Quibank app: now on the App Store — free, works offline. Get the app ↗

Chapters0/28 ▾
  1. 1.General Medical Knowledge & History0/4
    1. Taking and documenting the ophthalmic history
    2. Triage: which eye complaints cannot wait
    3. Eye anatomy and physiology the exam expects
    4. Systemic disease, medications and chart terms
  2. 2.Visual Acuity, Pupils & Motility0/4
    1. Measuring and recording visual acuity
    2. The pupil exam and the afferent pupillary defect
    3. Extraocular muscles and the cardinal positions of gaze
    4. Cover tests, stereopsis and convergence
  3. 3.Tonometry, Visual Fields & Biometry0/4
    1. Goldmann applanation tonometry and its errors
    2. Visual fields: reading defects and avoiding artifacts
    3. Biometry: measuring axial length for lens implants
    4. Supplemental tests: Amsler grid, color vision and glare
  4. 4.Refraction, Lensometry & Optics0/4
    1. Reading a prescription: transposition, spherical equivalent and vertex
    2. Retinoscopy and subjective refraction endpoints
    3. Lensometry: reading glasses and prism
    4. Keratometry and corneal astigmatism
  5. 5.Ophthalmic Imaging & Photography0/4
    1. Fluorescein angiography: filters, phases and what the frames show
    2. Dye studies: screening, side effects and reactions
    3. OCT, corneal topography and ultrasound
    4. Photography technique, slit-lamp lighting and labeling
  6. 6.Pharmacology, Microbiology & Surgical Assisting0/4
    1. How dilating, numbing and pressure-lowering drops work
    2. Giving drops and dyes safely
    3. Germs, hand hygiene and cleaning clinic equipment
    4. Sterile technique and assisting with minor surgery
  7. 7.Patient Services, Equipment & Ethics0/4
    1. Teaching patients to use eye drops and stay on treatment
    2. Communicating, identifying patients and triaging urgent calls
    3. Caring for tonometers, lenses and other equipment
    4. Consent, privacy, scope of practice and reporting errors
Saved in this browser. Move it to another device →
← Ophthalmic Imaging & Photography

Chapter 5 · Lesson 1 of 4

Fluorescein angiography: filters, phases and what the frames show

1 min read

Fluorescein glows when blue light hits it and gives off yellow-green light. The camera's exciter filter sends blue light into the eye, and the barrier filter in front of the sensor lets only the yellow-green glow through, so reflected blue light does not wash out the picture.

Before any dye is injected, the photographer takes color and red-free (green-filtered) photos. Blood and pigment look dark in red-free light, and these frames are the control that shows whether later brightness is true fluorescence or was there before the dye.

The dye arrives in a fixed order: a patchy choroidal flush, then the retinal arteries, then the arteriovenous phase with streams of dye along the vein walls, then full venous filling, and finally the late frames after the dye recirculates. Bright areas are named by how they change over time.

PatternEarlyLate
LeakageSmall and faintLarger, brighter, blurred edges
Window defectBrightFades, same size
StainingDimBrighter, fixed borders

Example: a spot is barely visible at 30 seconds and at 8 minutes it has doubled in size with fuzzy edges. Only leakage both grows and brightens, so that is the mechanism.

Check yourself

1. On an angiogram, an area of hyperfluorescence is small and faint early, then becomes larger and brighter with blurred margins on the late frames. Which mechanism does this describe?

2. Why are color and red-free (green-filtered) fundus photographs taken before the fluorescein is injected?

Facts checked 2026-11-15 · Sources: IJCAHPO: Certified Ophthalmic Assistant (COA), AAO EyeWiki: Fluorescein Angiography, FDA label: Fluorescite (fluorescein injection)