Quibank app: now on the App Store — free, works offline. Get the app ↗

Chapters0/28 ▾
  1. 1.General Medical Knowledge & History0/4
    1. Taking and documenting the ophthalmic history
    2. Triage: which eye complaints cannot wait
    3. Eye anatomy and physiology the exam expects
    4. Systemic disease, medications and chart terms
  2. 2.Visual Acuity, Pupils & Motility0/4
    1. Measuring and recording visual acuity
    2. The pupil exam and the afferent pupillary defect
    3. Extraocular muscles and the cardinal positions of gaze
    4. Cover tests, stereopsis and convergence
  3. 3.Tonometry, Visual Fields & Biometry0/4
    1. Goldmann applanation tonometry and its errors
    2. Visual fields: reading defects and avoiding artifacts
    3. Biometry: measuring axial length for lens implants
    4. Supplemental tests: Amsler grid, color vision and glare
  4. 4.Refraction, Lensometry & Optics0/4
    1. Reading a prescription: transposition, spherical equivalent and vertex
    2. Retinoscopy and subjective refraction endpoints
    3. Lensometry: reading glasses and prism
    4. Keratometry and corneal astigmatism
  5. 5.Ophthalmic Imaging & Photography0/4
    1. Fluorescein angiography: filters, phases and what the frames show
    2. Dye studies: screening, side effects and reactions
    3. OCT, corneal topography and ultrasound
    4. Photography technique, slit-lamp lighting and labeling
  6. 6.Pharmacology, Microbiology & Surgical Assisting0/4
    1. How dilating, numbing and pressure-lowering drops work
    2. Giving drops and dyes safely
    3. Germs, hand hygiene and cleaning clinic equipment
    4. Sterile technique and assisting with minor surgery
  7. 7.Patient Services, Equipment & Ethics0/4
    1. Teaching patients to use eye drops and stay on treatment
    2. Communicating, identifying patients and triaging urgent calls
    3. Caring for tonometers, lenses and other equipment
    4. Consent, privacy, scope of practice and reporting errors
Saved in this browser. Move it to another device →
← Tonometry, Visual Fields & Biometry

Chapter 3 · Lesson 3 of 4

Biometry: measuring axial length for lens implants

1 min read

Before cataract surgery the assistant often measures axial length, the length of the eye, which feeds the formula that chooses the lens implant. Axial length is a major source of error in that calculation, and an error of about 1 mm shifts the result by roughly 2.5 to 3 diopters, so small technique errors become refractive surprises after surgery.

  • Contact A-scan indents the cornea, so the eye measures shorter than it is. A shorter eye calls for a stronger lens, so the implant comes out too strong and the patient ends up nearsighted. Immersion keeps fluid between probe and cornea and avoids this.
  • The eye-type setting matters because sound travels at different speeds through a natural lens, an implant, aqueous and vitreous, or silicone oil. The wrong setting gives a systematically wrong length.
  • Optical biometry uses light, so a very dense cataract can block it. Measure that eye with immersion ultrasound; the fellow eye never substitutes.

Example: the right eye measures 23.1 mm and the left 24.0 mm, but the two eyes have nearly the same glasses prescription. Nothing explains a 0.9 mm difference, so remeasure both eyes until the readings repeat, rather than averaging them or passing them on.

Check yourself

1. Compared with immersion ultrasound, how does contact (applanation) A-scan biometry bias the measurement and the lens power calculated from it?

2. Axial length measures 23.4 mm in one eye and 24.6 mm in the fellow eye, and nothing in the refraction explains the difference. What should the assistant do?