Chapter 7 · Lesson 3 of 4
Caring for tonometers, lenses and other equipment
1 min read
A reusable applanation prism touches the tear film, so it must be disinfected between patients. After use on a red, watery eye, adenovirus is the concern. It is a tough virus that a quick alcohol wipe or dilute hydrogen peroxide does not reliably kill, so the prism is soaked in dilute sodium hypochlorite, rinsed well with water and air dried.
Heat sterilization ruins the prism, and the soak is kept to the time the protocol states, since the disinfectant can attack the cement and craze the surface. Inspect the tip for damage after reprocessing.
Example: a tonometer is knocked off the slit lamp and looks fine. Do not keep using it or subtract a little from each reading.
An instrument that is out of calibration gives believable but wrong numbers that drive treatment, so check its calibration with the weight bar as the manufacturer directs, log the event and send it for service if needed.
Coated lenses scratch easily. Blow loose dust off a slit lamp objective first, then wipe with a lens tissue moistened with cleaner. Put the cleaner on the tissue, not on the instrument, where it can run into the housing. Ammonia and acetone dissolve coatings and lens cement, and a dry cotton ball drags grit across the glass.
Check yourself
1. A reusable applanation prism has just been used on a patient with a red, watery eye. How should it be reprocessed before the next patient?
2. The objective lens of a slit lamp has a dusty film on it. How should it be cleaned?
Facts checked 2026-11-15 · Sources: IJCAHPO: Certified Ophthalmic Assistant (COA), CDC: Preventing Epidemic Keratoconjunctivitis (EKC), 45 CFR 164.510: Uses and disclosures requiring an opportunity to agree or object