Chapter 2 · Lesson 5 of 5
Priming, safety monitors and reused dialyzers
1 min read
Before treatment the circuit is primed and rinsed with normal saline. Priming pushes all air out of the dialyzer and lines so none can be pumped toward the patient, and the rinse washes out manufacturing residues and any storage germicide as the manufacturer directs. Saline has no protein, so it does not coat fibers or replace anticoagulation.
The air detector and venous line clamp are the last barrier against an air embolus. When the detector alarms, the pump stops and the clamp closes; the technician keeps it that way, finds where air entered, clears the circuit, and has the nurse assess the patient.
Overriding the detector, tapping the sensor or restarting the pump slowly would all let air into the venous return.
A confirmed blood leak, meaning the dialysate outflow tests positive for blood, means the membrane is torn and the blood has met non-sterile dialysate. The treatment ends and that blood is discarded rather than returned, and the nurse and biomedical staff are notified. Only a negative confirmation test points to a false alarm.
Example: a facility that reprocesses dialyzers hands over one for the next patient. Two checks come before connection: the label names this patient, and the rinsed dialyzer tests free of residual germicide, because germicide in the blood causes hemolysis.
Check yourself
1. A blood leak alarm sounds and a test of the dialysate outflow is positive for blood. What is the correct action?
2. Before a treatment the extracorporeal circuit is primed and rinsed with normal saline. What is the main purpose of this step?