34 Sterilization Methods Practice Questions & Answers
Every Sterilization Methods practice question from the Sterile Processing (CRCST) Practice Test, with the correct answer and a short explanation.
Start practice test →1. Saturated steam under pressure is used to sterilize surgical instrument sets. By what mechanism does moist heat actually destroy microorganisms, including spores?
- A.Pressure inside the chamber physically ruptures the cell wall, so the organism collapses and cannot recover.
- B.Moisture-laden heat coagulates and denatures the cell's proteins and enzymes, ending all metabolic activity.✓ Answer
- C.Heated vapor draws the water out of the cell, and the resulting desiccation leaves the spore unable to germinate.
- D.Steam chemically alkylates the organism's DNA, which blocks replication and leaves the cell unable to divide.
Moist heat sterilizes by hydrating and then coagulating and denaturing cellular proteins and enzymes, which is why saturated steam, not hot air, is required. Pressure only raises the temperature at which the steam condenses and does not kill on its own; dry heat kills by oxidation and ethylene oxide by alkylation.
Ref QB-VSBM2FSource: ANSI/AAMI ST79, principles of moist-heat steam sterilization; HSPA CRCST body of knowledge, Sterilization ProcessReport a problem with this question
2. A wrapped instrument tray is opened in the operating room. The tape on the outside has changed, but the chemical indicator placed deep inside the tray has not. Which explanation best accounts for this?
- A.External tape reacts at a lower temperature than an internal indicator, so a gap between them is expected.
- B.A pocket of air stayed trapped in the center of the set, so steam never contacted that surface at cycle temperature.✓ Answer
- C.Chamber pressure ran above the setting for the cycle, which pushed the internal indicator past its color endpoint.
- D.The tray left the chamber before drying had finished, and the residual moisture bleached the indicator's ink.
Air is heavier than steam and does not transfer heat the way condensing steam does, so a trapped pocket stays cooler and keeps steam off the instruments inside it. Trapped air is the most common cause of steam sterilization failure, and the set must be treated as unsterile and reprocessed; the external tape shows only that the package was exposed to a process.
Ref QB-EAV4TGSource: ANSI/AAMI ST79, air removal and steam penetration; CDC Guideline for Disinfection and Sterilization, steam sterilizationReport a problem with this question
3. The instructions for a rigid, nonporous instrument allow either a gravity-displacement or a dynamic-air-removal steam cycle. How do the two designs differ in the way air leaves the chamber?
- A.In gravity, a pump evacuates the chamber first; dynamic cycles let the air rise and vent through a port in the door.
- B.In gravity, repeated water sprays dilute the air; dynamic cycles heat the air until the steam supply absorbs it.
- C.In gravity, the jacket absorbs chamber air before steam enters; dynamic cycles add gravity plus a longer exposure.
- D.In gravity, entering steam pushes the heavier air down and out the drain; dynamic cycles remove air mechanically.✓ Answer
Gravity displacement is passive: steam is lighter than air, so it enters at the top and pushes air out through the bottom drain. Because that removal is slower and less complete, gravity cycles need longer exposure and suit nonporous, non-lumened items, while dynamic air removal actively pulls or flushes air out and is used for porous and lumened loads.
Ref QB-WSPX7GSource: ANSI/AAMI ST79, sterilizer cycle types: gravity-displacement and dynamic air removalReport a problem with this question
4. A device's instructions for use specify a dynamic-air-removal steam cycle. Which sterilizer design satisfies that instruction?
- A.A gravity-displacement sterilizer at the same exposure temperature, because the drain still carries chamber air away.
- B.A steam-flush pressure-pulse sterilizer, which removes air with repeated flushes and pulses above atmospheric.✓ Answer
- C.A tabletop gravity sterilizer set to its longest wrapped-goods cycle, because added exposure offsets residual air.
- D.Any steam sterilizer at all, because dynamic air removal describes the drying phase rather than the air-removal method.
Dynamic air removal covers both prevacuum and steam-flush pressure-pulse designs, so either satisfies an instruction that names it. The pressure-pulse design keeps chamber pressure above atmospheric instead of drawing a vacuum, so air cannot be sucked back in through a leak, and gravity displacement is not interchangeable with either one.
Ref QB-FA8FYHSource: ANSI/AAMI ST79, dynamic air removal cycles (prevacuum and steam-flush pressure-pulse)Report a problem with this question
5. A daily air-removal test pack of the Bowie-Dick type is run in a prevacuum sterilizer and passes. What does that result establish?
- A.That the exposure phase reached and held the temperature needed to kill the resistant spores on the sheet.
