27 Patient Care Equipment Practice Questions & Answers
Every Patient Care Equipment practice question from the Sterile Processing (CRCST) Practice Test, with the correct answer and a short explanation.
Start practice test →1. An infusion pump comes back from a nursing unit with dried blood streaked across its housing and keypad. The department uses an EPA-registered hospital disinfectant wipe. What should the technician do?
- A.Remove the visible blood first, because soil shields organisms from the disinfectant✓ Answer
- B.Send it to decontamination for steam processing, because blood makes it a critical item
- C.Spray the keypad with alcohol first, because alcohol lifts dried blood faster than detergent
- D.Wipe it once with the disinfectant wipe, because the product cleans and kills in one pass
Blood and other organic matter react with germicides and shield microorganisms, so visible soil must be removed before the disinfectant can work, and the product must then stay in contact for the time on its label. Blood on the housing does not change the pump's Spaulding category, because that category depends on what tissue the device contacts, and the pump touches only intact skin.
Ref QB-E8YB6GSource: CDC/HICPAC Guideline for Disinfection and Sterilization in Healthcare Facilities, Factors Affecting the Efficacy of Disinfection (organic and inorganic matter) and Recommendation on low-level disinfection of noncritical patient-care devicesReport a problem with this question
2. A disinfectant label says the surface must stay visibly wet for 3 minutes. A technician wipes a sequential compression device (SCD) pump and, after 30 seconds, towels it dry so it can be bagged. What is the correct practice?
- A.Towel it dry at 30 seconds, because studies show hospital disinfectants work within 1 minute
- B.Towel it dry now, because label contact times apply to instruments, not to equipment housings
- C.Keep it wet for the full 3 minutes, because the label contact time is legally binding✓ Answer
- D.Re-wipe it after towel-drying, because two short passes add up to the label contact time
CDC states that all applicable label instructions on EPA-registered products must be followed by law, and a user who chooses different exposure conditions takes on liability for off-label use. Studies of shorter contact times do not change the label, and two short passes interrupted by drying do not equal one continuous wet contact period.
Ref QB-YQGUWHSource: CDC/HICPAC Guideline for Disinfection and Sterilization in Healthcare Facilities, Recommendation on disinfecting noncritical medical devices with an EPA-registered hospital disinfectant according to the label's use directionsReport a problem with this question
3. A bladder scanner was just used on a patient on contact precautions, and the next patient needs it. No dedicated scanner is available. What must happen first?
- A.It can go straight to the next patient, because precautions apply to the patient, not to equipment
- B.It must get high-level disinfection, because contact precautions make it a semicritical item
- C.It must sit unused for 24 hours, because organisms on dry surfaces die off within a day
- D.It must be cleaned and disinfected, because shared equipment can carry the organism onward✓ Answer
CDC recommends that, when dedicated disposable devices are not available, noncritical patient-care equipment used on a patient on contact precautions be disinfected before it is used on another patient. The scanner touches intact skin, so it stays noncritical; precautions change how often it is disinfected, not its Spaulding level, and waiting a set time is not a substitute for disinfection.
Ref QB-F8BPBHSource: CDC/HICPAC Guideline for Disinfection and Sterilization in Healthcare Facilities, Recommendation on disinfecting noncritical patient-care equipment after use on a patient on contact precautionsReport a problem with this question
4. An enteral feeding pump returns to the department with the used feeding set still threaded through it. What is the technician's first step?
- A.Leave the set in place for the next patient, so the pump can go back out to the unit sooner
- B.Flush the set with disinfectant, so the pump and the set can both be reused on another patient
- C.Wipe the pump and the set together, so the set shields the pump channel during disinfection
- D.Remove and discard the used set, so every surface of the pump can be cleaned and disinfected✓ Answer
The CRCST outline lists removal of consumables as part of the care and handling of patient care equipment. A used feeding set is a disposable that went with that patient; it must come off before cleaning, because surfaces it covers cannot be reached by cleaning or disinfectant, and it is never passed to another patient.
