Chapter 4 · Lesson 2 of 4
Fixing artifact at its source
1 min read
Each artifact has a physical cause, and the correct first step removes that cause. Machine settings come last, if at all.
- AC interference: move phones, chargers and power cords away from the patient and keep lead wires straight along the limbs. Use the 60 Hz notch filter only if noise persists, because it hides the noise and can blunt sharp waveforms such as pacemaker spikes.
- Somatic tremor: warm a shivering patient with a blanket, have the patient lie flat with arms resting and supported, and for a constant hand tremor move the limb electrodes toward the trunk on the same limb (upper arms, thighs).
- Wandering baseline: clean off lotion or oil with alcohol and let it dry, replace dried or tacky electrodes with fresh ones from a sealed pouch, lay the cable on the bed so wires hang slack, and coach quiet breathing.
Example: the room is cold, an older patient is shivering, and every lead is buried in jagged noise even though the electrodes are new and secure. The electrodes are not the problem, so replacing them will not help; a blanket and a few minutes to settle will.
Check yourself
1. A tracing carries the uniform 60-cycle fuzz and the source of the interference has not yet been found. Why is engaging the 60 Hz notch filter treated as the last step rather than the first?
2. A technician takes electrodes from a foil pouch that was opened four days earlier and left on the counter. The gel looks dull and feels tacky rather than moist. Several leads then show a drifting baseline with intermittent sharp spikes. What is the correct response?
Facts checked 2026-11-15 · Sources: NHA Certified EKG Technician (CET) certification and test plan, AHA/ACCF/HRS Recommendations for the Standardization and Interpretation of the ECG, Part I (Circulation, 2007)