Dialysis Technician Cheat Sheet: Key Facts
Every key takeaway and common trap from the Dialysis Technician study guide on one page, grouped by chapter. Tap a lesson link to read the full explanation.
33 facts · 5 chaptersFacts checked 2026-11-15
Chapter 1
Renal Physiology & Dialysis Principles
Key takeaways
- Blood is filtered only at the glomerulus, and the tubules then reabsorb nearly all of that filtrate. → How healthy kidneys make urine
- Dialysis replaces filtering but not the kidney's hormones, so anemia and bone disease still need drug treatment. → What failing kidneys stop doing
- Diffusion moves solutes down a concentration gradient, while ultrafiltration uses pressure to remove water. → Diffusion, the membrane and clearance
- Countercurrent flow, a larger membrane, faster blood flow and longer treatment all raise clearance. → Diffusion, the membrane and clearance
- Weight gain between treatments converts one to one, so each kilogram is one liter of fluid to remove. → Ultrafiltration, osmosis and convection
Common trap
- Watch for answers that sound logical but run backward, such as slowing the blood so it spends more time in the fibers. Less blood presented to the membrane means less waste removed, so a slower blood flow lowers clearance. → Diffusion, the membrane and clearance
- In weight questions, distractors use the wrong conversion (100 mL per kilogram) or read the goal straight off one weight. The goal comes from the difference between the two weights, at 1,000 mL per kilogram. → Ultrafiltration, osmosis and convection
Quick reference
| Lost function | What follows |
|---|---|
| Erythropoietin production | The marrow makes fewer red cells; anemia is treated with a marrow-stimulating drug |
| Vitamin D activation | Less calcium is absorbed, calcium falls, phosphorus rises, and the parathyroid glands release more hormone |
| Sodium and water excretion | Volume expands; with renin release, blood pressure rises |
| Acid excretion and buffer balance | Bicarbonate falls and metabolic acidosis develops |
| Term | What moves | Driving force |
|---|---|---|
| Diffusion | Solute | Concentration difference |
| Ultrafiltration | Water | Pressure difference |
| Osmosis | Water | Toward the side with more dissolved particles |
| Convection (solvent drag) | Solute carried with water | Ultrafiltration |
Chapter 2
The Machine & Water Treatment
Key takeaways
- Carbon tanks run in series ahead of the reverse osmosis unit, and carbon is the stage that removes chlorine and chloramine. → The water treatment train
- A high chlorine result between the tanks means test after the second tank; high there too means no treatments. → Testing for chlorine and chloramine
- Dialysate is checked with a separate calibrated meter, and a machine that differs by more than the manufacturer's tolerance is taken out of service. → Dialysate: concentrates, conductivity and temperature
- Rising venous pressure at a steady pump speed means a narrowing path, and a sudden drop means a possible disconnection. → Reading the circuit pressures
- Safety stops from the air detector or blood leak detector are respected until the cause is found, never overridden. → Dialysate: concentrates, conductivity and temperature
Common trap
- Ultrafilters, softeners and reverse osmosis membranes do not remove chloramine. Any answer that relies on one of them to make chlorinated water safe is wrong; only the carbon stage does that job. → The water treatment train
- Having a second person read the same machine display is not verification. It is the instrument that is in doubt, not the person reading it, so only an independent meter counts. → Dialysate: concentrates, conductivity and temperature
- Widening the alarm limits, lowering the alarm floor, or topping up the drip chamber so the machine will keep running never fixes a pressure problem. These answers silence the warning while the cause, such as a clot or a disconnected needle, keeps getting worse. → Reading the circuit pressures
Quick reference
| Stage | What it does | If it fails |
|---|---|---|
| Softener | Trades calcium and magnesium for sodium by ion exchange; regenerated with salt from the brine tank | Hardness passes on and scales the reverse osmosis membrane |
| Carbon tanks | Adsorb chlorine and chloramine | Oxidant reaches patients and also damages the membrane |
| Reverse osmosis | Rejects dissolved ions, bacteria and endotoxin | Contaminants such as aluminum reach the dialysate |
| Deionizer, if used | Polishes out remaining ions; must be followed by an ultrafilter | The resin bed grows bacteria that release endotoxin |
| Reading | Likely cause | First action |
|---|---|---|
| Arterial (pre-pump) more and more negative, line pulling flat | Inflow cannot keep up: kink, clamp, needle against the vessel wall | Lower pump speed, inspect needle and line |
| Venous rising steadily, dark drip chamber | Clot in the drip chamber, dialyzer or venous needle | Check for clots and kinks, notify the nurse |
