Quibank app: now on the App Store — free, works offline. Get the app ↗

Chapters0/23 ▾
← The Machine & Water Treatment

Chapter 2 · Lesson 1 of 5

The water treatment train

1 min read

A patient's blood meets a very large volume of dialysate through the membrane at every treatment, so the water used to make it must be far cleaner than drinking water.

Each stage of the treatment system removes a specific problem and often protects the stage after it. Exam questions usually ask why a component sits where it does, or what happens when it fails.

StageWhat it doesIf it fails
SoftenerTrades calcium and magnesium for sodium by ion exchange; regenerated with salt from the brine tankHardness passes on and scales the reverse osmosis membrane
Carbon tanksAdsorb chlorine and chloramineOxidant reaches patients and also damages the membrane
Reverse osmosisRejects dissolved ions, bacteria and endotoxinContaminants such as aluminum reach the dialysate
Deionizer, if usedPolishes out remaining ions; must be followed by an ultrafilterThe resin bed grows bacteria that release endotoxin

Example: a new technician asks why carbon is not placed after the reverse osmosis unit. The answer has two parts: the membrane cannot remove chlorine or chloramine, and those oxidants would degrade the membrane itself. Carbon upstream solves both problems at once.

Check yourself

1. A new technician asks why the carbon tanks are plumbed ahead of the reverse osmosis unit rather than after it. What is the best answer?

2. The hardness test on water leaving the softener is suddenly high and the brine tank is empty. What is the consequence if the system keeps running this way?

Facts checked 2026-11-15 · Sources: NNCC: Certified Clinical Hemodialysis Technician (CCHT), 42 CFR Part 494: Conditions for Coverage for ESRD Facilities

Dialysis Technician cheat sheet: every key fact on one page →