- B.That the air was removed well enough for steam to penetrate quickly, with no leak or non-condensable gas present.✓ Answer
- C.That the sterilizer met every critical parameter for the load that follows, so no biological challenge is needed today.
- D.That the chamber drain, the door gasket and the jacket passed the preventive maintenance scheduled for that week.
The Bowie-Dick style test is a chemical indicator designed for one specific test: air removal and rapid steam penetration. It detects leaks, vacuum pump problems and non-condensable gases, contains no spores, says nothing about lethality, and never substitutes for biological monitoring.
Ref QB-XS2ADHSource: ANSI/AAMI ST79, Bowie-Dick (air-removal) testing; ISO 11140 indicators for a specific testReport a problem with this question
6. Under what conditions is the daily air-removal test of a dynamic-air-removal sterilizer performed?
- A.In a fully loaded chamber at the end of the day, with the pack on the top shelf and away from the chamber door.
- B.In an empty cold chamber at any convenient hour, with the pack placed at the geometric center of the chamber.
- C.In an empty, prewarmed chamber before the first processed load, with the pack on the bottom shelf over the drain.✓ Answer
- D.In the first patient-item load of the day, with the pack tucked inside the densest wrapped tray in that load.
The chamber is empty so the test pack is the only challenge, and a warm-up cycle is run first so that a cold chamber does not itself generate condensate and a false failure. The pack sits over the drain because that is the coolest, hardest place for steam to reach, and the test is run each day the sterilizer is used, before the first processed load.
Ref QB-GAKS5KSource: ANSI/AAMI ST79, daily air-removal (Bowie-Dick) test procedureReport a problem with this question
7. The morning air-removal test sheet shows a pale, unchanged patch in the center while the edges changed uniformly. What is the correct next step?
- A.Rerun the same pack through a second cycle, since the greater total exposure will finish the change and confirm it.
- B.Release the sterilizer for use, since the edges responded and a pale central area is normal in a folded test pack.
- C.Run the day's loads but add a second internal indicator to every tray, and report the pale center at the next service.
- D.Repeat the test in an empty chamber; if the second pack also fails, hold the sterilizer out of service for repair.✓ Answer
A lighter or unchanged center is the classic pattern of air trapped in the pack, caused by a leak, a vacuum pump fault or non-condensable gases in the steam. The test is repeated once to confirm, and a second failure means the sterilizer may not process patient items until the cause is found and corrected; extra indicators or a longer run do not fix air removal.
Ref QB-YSDMJJSource: ANSI/AAMI ST79, interpretation of air-removal test results and removal of a sterilizer from serviceReport a problem with this question
8. Physical, chemical and biological monitors are all applied to a steam load. Which statement describes what they can establish?
- A.Only the biological indicator gives direct evidence of lethality; physical and chemical monitors show conditions.✓ Answer
- B.Only the chemical indicator gives direct evidence of lethality, since its dye reacts to time, heat and moisture at once.
- C.Only the printout gives direct evidence of lethality; indicators and spore tests are secondary checks on packaging.
- D.All three give equally direct evidence of lethality, so a single passing result of any kind releases the load for use.
Physical monitors report what the chamber actually did, and chemical indicators show that specified conditions were met where the indicator sat, but neither challenges the process with living organisms. The biological indicator carries a known population of highly resistant spores, so it is the only device that directly measures the lethality of the cycle.
Ref QB-HAX4BMSource: ANSI/AAMI ST79, physical, chemical and biological monitoring of sterilization processesReport a problem with this question
9. A supervisor proposes replacing the routine biological indicator challenge with an integrating (Type 5) chemical indicator, arguing that it responds to all the critical variables. Why is that unacceptable?
- A.The integrator responds only to temperature, so it cannot confirm that steam and pressure were present in the chamber.
- B.The integrator is a single-use device, and its result cannot be filed with the cycle records a recall investigation needs.
- C.The integrator is calibrated to one named cycle only, so it would have to be replaced whenever the load contents change.
- D.The integrator carries no living spores, so it can never show that a resistant microbial population was killed.✓ Answer
An integrating indicator is calibrated against the performance of a biological indicator, but it is still a chemical reaction and contains no living organisms, so it cannot demonstrate spore kill. Biological monitoring at the required intervals, and in implant loads, cannot be satisfied by any chemical indicator.
Ref QB-ZSRXQKSource: ANSI/AAMI ST79, biological monitoring requirements; ISO 11140 integrating indicatorsReport a problem with this question
10. A technician places an emulating (Type 6) indicator labeled for one specific exposure cycle into a tray that will be run in a different cycle. Why is that wrong?
