Spotting a misplaced or swapped electrode

1 min read

The technician does not interpret the tracing, but must recognize patterns that point to an acquisition error rather than to the heart. The response to every pattern below is the same: correct the electrodes and repeat the tracing before it is mounted or sent.

What the tracing showsLikely cause
Lead I inverted, aVR upright, chest leads normalRight and left arm electrodes reversed
One limb lead nearly flat, others normalRight leg electrode swapped with another limb electrode
rSr' complexes and inverted P or T waves in V1 and V2V1 and V2 placed one or two spaces too high
R-wave progression reverses between two neighboring chest leadsChest electrodes out of order

Example: lead II is a flat line and everything else looks normal. Lead II is built from the right arm and the left leg, so a flat lead II fits a right arm and right leg swap: the lead now measures two points at almost the same potential.

A flat lead III would instead fit a left arm and right leg swap. Either way, fix the cables and record again.

Check yourself

1. A tracing shows an inverted P wave, QRS and T wave in lead I with an upright aVR, while R-wave progression across V1 to V6 is entirely normal. What should the technician conclude?

2. A technician notices that the R wave is taller in V3 than in V4 and that those two leads look interchanged, while every other lead is normal. What should the technician do first?

Facts checked 2026-11-15 · Sources: AHA/ACCF/HRS Recommendations for the Standardization and Interpretation of the ECG, Part I (Circulation, 2007), NHA Certified EKG Technician (CET) certification and test plan

EKG Technician (CET) cheat sheet: every key fact on one page →