Photography technique, slit-lamp lighting and labeling
1 min read
Before fundus photos, the photographer focuses the crosshair in the eyepiece with accommodation relaxed, turning from plus toward minus. The crosshair sits in the same plane as the sensor, so this cancels the photographer's own refractive error; then a fundus that looks sharp in the eyepiece will record sharp.
If the camera is at the right distance but one edge shows a bright crescent and the other side is dark, part of the illumination ring is being blocked by the iris. Recenter the camera or improve dilation.
| Slit-lamp technique | Best for |
|---|---|
| Retroillumination | Posterior capsule haze, seen dark against the red reflex |
| Specular reflection | The endothelial cell mosaic |
| Sclerotic scatter | Subtle corneal haze |
| Conical beam | Cells and flare |
Specular microscopy images the endothelium and reports cell density and the variation in cell size and shape. Nystagmus is recorded on video, labeled with the gaze position and head posture, because a still frame cannot show movement. Left-eye photos filed under the right eye are corrected before the study is filed, since laterality guides treatment.
Clinical photographs are protected health information, and agreeing to be photographed for care does not cover publication, outside lectures or marketing. Identifiable photos used for a lecture or social media therefore need a separate signed authorization from the patient that names the intended use.
Check yourself
1. With the camera at the correct working distance, a fundus photograph shows a bright crescent of light along one edge of the frame with the opposite side underexposed. What is the most likely cause?
2. The practice wants to use an identifiable external photograph of a patient in a lecture and on its social media. What is required?
Facts checked 2026-11-15 · Sources: IJCAHPO: Certified Ophthalmic Assistant (COA), AAO EyeWiki: Fluorescein Angiography, FDA label: Fluorescite (fluorescein injection)