Dialysate: concentrates, conductivity and temperature
1 min read
The machine makes dialysate by proportioning acid concentrate and bicarbonate concentrate separately into a large stream of treated water. The two concentrates are never combined at full strength, because the calcium and magnesium in the acid concentrate would react with bicarbonate, form insoluble carbonates, and release carbon dioxide. Dilution keeps those salts in solution.
Conductivity stands in for the total ion content of the dialysate. Low conductivity means the fluid is too dilute, hypotonic to blood, so water moves into red cells until they burst; this is the more rapidly lethal direction. Dialysate that is too hot also destroys red cells, releasing potassium and turning venous blood translucent cherry-red, with back and chest pain.
When conductivity or temperature is out of range the machine goes into bypass, which diverts dialysate to the drain so it never meets the blood. Dialysis effectively stops in bypass, so the cause is fixed rather than the bypass defeated.
Before the patient is connected, the dialysate is measured with a separate, calibrated meter and both readings are recorded, because the machine's own cell can drift and cannot check itself.
Example: the machine shows a value inside the range and the meter shows a value that is also inside the range but differs by more than the manufacturer's tolerance. The two still disagree, so the machine is tagged, reported to biomedical staff, and the patient moves to a machine that verifies.
Check yourself
1. A conductivity alarm shows a value well below the set range. Compared with correctly proportioned dialysate, what would this fluid do to the patient's blood?
2. The machine display shows conductivity and temperature inside the set range and no alarms are active. What must the technician still do before the patient is connected?
Facts checked 2026-11-15 · Sources: NNCC: Certified Clinical Hemodialysis Technician (CCHT), 42 CFR Part 494: Conditions for Coverage for ESRD Facilities