Ref QB-ZQU64KSource: HSPA CRCST Exam Content Outline, Section 6 (Patient Care Equipment and Distribution), Care and Handling: removal of consumables, cleaningReport a problem with this question
5. The emergency department calls for an infusion pump. The only clean pump on the shelf powers on normally, but its preventive maintenance (PM) label expired last month. What should the technician do?
- A.Send it with a note asking the ED to call Biomed, so the PM can be done at the bedside
- B.Send it to the ED, because a pump that powers on and runs is safe whatever the label says
- C.Hold it back for Biomedical Engineering, because an expired PM label means noncompliance✓ Answer
- D.Peel off the expired label and send it, because the label date is just paperwork for Biomed
The CRCST outline names a missing or expired preventive maintenance label as equipment out of compliance, with referral to Biomedical Engineering as the corrective action. Powering on does not show that the safety and performance checks done at PM are current, so the pump is not distributed, and removing the label would hide the problem rather than fix it.
Ref QB-G8PZGJSource: HSPA CRCST Exam Content Outline, Section 6, Inspection and Corrective action plan for equipment out of compliance (missing/expired preventative maintenance label, Biomedical Engineering)Report a problem with this question
6. While cleaning a hypothermia unit, a technician finds that the outer jacket of its power cord is frayed and the inner wires show. What should happen to the unit?
- A.Send it to a low-acuity unit instead, so the damaged cord is used where patient risk is lower
- B.Wrap the frayed spot in electrical tape, so the unit can go out while a repair is scheduled
- C.Clean and distribute it with a note to the unit, so nursing staff can watch the cord during use
- D.Remove it from service and tag it for Biomedical Engineering, so no one uses it until repaired✓ Answer
OSHA's electrical safe-work rule requires that cord-and-plug equipment with a defect that might expose an employee to injury be removed from service, and that no one use it until the repairs and tests needed to make it safe are done. The CRCST outline lists frayed cords as an inspection finding; tape, warnings or a lower-acuity location do not make the cord safe.
Ref QB-2R7H9MSource: OSHA 29 CFR 1910.334(a)(2)(ii), defective portable cord- and plug-connected equipment removed from service; HSPA CRCST Exam Content Outline, Section 6, Equipment inspection (frayed cords, damage)Report a problem with this question
7. A portable suction unit's power cord has a flattened, crushed section where a bed wheel rolled over it. The unit powers on and passes its startup self-test. How should it be handled?
- A.Keep it in clean storage as a backup, because it should be used only in emergencies
- B.Take it out of service for repair, because a crushed jacket can signal internal damage✓ Answer
- C.Distribute it, because a passed self-test shows that the cord has no internal damage
- D.Distribute it after taping the crushed part, because tape restores the outer jacket
OSHA requires cord-and-plug equipment to be inspected for external defects and also for evidence of possible internal damage, naming a pinched or crushed outer jacket as an example. A self-test checks the device's functions, not the integrity of the cord's insulation, so a crushed cord means removal from service until it is repaired and tested.
Ref QB-H82CPKSource: OSHA 29 CFR 1910.334(a)(2)(i)-(ii), visual inspection for evidence of possible internal damage (pinched or crushed outer jacket) and removal of damaged items from serviceReport a problem with this question
8. A blood warmer must be shipped to its manufacturer for repair. One internal part could not be decontaminated. Under the OSHA Bloodborne Pathogens Standard, what is required?
- A.Hold it until that part can be autoclaved, since contaminated items may never be shipped
- B.Ship it in its original box, since the maker expects returned devices to be contaminated
- C.Label which part remains contaminated and tell the servicer, since they must take precautions✓ Answer
- D.Clean the outer case and ship it unlabeled, since inner parts fall outside the OSHA rule
The standard requires equipment that may be contaminated to be examined and decontaminated as necessary before servicing or shipping. Where part of it cannot feasibly be decontaminated, a readily observable biohazard label stating which portions remain contaminated must be attached, and that information must reach affected employees, the servicing representative or the manufacturer before handling or shipping so that appropriate precautions can be taken.
