Reading the circuit pressures
1 min read
Each pressure monitor tells you about resistance in one part of the blood circuit. Read every change against the pump speed: if the pump has not changed, the reading moved because something in the circuit or the access changed.
| Reading | Likely cause | First action |
|---|---|---|
| Arterial (pre-pump) more and more negative, line pulling flat | Inflow cannot keep up: kink, clamp, needle against the vessel wall | Lower pump speed, inspect needle and line |
| Venous rising steadily, dark drip chamber | Clot in the drip chamber, dialyzer or venous needle | Check for clots and kinks, notify the nurse |
| Venous falling suddenly toward zero | Line separation or dislodged venous needle | Stop the pump, clamp, check the needle and every connection |
| Transmembrane pressure climbing with the same goal | Fibers clotting, less membrane left | Notify the nurse so anticoagulation and the circuit can be evaluated |
Example: ten minutes into a run the arterial pressure keeps getting more negative. Speeding up the pump to pull past the problem only makes the suction worse and can damage the access and the red cells, so the pump is slowed while the inflow is checked.
Check yourself
1. Two hours into a treatment the venous pressure has risen steadily although the blood pump speed has not been changed, and the venous drip chamber looks dark. What is the most likely cause?
2. Ten minutes into treatment the pre-pump arterial pressure becomes steadily more negative and the line pulls flat with each pump stroke. What should the technician do FIRST?