Chapters0/30 ▾
0 of 30 lessons finished
- 1.Asepsis & Sterile Technique0/5
- Where sterile ends: gowns, tables, movement and the prep
- Breaks in technique: strike-through, glove punctures and wound class
- Judging packages, trays and solutions before they reach the field
- Reading sterilization indicators and biological results
- Cleaning first, Spaulding, and choosing a sterilization method
- 2.Instrumentation & Sutures0/5
- 3.Surgical Procedures0/5
- 4.Patient Safety & Counts0/5
- 5.Professional Practice0/5
- Informed consent: who explains, who witnesses, who may sign
- Roles, scope of practice and owning your errors
- Documentation: count records, corrections and occurrence reports
- Legal doctrines: negligence, res ipsa loquitur, respondeat superior, battery and abandonment
- Privacy, do-not-resuscitate orders and advance directives
- 6.Basic Science0/5
Chapter 2 · Lesson 5 of 5
Hemostasis categories, electrosurgery and powered equipment
1 min read
Hemostasis is grouped by how it works. Mechanical methods physically stop the flow: clamps, ligatures, clips, pressure, packing and bone wax, which simply plugs the channels in cut bone. Thermal methods use energy devices. Chemical methods use agents that act on clotting or on vessels.
In bipolar electrosurgery the two poles are the two forceps tips, so current flows only through the tissue between them and no return electrode is needed. Monopolar current crosses the patient to a return electrode, and in laparoscopy it can stray.
Insulation failure means current leaks through a crack. Direct coupling means the active tip touches another metal instrument. Capacitive coupling means current passes through intact insulation into a nearby conductor, such as a metal cannula.
- A lit fiberoptic cable is never left on the drapes; its end concentrates enough light to scorch or ignite cloth.
- With a power drill, irrigate steadily to prevent heat damage to bone, and set the safety lock before every handoff or bur change.
Example: a burn appears on bowel far from the tip, the insulation is intact and nothing touched the electrode. Only capacitive coupling fits all three facts.
Check yourself
1. During laparoscopy a loop of bowel is burned by an active monopolar electrode that never touched it; the insulation is later found intact and the tip touched no other instrument. Which mechanism explains the injury?
2. A bleeding cut sternal edge is sealed with bone wax. Which category of hemostasis does that action belong to?
Facts checked 2026-11-15 · Sources: NBSTSA: Surgical Technologist Certifying Examination Content Outline, Association of Surgical Technologists: Standards of Practice