Chapter 4 · Lesson 4 of 4
Machine settings, honest records and artifact versus a real rhythm
1 min read
The standard settings are 25 mm/sec paper speed and 10 mm/mV sensitivity. Readers assume those values, so any change must be marked on the tracing.
At half standardization (5 mm/mV) every complex prints at half its true height. At 50 mm/sec every interval prints twice as wide, so a rate calculated as if the paper were standard comes out at half the true rate.
Example: at 50 mm/sec, R waves are 30 small boxes apart. Read as standard, 1500 ÷ 30 = 50, but each box is really 0.02 sec, so the true rate is 3000 ÷ 30 = 100. At 5 mm/mV, a complex printed 8 mm tall is 1.6 mV, which would be 16 mm at standard sensitivity.
- A tracing is a medical record: never smooth, erase or trim it, and never mount only the clean leads. If artifact remains after every fix, record the best tracing possible and note that it could not be cleared.
- An ST segment that rises and falls with a drifting baseline is a technical fault; fix the drift and repeat instead of reporting it.
- On a chaotic monitor strip, normal QRS complexes still marching through the noise at the old rate suggest artifact, but the technician still goes to check the patient.
Check yourself
1. Complexes in V2 and V3 are so tall that they overlap the adjacent channels, so the technician reduces sensitivity to 5 mm/mV and records a readable tracing. What must accompany that change?
2. A patient with an involuntary tremor cannot stay still. After the room is warmed, the limbs supported and the sites re-prepped, a fine tremor artifact remains in the limb leads, although all complexes are measurable. How should the technician handle the tracing?