Dialysis Technician Cheat Sheet: Key Facts

Every key takeaway and common trap from the Dialysis Technician study guide on one page, grouped by chapter. Tap a lesson link to read the full explanation.

33 facts · 5 chaptersFacts checked 2026-11-15
  1. Renal Physiology & Dialysis Principles
  2. The Machine & Water Treatment
  3. Vascular Access
  4. Intradialytic Complications
  5. Infection Control, Safety & Professional Role

Chapter 1

Renal Physiology & Dialysis Principles

Key takeaways

Common trap

  • Watch for answers that sound logical but run backward, such as slowing the blood so it spends more time in the fibers. Less blood presented to the membrane means less waste removed, so a slower blood flow lowers clearance. → Diffusion, the membrane and clearance
  • In weight questions, distractors use the wrong conversion (100 mL per kilogram) or read the goal straight off one weight. The goal comes from the difference between the two weights, at 1,000 mL per kilogram. → Ultrafiltration, osmosis and convection

Quick reference

What failing kidneys stop doing →
Lost functionWhat follows
Erythropoietin productionThe marrow makes fewer red cells; anemia is treated with a marrow-stimulating drug
Vitamin D activationLess calcium is absorbed, calcium falls, phosphorus rises, and the parathyroid glands release more hormone
Sodium and water excretionVolume expands; with renin release, blood pressure rises
Acid excretion and buffer balanceBicarbonate falls and metabolic acidosis develops
Ultrafiltration, osmosis and convection →
TermWhat movesDriving force
DiffusionSoluteConcentration difference
UltrafiltrationWaterPressure difference
OsmosisWaterToward the side with more dissolved particles
Convection (solvent drag)Solute carried with waterUltrafiltration

Chapter 2

The Machine & Water Treatment

Key takeaways

Common trap

  • Ultrafilters, softeners and reverse osmosis membranes do not remove chloramine. Any answer that relies on one of them to make chlorinated water safe is wrong; only the carbon stage does that job. → The water treatment train
  • Having a second person read the same machine display is not verification. It is the instrument that is in doubt, not the person reading it, so only an independent meter counts. → Dialysate: concentrates, conductivity and temperature
  • Widening the alarm limits, lowering the alarm floor, or topping up the drip chamber so the machine will keep running never fixes a pressure problem. These answers silence the warning while the cause, such as a clot or a disconnected needle, keeps getting worse. → Reading the circuit pressures

Quick reference

The water treatment train →
StageWhat it doesIf it fails
SoftenerTrades calcium and magnesium for sodium by ion exchange; regenerated with salt from the brine tankHardness passes on and scales the reverse osmosis membrane
Carbon tanksAdsorb chlorine and chloramineOxidant reaches patients and also damages the membrane
Reverse osmosisRejects dissolved ions, bacteria and endotoxinContaminants such as aluminum reach the dialysate
Deionizer, if usedPolishes out remaining ions; must be followed by an ultrafilterThe resin bed grows bacteria that release endotoxin
Reading the circuit pressures →
ReadingLikely causeFirst action
Arterial (pre-pump) more and more negative, line pulling flatInflow cannot keep up: kink, clamp, needle against the vessel wallLower pump speed, inspect needle and line
Venous rising steadily, dark drip chamberClot in the drip chamber, dialyzer or venous needleCheck for clots and kinks, notify the nurse
Venous falling suddenly toward zeroLine separation or dislodged venous needleStop the pump, clamp, check the needle and every connection
Transmembrane pressure climbing with the same goalFibers clotting, less membrane leftNotify the nurse so anticoagulation and the circuit can be evaluated

Chapter 3

Vascular Access

Key takeaways

Common trap

  • Answers that sound protective for grafts are often wrong: buttonhole is never used on a synthetic graft, because a scar track cannot form through the graft wall. Rotation, not a constant site, is the routine method for grafts. → Placing the needles

Quick reference

Look, listen and feel before every use →
FindingLikely meaning
Hard water-hammer pulse, high-pitched systolic-only bruit, whole arm swollenOutflow stenosis
No thrill, no bruit, access cool and firm like a cordThrombosis (clotted access)
Redness, warmth, tenderness, drainage at a needle siteAccess infection
Hand cold, pale, numb and aching, worst during dialysisSteal syndrome: the access diverts blood from the hand

Chapter 4

Intradialytic Complications

Key takeaways

Common trap

  • Stopping the blood pump feels like the safe answer for low blood pressure, but blood that stops moving clots in the circuit and the patient loses that volume. The pump is stopped only when continued circulation would harm the patient, such as air, hemolysis, a blood leak, a dislodged needle or cardiac arrest. → Intradialytic hypotension and cramps

Quick reference

Blood loss, dialyzer reactions and fever →
PictureLikely causeTechnician response
Hives, swollen lips, wheezing in the first minutesType A (anaphylactic) dialyzer reactionStop, clamp without returning blood, call the nurse
Chest and back pain appearing later in the runType B dialyzer reaction (milder)Report; the nurse decides whether to continue with oxygen and observation
Chills, new fever and falling pressure an hour inPyrogenic reaction from endotoxin, or a catheter bloodstream infectionFull vital signs, keep the patient warm, notify the nurse immediately

Chapter 5

Infection Control, Safety & Professional Role

Key takeaways

Common trap

Quick reference

Hepatitis B, hepatitis C and staff vaccination →
Patient statusRoom and machineDialyzer reuse
HBsAg positiveSeparate room, dedicated machine and suppliesNot reused
Anti-HCV positiveUsual station and machine, routine precautionsMay take part