- A.An emulating indicator reacts to the variables of the one cycle it is labeled for, so its reading means nothing here.✓ Answer
- B.An emulating indicator has to be paired with a spore strip in the same tray, and without that pairing it cannot be read.
- C.An emulating indicator may be used only outside a package, so it cannot be placed inside a tray in any cycle at all.
- D.An emulating indicator outranks an integrator, so using one in a routine tray wastes the highest grade made anywhere.
An emulating indicator is a cycle-verification device: it is designed and labeled for the stated variables of one named cycle, so in a different cycle its endpoint no longer corresponds to anything meaningful. The numbering of indicator types describes what an indicator responds to and is not a quality ranking.
Ref QB-DRJ8TVSource: ISO 11140 emulating (cycle-verification) indicators; ANSI/AAMI ST79, selection and use of chemical indicatorsReport a problem with this question
11. A technician incubates the day's processed biological indicator together with an unprocessed control from the same lot. The processed indicator shows no growth, and the control also shows no growth. How is this interpreted?
- A.The run fails: no growth in the control means the sterilizer contaminated that vial, so the load is recalled.
- B.The run passes: no growth in either vial confirms the spores were killed and that the incubator was right.
- C.The run is invalid: with no growth in the control, neither the spores nor the incubator can be trusted at all.✓ Answer
- D.The run is unresolved for implants only: other items are released and the control is repeated the next day.
The unprocessed control from the same lot proves two things at once: that the spore population was viable and that the incubator was working. If the control does not grow, nothing can be concluded from the processed indicator, so acceptance requires the processed indicator negative and the control positive.
Ref QB-W8RDDWSource: ANSI/AAMI ST79, biological indicator controls and interpretation of resultsReport a problem with this question
12. A full-size dynamic-air-removal sterilizer is monitored with a process challenge device holding a biological indicator. Where and under what chamber condition is that device placed?
- A.In the part of a fully loaded chamber that steam reaches last, low and toward the front near the drain.✓ Answer
- B.On the top shelf of an empty chamber, so that nothing else takes up steam and the device gets the full flow.
- C.Inside the heaviest tray of the load and wrapped with the instruments, so it shares that tray's exact conditions.
- D.Anywhere in the chamber provided the load is light, since the device itself is the challenge rather than its position.
A process challenge device has to present a challenge at least equal to the most difficult item routinely processed, and it only does that where the sterilant arrives last. For a full-size chamber that means a chamber loaded as it is for routine work, with the device low and near the door over the drain.
Ref QB-ERWJYWSource: ANSI/AAMI ST79, process challenge devices and placement for routine biological monitoringReport a problem with this question
13. A steam sterilizer has just undergone a major repair to its vacuum system. What must happen before it processes items for patient use again?
- A.It is requalified with challenge cycles monitored by a biological indicator, and stays out of service until they pass.✓ Answer
- B.It is returned to use and waits for the next scheduled spore test, with earlier loads simply noted in the cycle log.
- C.It is released once the service report is filed, since the repair company has already verified the unit's function.
- D.It is run through one empty warm-up cycle, and the first load is released on the strength of its chemical indicators.
Qualification testing with biological indicator challenge devices, plus air-removal testing on a dynamic-air-removal unit, is required after installation, relocation, a malfunction, a major repair, a process failure or a change in utilities. Consecutive passing cycles are needed, and the sterilizer may not process patient items until they pass; a vendor's report does not release it.
Ref QB-X84RJXSource: ANSI/AAMI ST79, qualification testing after installation, relocation, malfunction or major repairReport a problem with this question
14. An implant was processed in a load monitored with a process challenge device holding both a biological indicator and an integrating indicator. The biological indicator is still incubating and the implant is needed for a documented emergency. What does the standard allow?
- A.Release on the passing integrator alone and discard the biological indicator, since the load has gone to the operating room.
- B.Release on the passing integrator and physical monitors, record the early release on an exception form, and still read it.✓ Answer
- C.Release the implant and log it as a routine issue, because an emergency case adds no further documentation duty at all.
- D.Release the implant only once the biological indicator has been read out, even if the procedure has to wait until then.
Implant loads are quarantined until the biological indicator has been read, and that remains the default. Only in a documented emergency may the implant be released early on the strength of the integrating indicator and the cycle record, and even then the biological indicator is still incubated, the result documented, and the early release recorded on an exception form with follow-up.