Ref QB-3RJUENSource: OSHA 29 CFR 1910.1030(d)(2)(xiv)(A)-(B), equipment that may be contaminated: label stating which portions remain contaminated and notice to the servicing representativeReport a problem with this question
9. A technician will clean a returned portable suction unit whose canister lid is splattered with secretions, using a spray-and-wipe method that could spatter. Under the OSHA Bloodborne Pathogens Standard, which protection is required?
- A.A face shield alone, because gloves are optional when using a disinfectant spray
- B.Gloves and a gown, because spatter at a counter reaches clothing, not the face
- C.Gloves alone, because eye protection is needed only when handling sharps
- D.Gloves plus a mask with goggles or a face shield, because spatter can reach the face✓ Answer
OSHA requires masks in combination with eye protection, such as goggles or glasses with solid side shields, or chin-length face shields, whenever splashes, spray, spatter or droplets of potentially infectious material may be generated and contamination of the eyes, nose or mouth can be reasonably anticipated. Gloves are still required for hand contact with contaminated surfaces.
Ref QB-J8DPUMSource: OSHA 29 CFR 1910.1030(d)(3)(x), masks, eye protection and face shieldsReport a problem with this question
10. A specialty bed is rented from a vendor for one patient. Which practice fits the tracking of rented equipment named in the CRCST outline?
- A.Log its receipt and ID on arrival, then log its return so the rental can be closed out✓ Answer
- B.Log it only when it goes back, because the vendor's invoice already shows the delivery
- C.Add it to the hospital's owned inventory, because all equipment on site is an asset
- D.Skip logging it, because a rental stays the vendor's responsibility until it is sent back
The CRCST outline lists rental and loaned equipment, along with receipt and distribution, among the processes for tracking equipment. Recording the unit when it arrives and when it leaves lets the department know where it is, confirm it went back, and stop paying for it; rented equipment remains vendor property, so it does not belong in the owned inventory.
Ref QB-7RH4ECSource: HSPA CRCST Exam Content Outline, Section 6, Equipment Tracking and Documentation: processes for tracking equipment (rental, loaned, receipt and distribution)Report a problem with this question
11. A new room for cleaning returned patient care equipment is being set up. Under the OSHA Bloodborne Pathogens Standard, what determines how often its surfaces are cleaned and by what method?
- A.A written schedule the employer sets by location, surface type, soil present and tasks done✓ Answer
- B.A schedule set by the disinfectant vendor, whose product label fixes how often rooms are cleaned
- C.An as-needed approach, cleaning only the surfaces that show visible soil at the time
- D.A daily housekeeping wipe-down, the single frequency OSHA sets for every hospital room
The standard requires the employer to keep the worksite clean and sanitary and to determine and implement an appropriate written schedule for cleaning and method of decontamination, based on the location in the facility, the type of surface, the type of soil present and the tasks performed there. OSHA does not set one frequency for all rooms, and cleaning only visible soil would miss contamination that cannot be seen.
Ref QB-Q8Q9ZCSource: OSHA 29 CFR 1910.1030(d)(4)(i), housekeeping: written schedule for cleaning and method of decontaminationReport a problem with this question
12. Clean, battery-powered infusion pumps are stored unplugged on open shelves, and nursing units keep reporting that pumps arrive with low batteries. Which change addresses this?
- A.Fit a new battery before every delivery, because stored pumps drain too quickly
- B.Store clean pumps plugged in at a charging station, so they go out fully charged✓ Answer
- C.Charge the pumps from outlets in the soiled area, because that room has free plugs
- D.Ask the units to charge each pump on arrival, because charging is the user's job
The CRCST outline lists charging stations and plugs as work area requirements, and equipment charging as part of the location and storage of equipment, which is to be dry, clean and secure. Keeping clean pumps on charge in clean storage means they leave ready to use; charging in the soiled area would put clean equipment back into contaminated space and break one-way flow.
