Biometry: measuring axial length for lens implants
1 min read
Before cataract surgery the assistant often measures axial length, the length of the eye, which feeds the formula that chooses the lens implant. Axial length is a major source of error in that calculation, and an error of about 1 mm shifts the result by roughly 2.5 to 3 diopters, so small technique errors become refractive surprises after surgery.
- Contact A-scan indents the cornea, so the eye measures shorter than it is. A shorter eye calls for a stronger lens, so the implant comes out too strong and the patient ends up nearsighted. Immersion keeps fluid between probe and cornea and avoids this.
- The eye-type setting matters because sound travels at different speeds through a natural lens, an implant, aqueous and vitreous, or silicone oil. The wrong setting gives a systematically wrong length.
- Optical biometry uses light, so a very dense cataract can block it. Measure that eye with immersion ultrasound; the fellow eye never substitutes.
Example: the right eye measures 23.1 mm and the left 24.0 mm, but the two eyes have nearly the same glasses prescription. Nothing explains a 0.9 mm difference, so remeasure both eyes until the readings repeat, rather than averaging them or passing them on.
Check yourself
1. Compared with immersion ultrasound, how does contact (applanation) A-scan biometry bias the measurement and the lens power calculated from it?
2. Axial length measures 23.4 mm in one eye and 24.6 mm in the fellow eye, and nothing in the refraction explains the difference. What should the assistant do?