Ref QB-FR9W5YSource: ANSI/AAMI ST79, monitoring of implant loads and documented emergency releaseReport a problem with this question
15. A department owns a single set of a frequently needed instrument and runs it through an immediate-use steam cycle between cases every day. How should this practice be judged?
- A.It is acceptable because the set is cleaned and decontaminated the same way before each immediate-use cycle is run.
- B.It is a misuse of immediate-use sterilization, which exists for an urgent item and not to cover an inventory shortage.✓ Answer
- C.It is acceptable as long as every cycle is logged and the set travels to the operating room in a closed container.
- D.It is acceptable because an immediate-use cycle reaches the same exposure temperature as a terminal wrapped cycle.
Immediate-use steam sterilization is reserved for a situation where an item is needed right away and no sterile replacement is available. Using it day after day to work around a permanent shortage of instruments is exactly the convenience use the standard prohibits, and correct cleaning or good record keeping does not make it acceptable.
Ref QB-Y8F3RYSource: ANSI/AAMI ST79, immediate-use steam sterilization; AORN Guideline for SterilizationReport a problem with this question
16. What must the record of an immediate-use steam cycle make possible?
- A.Tracing the load contents and cycle data to the individual patient on whom the items were used, and to the operator.✓ Answer
- B.Showing that the items were placed into sterile storage and were rotated with the rest of the department's stock.
- C.Proving that the set carried an expiration date and a shelf-life statement before it went to the operating room.
- D.Confirming that the set was wrapped in a validated barrier so it could be held for the next case on that day's list.
An immediate-use item is transported aseptically and used at once rather than stored, so there is no quarantine period and no shelf life to record. The record therefore has to support traceability to the patient, along with the load contents, the cycle parameters, the operator and the reason the cycle was run.
Ref QB-GRM8BZSource: ANSI/AAMI ST79, immediate-use steam sterilization documentation and traceabilityReport a problem with this question
17. A mixed steam load holds wrapped metal instrument trays, textile packs, basins and peel pouches. How should the chamber be loaded?
- A.Textile packs low, metal trays above them, basins flat on the shelf, and pouches laid flat and stacked to save room.
- B.Everything packed tightly so nothing shifts, with the pouches placed inside the wrapped trays so they cannot slide.
- C.Metal trays low, textiles above them, basins tilted on edge and pouches on edge, with space between every item.✓ Answer
- D.Metal trays and textiles side by side and touching on one shelf, with basins upright so they can hold the condensate.
Condensate runs downward, so metal placed above porous goods drips onto them and soaks the barrier. Basins and solid-bottom containers are tilted on edge so water runs out, pouches stand on edge so steam enters and the paper can dry, and space between items lets steam circulate and reach every surface.
Ref QB-Z8TDWZSource: ANSI/AAMI ST79, sterilizer loading and load configurationReport a problem with this question
18. Wrapped trays come out damp from one steam sterilizer, while the same trays run dry in another sterilizer in the same department. Which cause best explains this?
- A.That chamber runs hotter than the cycle calls for, and the superheated steam lays a film of water on the wrapper.
- B.The internal indicators in the trays take up steam during exposure and then give the moisture back during drying.
- C.The trays are cooling too slowly on the cart, so the warm air around them keeps the wrappers from drying off.
- D.The steam reaching that chamber carries too much water, so condensate forms faster than the drying phase removes it.✓ Answer
Steam that carries excess water, typically from boiler carryover or a supply problem serving that one machine, deposits more condensate than the drying phase can drive off, which is a classic wet-pack cause. A wet pack is considered contaminated: it cannot be dried, re-taped or used, and the contents must be fully reprocessed.
Ref QB-HRYJG2Source: ANSI/AAMI ST79, steam quality and causes of wet packsReport a problem with this question
19. A staging rack used for cooling sits directly under a ceiling air diffuser, and packages are unloaded onto it straight from the chamber. What is the problem?
- A.Moving air dries the wrappers faster than the sterilizer's own drying phase, which makes the barrier brittle and easy to tear.
- B.The diffuser keeps the packages warm for longer, so the indicators inside them go on reacting and overshoot the endpoint.
- C.Cold air blowing onto hot packages drives condensation inside the wrapper, and that moisture carries contamination in.✓ Answer
- D.Air from a diffuser is unfiltered, so airborne dust settles on the wrappers and has to be wiped off before they are stored.
Cooling a hot package too quickly makes water condense on and within the barrier, and a damp wrapper wicks organisms from the outside to the contents. Loads are cooled undisturbed on a rack away from vents, drafts and cold surfaces, are never set hot onto a cold solid counter, and are not handled or transferred until they are cool.