Ref QB-8RVEKDSource: HSPA CRCST Exam Content Outline, Section 6, Work area requirements (charging stations, plugs, one way flow) and Location and storage of equipment (dry, clean, secure, equipment charging)Report a problem with this question
13. After cleaning and checking an infusion pump, a technician returns it to a unit. Nurses there often cannot tell processed pumps from used ones parked in the hallway. Which practice addresses this?
- A.Write the last patient's room on it, so nurses know where the pump came from
- B.Park clean and used pumps together, so nurses can grab whichever one is closest
- C.Wipe the pump again on the unit, so nurses can watch it being cleaned before use
- D.Put a clean label or an equipment cover on it, so it is marked as ready for use✓ Answer
The CRCST outline lists equipment covers and clean labels or stickers among the supplies needed in the patient care equipment work area. A visible clean marker separates processed equipment from used equipment at the point of use, so a pump that has not been cleaned is not picked up by mistake; writing a room number would add patient information without showing processing status.
Ref QB-R83M6ESource: HSPA CRCST Exam Content Outline, Section 6, Work Area Preparation: supplies needed (equipment covers, clean labels/stickers)Report a problem with this question
14. Two hours into the shift, the disposable plastic cover over the touchscreen of the equipment-tracking terminal in the return area is splashed with blood. Under the OSHA Bloodborne Pathogens Standard, when must the cover be removed and replaced?
- A.As soon as feasible, because the cover has now become overtly contaminated with blood✓ Answer
- B.At the next weekly cleaning, because covers exist to cut how often screens are cleaned
- C.At the end of the shift, because covers are changed on a fixed end-of-shift schedule
- D.Once the cover tears, because an intact cover still keeps the screen beneath it clean
OSHA requires protective coverings such as plastic wrap used over equipment and surfaces to be removed and replaced as soon as feasible when they become overtly contaminated, or at the end of the work shift if they may have become contaminated during it. End-of-shift replacement applies only to covers that may have been contaminated, not to covers visibly splashed with blood.
Ref QB-9R8SRESource: OSHA 29 CFR 1910.1030(d)(4)(ii)(B), protective coverings used to cover equipment and environmental surfacesReport a problem with this question
15. The ICU requests a hypothermia unit. Its instructions for use (IFU) call for attaching a disposable component and running a function check before the unit is used. How should it be prepared for distribution?
- A.Send it bare and untested, because the ICU nurse will assemble and test it at the bedside
- B.Assemble it with its disposables and run the IFU check, because the IFU sets the testing✓ Answer
- C.Run the check only if the unit looks damaged, because routine checks are for suspect units
- D.Send it assembled but untested, because it passed a function check after the last patient
The CRCST outline lists equipment assembly for distribution, including disposable components, and equipment testing per the manufacturer's IFU as steps of equipment distribution. A check done after a previous patient does not show the unit works now that it has been cleaned and assembled, and the IFU, not the device's appearance, decides when testing is done.
Ref QB-S8EXBFSource: HSPA CRCST Exam Content Outline, Section 6, Equipment Distribution: equipment assembly (disposable components, IFU) and equipment testing per manufacturer's IFUReport a problem with this question
16. The department receives a new model of infusion pump, and a technician is unsure which cleaning agents its plastic housing and display can tolerate. What should guide the choice?
- A.The method used for the old model, because pumps from one maker share their materials
- B.The operator manual or IFU, because it lists the cleaning agents the device can tolerate✓ Answer
- C.The unit nurses' usual routine, because users know the device best from daily handling
- D.The strongest disinfectant on hand, because a stronger chemical protects patients best
The CRCST outline lists the operator manual among the supplies needed for patient care equipment work, and CDC notes that FDA asks manufacturers to include a validated cleaning and disinfection protocol in device labeling, which users should know. A stronger or unlisted chemical can damage plastics and displays, and materials can change between models even from the same maker.
