Two carbon tanks are installed in series. The first, the worker tank, does most of the adsorbing; the second, the polisher, is a backup.
The routine sample is drawn from the port between the two tanks, because a high result there reveals that the worker tank is exhausted while the polisher is still protecting patients. Sampling only at the end of the system would hide that early warning.
- Result between the tanks over the facility limit: test right away after the polisher tank.
- Result after the polisher within limit: treatment may proceed under protocol while the worker tank is serviced.
- Result after the polisher also over the limit: hold all treatments, stop any running, notify the nurse in charge and biomedical staff, and keep the water out of use.
Chloramine is an oxidant that attacks red cells directly. When it slips past exhausted carbon, the pattern is unexplained hemolytic anemia across patients who share the same water: falling hemoglobin, rising erythropoietin needs, and sometimes chest or back pain. Endotoxin looks different, causing chills and fever during the run, and aluminum injury shows up as bone and nerve disease.
Example: the early-morning test between the tanks is high. Rebedding both tanks or switching to bypass is not the first step; the next sample after the polisher decides whether the shift can start.
Check yourself
1. A technician is preparing the water treatment system before the first patients of the day. Where should the sample for chlorine testing be drawn, and why?
2. Over two weeks several patients on the same shift show falling hemoglobin and rising erythropoietin requirements, and one reports chest and back pain during the run. Which water problem best explains this pattern?
Facts checked 2026-11-15 · Sources: NNCC: Certified Clinical Hemodialysis Technician (CCHT), 42 CFR Part 494: Conditions for Coverage for ESRD Facilities
Dialysis Technician cheat sheet: every key fact on one page →