carbonate precipitation✓ Answer
    • C.Bubbles that set off the machine's air detector alarm
    • D.Dextrose breakdown that raises the dialysate glucose level

    Overagitation drives CO2 out of the bicarbonate solution, which raises the pH and favors carbonate formation; after proportioning, calcium and magnesium carbonate can precipitate in the machine's fluid paths. Losing that calcium can lower the dialysate calcium and the patient's serum calcium, which is why mixers need a timer or a staff cut-off. Bicarbonate concentrate contains no dextrose; dextrose is in the acid concentrate.

    Ref QB-Q5GJ2RSource: CMS ESRD Interpretive Guidance V234: ANSI/AAMI RD52:2004 5.4.4.3 and rationale A.5.4.4.3 (overmixing bicarbonate causes CO2 loss, higher pH and calcium/magnesium carbonate precipitation)Report a problem with this question

  40. 40. A batch of acid concentrate has just been mixed from powder in the facility's mixing tank. Before it is released to the distribution system, how should correct mixing be verified?

    • A.Test pH, since an off-strength mix will show a clear pH change
    • B.Check that no powder is visible and the solution looks clear
    • C.Test conductivity or specific gravity against the expected range✓ Answer
    • D.Run the mixer for an extra full cycle so it is surely dissolved

    AAMI RD52, as adopted by CMS, says acid and bicarbonate concentrates may be tested by conductivity or with a hydrometer, and the batch is not released until the results are checked against the expected range and recorded. pH is not an appropriate indicator of proper dissolution, because large changes in concentration produce little change in pH, and a clear-looking solution proves nothing about strength.

    Ref QB-7NADFQSource: CMS ESRD Interpretive Guidance V229: ANSI/AAMI RD52:2004 6.4.1 (concentrates tested by conductivity or hydrometer; pH inappropriate as indicator of proper dissolution; not used until tests completed)Report a problem with this question

  41. 41. Midway through a treatment, the physician orders a switch to an acid concentrate with a different potassium content. After the new jug is connected, what must the technician do before dialysate from it reaches the patient?

    • A.Trust the machine display once the alarms have cleared
    • B.Wait for the next treatment to run any new checks
    • C.Check only the dialysate temperature on the screen
    • D.Check pH and conductivity with an independent device✓ Answer

    CMS guidance requires a pH test with a hand-held meter or adequately sensitive strips before every treatment and whenever a different acid concentrate composition is used, and an independent conductivity test at those times when the manufacturer requires it. Certain mix-ups involving the acid concentrate can give an acceptable conductivity with a very low pH, so the machine display alone is not enough.

    Ref QB-P558VPSource: CMS ESRD Interpretive Guidance V250: ANSI/AAMI RD52:2004 5.6 (pH test before every treatment and whenever a different acid concentrate composition is used; acceptable conductivity possible with very low pH)Report a problem with this question

  42. 42. A patient's pre-treatment weight is 2.5 kg above the target weight, and the nurse asks the technician to set the ultrafiltration goal to remove exactly that weight gain. What volume should be entered?

    • A.250 mL
    • B.2,500 mL✓ Answer
    • C.1,136 mL
    • D.5,500 mL

    Fluid weight is converted at 1 kg = 1,000 mL, so a 2.5 kg gain equals a 2,500 mL ultrafiltration goal. The 5,500 mL distractor comes from converting kilograms to pounds (2.5 x 2.2) and the 1,136 mL figure from dividing by 2.2, both mixing up weight units with volume.

    Ref QB-6NX2APSource: NNCC CCHT Certification Preparation Guide, sample item 16 rationale (fluid removal conversion 1 kg = 1,000 mL)Report a problem with this question

Other topics in this exam

Practice questions based on the NNCC CCHT content outline, CDC recommendations for hemodialysis settings, and the OSHA Bloodborne Pathogens Standard. Not affiliated with or endorsed by NNCC, and not medical advice. Machine settings, dialysate composition, flow rates, alarm thresholds, water-quality limits and laboratory targets are deliberately never the answer to a question here — those are set by your machine's manufacturer, your facility's protocol and the physician's prescription. Always follow your facility's policies and your patient's prescribed treatment. About the CCHT exam (NNCC) →