Ref QB-ARK4XFSource: HSPA CRCST Exam Content Outline, Section 6, Work Area Preparation (operator manual); CDC/HICPAC Guideline for Disinfection and Sterilization in Healthcare Facilities, FDA request for a validated cleaning and disinfection protocol in device labelingReport a problem with this question
17. The emergency department urgently needs a portable suction unit. None are in clean storage, but a technician sees one parked in a hallway outside a patient room with no clean label. What should the technician do?
- A.Bring the hallway unit back for processing, because an unlabeled unit counts as used✓ Answer
- B.Tell the ED to borrow one from another unit, because sourcing it is not the department's job
- C.Wipe the hallway unit on the way to the ED, because a quick wipe covers processing
- D.Take the hallway unit straight to the ED, because an urgent call outranks processing
An unlabeled unit outside a patient room must be treated as used: CDC calls for noncritical patient-care devices to be disinfected after use on each patient or on another regular basis, and the CRCST outline adds inspection and a current preventive maintenance label before distribution. A hurried wipe in transit skips the inspection and contact time, so the technician brings the unit back for processing and can tell the ED when one will be ready.
Ref QB-T8S9HGSource: CDC/HICPAC Guideline for Disinfection and Sterilization in Healthcare Facilities, Recommendation that noncritical patient-care devices be disinfected when visibly soiled and on a regular basis; HSPA CRCST Exam Content Outline, Section 6, Equipment inspection and distributionReport a problem with this question
18. A unit manager asks why bedside blood pressure cuffs are disinfected on the unit instead of being sent to central processing like surgical instruments. What is the reason?
- A.They touch mucous membranes only briefly, so a quick wipe replaces high-level disinfection
- B.They are single-use items, so they are thrown away after each patient, not reprocessed
- C.They touch only intact skin, a good barrier, so low-level disinfection where used is enough✓ Answer
- D.They are too bulky for central processing, so units handle them whatever their risk level
CDC classifies items that contact intact skin but not mucous membranes as noncritical, because intact skin is an effective barrier to most microorganisms, and lists blood pressure cuffs as an example. It notes that most noncritical reusable items may be decontaminated where they are used and need not go to central processing, and it recommends low-level disinfection for such equipment.
Ref QB-BRXE4HSource: CDC/HICPAC Guideline for Disinfection and Sterilization in Healthcare Facilities, Noncritical Items (Spaulding classification) and Recommendation on low-level disinfection of noncritical patient-care equipmentReport a problem with this question
19. To be extra safe, a technician proposes soaking reusable blood pressure cuffs in the department's glutaraldehyde high-level disinfectant. According to CDC guidance, what do these cuffs call for?
- A.A plain water rinse, because intact skin makes any disinfectant unnecessary for them
- B.An EPA-registered low-level disinfectant, because the cuffs touch only intact skin✓ Answer
- C.Steam sterilization, because the cuffs touch many different patients in a single day
- D.A glutaraldehyde soak, because shared items need the highest level the department has
CDC recommends low-level disinfection for noncritical equipment such as a blood pressure cuff that touches intact skin, using an EPA-registered hospital disinfectant per the label, and notes that EPA-registered quaternary ammonium compounds are appropriate for this use. High-level disinfection is the minimum for semicritical items that touch mucous membranes or nonintact skin, so a glutaraldehyde soak adds chemical exposure without being required, while water alone does not disinfect.
Ref QB-U85MPHSource: CDC/HICPAC Guideline for Disinfection and Sterilization in Healthcare Facilities, Recommendation on low-level disinfection of noncritical patient-care equipment (blood pressure cuff) and section on quaternary ammonium compoundsReport a problem with this question
20. A pump's three-prong plug will not fit an old two-slot outlet in a patient room. A coworker offers an adapter that leaves the grounding prong unconnected. What does OSHA's electrical safe-work rule say?