| Venous falling suddenly toward zero | Line separation or dislodged venous needle | Stop the pump, clamp, check the needle and every connection |
| Transmembrane pressure climbing with the same goal | Fibers clotting, less membrane left | Notify the nurse so anticoagulation and the circuit can be evaluated |
Chapter 3
Vascular Access
Key takeaways
- A native fistula is preferred because it uses the patient's own vessels, so it infects less and stays open longer. → Fistula, graft or catheter
- A healthy access has a soft continuous thrill and a low continuous bruit; a hard pulse or a high-pitched systolic bruit suggests stenosis. → Look, listen and feel before every use
- The venous needle points toward the heart and sits downstream of the arterial needle to limit recirculation. → Placing the needles
- Rotate sites rope-ladder style, and use buttonhole cannulation only in fistulas, never in grafts. → Placing the needles
- Signs of clotting, infection or steal mean the access is not used until the nurse has been told. → Look, listen and feel before every use
Common trap
- Answers that sound protective for grafts are often wrong: buttonhole is never used on a synthetic graft, because a scar track cannot form through the graft wall. Rotation, not a constant site, is the routine method for grafts. → Placing the needles
Quick reference
| Finding | Likely meaning |
|---|---|
| Hard water-hammer pulse, high-pitched systolic-only bruit, whole arm swollen | Outflow stenosis |
| No thrill, no bruit, access cool and firm like a cord | Thrombosis (clotted access) |
| Redness, warmth, tenderness, drainage at a needle site | Access infection |
| Hand cold, pale, numb and aching, worst during dialysis | Steal syndrome: the access diverts blood from the hand |
Chapter 4
Intradialytic Complications
Key takeaways
- Yawning, sighing, restlessness, nausea and cramping are early warnings of hypotension, before the pressure collapses. → Intradialytic hypotension and cramps
- For simple hypotension, per protocol, turn ultrafiltration off and lay the patient flat with legs raised, but keep the blood pump running. → Intradialytic hypotension and cramps
- Disequilibrium threatens new patients with very high urea, because urea leaves the blood faster than the brain. → Look-alikes: low sugar, disequilibrium and mental status
- Air embolism, hemolysis and a type A dialyzer reaction mean stop, clamp and never return the blood. → Blood loss, dialyzer reactions and fever
- The access and line connections stay uncovered and visible so a dislodged needle is seen at once. → Blood loss, dialyzer reactions and fever
Common trap
- Stopping the blood pump feels like the safe answer for low blood pressure, but blood that stops moving clots in the circuit and the patient loses that volume. The pump is stopped only when continued circulation would harm the patient, such as air, hemolysis, a blood leak, a dislodged needle or cardiac arrest. → Intradialytic hypotension and cramps
Quick reference
| Picture | Likely cause | Technician response |
|---|---|---|
| Hives, swollen lips, wheezing in the first minutes | Type A (anaphylactic) dialyzer reaction | Stop, clamp without returning blood, call the nurse |
| Chest and back pain appearing later in the run | Type B dialyzer reaction (milder) | Report; the nurse decides whether to continue with oxygen and observation |
| Chills, new fever and falling pressure an hour in | Pyrogenic reaction from endotoxin, or a catheter bloodstream infection | Full vital signs, keep the patient warm, notify the nurse immediately |
Chapter 5
Infection Control, Safety & Professional Role
Key takeaways
- Change gloves and clean your hands between every patient and every station, even when only the machine is touched. → Infection control at the dialysis station
- Anything taken to a station is used for that patient, discarded, or disinfected, and never returned to a clean cart. → Infection control at the dialysis station
- HBsAg-positive patients dialyze in a separate room on a dedicated machine; anti-HCV-positive patients need routine precautions only. → Hepatitis B, hepatitis C and staff vaccination
- After a needlestick, wash with soap and water first, then report right away. → Sharps, exposures and emergencies
- Prescription changes, a patient stopping early, and suspected abuse all go to the nurse, and the technician documents. → Patient rights: privacy, refusal and grievances
Common trap
- Questions about hepatitis C often offer isolation, a dedicated machine or a move to the last shift. None of these applies; only HBsAg-positive patients need isolation measures. → Hepatitis B, hepatitis C and staff vaccination
Quick reference
| Patient status | Room and machine | Dialyzer reuse |
|---|---|---|
| HBsAg positive | Separate room, dedicated machine and supplies | Not reused |
| Anti-HCV positive | Usual station and machine, routine precautions | May take part |