Quibank app: now on the App Store — free, works offline. Get the app ↗

Chapters0/28 ▾
  1. 1.General Medical Knowledge & History0/4
    1. Taking and documenting the ophthalmic history
    2. Triage: which eye complaints cannot wait
    3. Eye anatomy and physiology the exam expects
    4. Systemic disease, medications and chart terms
  2. 2.Visual Acuity, Pupils & Motility0/4
    1. Measuring and recording visual acuity
    2. The pupil exam and the afferent pupillary defect
    3. Extraocular muscles and the cardinal positions of gaze
    4. Cover tests, stereopsis and convergence
  3. 3.Tonometry, Visual Fields & Biometry0/4
    1. Goldmann applanation tonometry and its errors
    2. Visual fields: reading defects and avoiding artifacts
    3. Biometry: measuring axial length for lens implants
    4. Supplemental tests: Amsler grid, color vision and glare
  4. 4.Refraction, Lensometry & Optics0/4
    1. Reading a prescription: transposition, spherical equivalent and vertex
    2. Retinoscopy and subjective refraction endpoints
    3. Lensometry: reading glasses and prism
    4. Keratometry and corneal astigmatism
  5. 5.Ophthalmic Imaging & Photography0/4
    1. Fluorescein angiography: filters, phases and what the frames show
    2. Dye studies: screening, side effects and reactions
    3. OCT, corneal topography and ultrasound
    4. Photography technique, slit-lamp lighting and labeling
  6. 6.Pharmacology, Microbiology & Surgical Assisting0/4
    1. How dilating, numbing and pressure-lowering drops work
    2. Giving drops and dyes safely
    3. Germs, hand hygiene and cleaning clinic equipment
    4. Sterile technique and assisting with minor surgery
  7. 7.Patient Services, Equipment & Ethics0/4
    1. Teaching patients to use eye drops and stay on treatment
    2. Communicating, identifying patients and triaging urgent calls
    3. Caring for tonometers, lenses and other equipment
    4. Consent, privacy, scope of practice and reporting errors
Saved in this browser. Move it to another device →
← Ophthalmic Imaging & Photography

Chapter 5 · Lesson 3 of 4

OCT, corneal topography and ultrasound

1 min read

Optical coherence tomography builds a cross-section of the retina from reflected light. It compares light coming back from the tissue with a reference beam, much as ultrasound times echoes, but it needs no dye, no injection and no contact with the eye.

Scan quality matters: a dry tear film, a blink or a cloudy lens scatters the beam, lowers the signal strength and can make a layer look falsely thin. A low-quality scan is repeated after lubricating the eye and asking for a full blink, not saved as a result.

Placido-disc topography reads the reflection of lighted rings off the tear film. The standard color scale shows curvature: warm red and orange mean steeper, higher-power cornea; cool blue means flatter cornea.

It does not measure thickness. Because the rings reflect from the tear film, contact lens warpage and a dry or broken tear film give broken, irregular rings; leave lenses out beforehand, blink and measure again.

Example: a patient has a dense vitreous hemorrhage and the physician cannot see the retina. Photography, angiography and OCT all depend on light reaching the retina and returning, so all three fail. B-scan ultrasound sends sound through the blood and can still show a detachment or a mass.

Check yourself

1. A patient has a dense vitreous hemorrhage and no view of the fundus. Which study best assesses the retina behind it?

2. A patient asks how an optical coherence tomography scan produces its image. Which explanation is accurate?