- A.The ground prong may be cut off instead, because two-prong plugs fit all outlets
- B.The adapter is acceptable briefly, because grounding matters only on long runs
- C.The pump must go to a grounded outlet, because ground continuity has to be kept✓ Answer
- D.The adapter is acceptable, because the pump's battery takes over if a fault occurs
OSHA states that attachment plugs and receptacles may not be connected or altered in a way that would prevent proper continuity of the equipment grounding conductor, and that the plug and receptacle contacts must first be checked for proper mating configuration. An adapter that leaves the ground unconnected and a cut-off prong both defeat the grounding path; the unit should be plugged into a proper grounded receptacle, with the outlet reported for repair.
Ref QB-E8ABY9Source: OSHA 29 CFR 1910.334(a)(2)(iii), (a)(3)(ii) and (a)(3)(iii), plug and receptacle mating configuration, continuity of the equipment grounding conductor, and adapters that interrupt grounding continuityReport a problem with this question
21. A department records equipment loans on paper cards at the dispatch desk and also scans each pump's barcode into a computer when it is delivered to a unit. How does the CRCST content outline classify this kind of tracking setup?
- A.A manual system
- B.A computer system
- C.A hybrid system✓ Answer
- D.An RFID system
The CRCST outline names four kinds of equipment tracking systems: manual, computer, Radio Frequency Identification (RFID) and hybrid. Paper cards are a manual method and barcode scans into a computer are a computer method, so a setup that uses both at once is a hybrid. RFID reads radio tags rather than printed barcodes, so barcode scanning alone does not make a system RFID.
Ref QB-WQ46D9Source: HSPA CRCST Exam Content Outline (rev. Nov 2023), Section 6.III.A Systems used (manual, computer, RFID, hybrid)Report a problem with this question
22. A sequential compression device (SCD) pump comes back with its compression sleeves still attached, and the sleeves are labeled single-use. A coworker suggests wiping the sleeves and sending them out again with the pump to save money. What is correct?
- A.Discard them; in-house reuse would be regulated as manufacturing✓ Answer
- B.Reuse them after a low-level disinfectant wipe and contact time
- C.Reuse them after a high-level disinfectant soak and full rinse
- D.Keep them for reuse on the same unit only, after a disinfectant wipe
SCD sleeves are a supply that goes with the pump, but sleeves labeled single-use are removed and discarded, while the pump itself is cleaned. CDC recommends following the FDA enforcement document on single-use devices: a hospital that reprocesses a single-use device is treated as the manufacturer of that device and must meet the same regulatory standards as the original manufacturer. Choosing a stronger disinfectant or limiting reuse to one unit does not remove that requirement.
Ref QB-F8NN5BSource: CDC Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), Recommendation 20.a, Reuse of Single-Use Medical Devices; HSPA CRCST Content Outline Section 6.II.A (SCD sleeves)Report a problem with this question
23. An IV pump returns from a unit with an uncapped needle still attached to the end of the administration tubing hanging from it. What should the technician do?
- A.Recap the needle, then pull it off the tubing
- B.Discard needle and tubing in a sharps container✓ Answer
- C.Snip the needle off the tubing with scissors
- D.Pull the needle off and put it in the trash bag
The OSHA Bloodborne Pathogens Standard bars bending, recapping or removing contaminated needles, and prohibits shearing or breaking them, because each of these brings the hand close to the point. Contaminated sharps must be discarded as soon as feasible in closable, puncture-resistant, leakproof, labeled containers, so the needle goes into the sharps container still attached to its tubing.
Ref QB-XQFGJASource: OSHA 29 CFR 1910.1030(d)(2)(vii) and (d)(4)(iii)(A)(1)Report a problem with this question
24. A technician has just washed up, and her hands are still wet when she needs to unplug a pump from an energized outlet at the charging station. Under OSHA's electrical safe-work rule, what should she do?
- A.Pull the cord quickly so contact is brief
- B.Unplug it with wet hands, then dry the plug
- C.Switch off the pump, then unplug with wet hands
- D.Dry her hands fully before touching the plug✓ Answer
OSHA states that employees' hands may not be wet when plugging and unplugging flexible cords and cord-and-plug equipment if energized equipment is involved, because water gives current a path into the body. Turning off the pump does not de-energize the outlet, and yanking the cord adds damage to the cord without removing the shock risk.
Ref QB-G8ZYACSource: OSHA 29 CFR 1910.334(a)(5)(i)Report a problem with this question
25. To move an infusion pump from a low shelf onto a delivery cart, a technician lifts it by its attached power cord. What does OSHA's electrical safe-work rule say about this?
- A.Cords may not be used to raise or lower equipment✓ Answer
- B.Cords may lift a unit once it is unplugged
- C.Units under 10 pounds may be carried by the cord
- D.A strain-relief boot makes cord lifting acceptable
OSHA requires portable equipment to be handled in a way that will not damage it and states that flexible cords connected to equipment may not be used for raising or lowering the equipment. Hanging the unit's weight on the cord strains the conductors and the cord entry, which can cause the hidden internal damage that later becomes a shock hazard, whatever the unit weighs or whether it is plugged in.
Ref QB-YQTTQBSource: OSHA 29 CFR 1910.334(a)(1) HandlingReport a problem with this question
26. An older bedside monitor arrives for processing, and neither the manufacturer nor the department can supply cleaning instructions for it. What does CDC's environmental infection control guideline advise for its surfaces?
- A.Low-temperature sterilization cycle for the whole monitor housing
- B.Detergent/disinfectant clean, then hospital disinfectant per label✓ Answer
- C.High-level disinfectant soak, then a sterile-water rinse of housing
- D.Plain tap-water wipe only, so the plastic housing is not damaged
CDC says to follow the manufacturer's instructions for noncritical medical equipment, and, when none are available, to clean the surfaces with a detergent/disinfectant, which may be followed by an EPA-registered hospital disinfectant used according to its label. A monitor touches only intact skin, so it is noncritical and does not call for sterilization or high-level disinfection, and soaking it would push fluid into the electronics; plain water removes soil but does not disinfect.
Ref QB-H8CBGDSource: CDC Guidelines for Environmental Infection Control in Health-Care Facilities (2003, updated 2019), Environmental Services Recommendations C and D.1Report a problem with this question
27. A technician plans to disinfect a stretcher's entire mattress and frame with 70% isopropyl alcohol from a squeeze bottle. What does CDC's environmental infection control guideline say about this?
- A.Alcohol is fine on any surface if it stays wet 1 minute
- B.Alcohol must be diluted to 35% for use on large surfaces
- C.Alcohol is required because it is a high-level disinfectant
- D.Alcohol should not be used to disinfect large surfaces✓ Answer
CDC's environmental guideline says not to use alcohol to disinfect large environmental surfaces. CDC's disinfection guideline explains why: alcohols evaporate rapidly, so the needed contact time is hard to reach unless an item is immersed, and repeated use swells and hardens rubber and some plastics. A large noncritical surface such as a stretcher is instead disinfected with an EPA-registered hospital disinfectant used per its label.
Ref QB-ZQ66WCSource: CDC Guidelines for Environmental Infection Control in Health-Care Facilities (2003, updated 2019), Environmental Services Recommendation D.2; CDC Guideline for Disinfection and Sterilization (2008), Alcohol sectionReport a problem with this question
Other topics in this exam
Practice questions based on the HSPA CRCST Certification Exam Content Outline, ANSI/AAMI ST79 sterility-assurance practice, and CDC/HICPAC disinfection and sterilization guidance. CRCST is a mark of the Healthcare Sterile Processing Association; this site is not affiliated with or endorsed by HSPA. Always follow the device manufacturer's written instructions for use, your facility's policies, and the current edition of the applicable standards. About the CRCST exam →