Ref QB-296R33Source: ANSI/AAMI ST79, cooling and post-sterilization handling of sterilized itemsReport a problem with this question
20. A heat-sensitive device is processed with ethylene oxide. Why must the aeration phase be completed before the item is handled or issued?
- A.Aeration allows the external indicator to finish changing color, and the load is not released before that change is read.
- B.Aeration is the phase in which the gas finishes killing the spores, so an early removal leaves the process incomplete.
- C.Aeration restores the humidity the device lost inside the chamber, without which its plastic parts grow brittle and crack.
- D.Aeration lets the gas taken up by the device and its packaging dissipate, since that residue harms patients and staff.✓ Answer
Ethylene oxide is absorbed by polymers and packaging materials, and residual gas is toxic, so aeration is a mandatory phase that drives the residue off before anyone handles the item. Sterilization itself occurs by alkylation during the exposure phase, and it is the long aeration requirement that gives ethylene oxide the longest turnaround of the low-temperature methods.
Ref QB-M8BFCTSource: ANSI/AAMI ST41, ethylene oxide sterilization and aeration; OSHA ethylene oxide standard, 29 CFR 1910.1047Report a problem with this question
21. A rigid device with a long, narrow lumen is being considered for a hydrogen peroxide cycle. What determines whether it may be processed that way?
- A.Whether the lumen can be flushed with sterile water first, because a wetted channel helps carry vapor along its length.
- B.Whether a longer exposure can be selected, since added time will move the vapor through any channel a device may have.
- C.Whether the device and sterilizer instructions accept a lumen of that bore and length, since the vapor must reach its end.✓ Answer
- D.Whether the device is metal, because a metal channel conducts the vapor while a polymer channel absorbs and stops it.
Hydrogen peroxide sterilizes by diffusing to every surface, and a long narrow channel limits how far the vapor penetrates, so the lumen claims validated in the device and sterilizer instructions decide the question. The item must also be completely dry, because residual moisture will cancel the cycle.
Ref QB-4R5ASSSource: ANSI/AAMI ST58, low-temperature chemical sterilization; device and sterilizer instructions for use, lumen claimsReport a problem with this question
22. A procedure tray needs a sterile jar of petroleum-based lubricant and a packet of powder. Steam cannot penetrate either item. Which method and typical cycle fit these items?
- A.Prevacuum steam at 132°C for 4 minutes
- B.Ethylene oxide at 55°C, then 12 h aeration
- C.Dry heat (hot air) at 170°C for 60 minutes✓ Answer
- D.Gravity steam at 121°C for 30 minutes
CDC limits dry heat to materials that moist heat would damage or cannot penetrate, and it names powders and petroleum products as examples. Its most common hot-air time-temperature pairs are 170°C for 60 minutes, 160°C for 120 minutes and 150°C for 150 minutes. Any steam cycle fails because steam cannot reach the inside of oils or powders, whatever the exposure time. Ethylene oxide is reserved for heat- and moisture-sensitive devices, and these items tolerate heat.
Ref QB-N8PSHUSource: CDC Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, updated 2024), Dry-Heat Sterilizers section (p. 69-70): powders and petroleum products; 170°C/60 min, 160°C/120 min, 150°C/150 minReport a problem with this question
23. A department installs its first ethylene oxide sterilizer and must stock biological indicators for it. Which organism and incubation temperature apply to EO monitoring?
- A.B. atrophaeus spores, incubated at 35–37°C✓ Answer
- B.G. stearothermophilus spores, incubated at 55–60°C
- C.B. atrophaeus spores, incubated at 55–60°C
- D.G. stearothermophilus spores, incubated at 35–37°C
Each indicator organism is chosen because it is especially resistant to one sterilizing agent. Bacillus atrophaeus spores monitor ethylene oxide and dry heat, while Geobacillus stearothermophilus spores monitor steam, hydrogen peroxide gas plasma and liquid peracetic acid. The BI must also be incubated at the temperature that organism grows at: 35–37°C for B. atrophaeus and 55–60°C for G. stearothermophilus. At the wrong temperature, surviving spores may not grow, and a failed cycle would look like a pass.
Ref QB-5RGMXTSource: CDC Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, updated 2024), Monitoring section (p. 77): B. atrophaeus for ETO and dry heat, incubated 35-37°C; G. stearothermophilus for steam, H2O2 gas plasma and peracetic acid, incubated 55-60°CReport a problem with this question
24. An industrial hygienist reviews the monitoring badges of staff who unload the ethylene oxide sterilizer and aerator. Which pair of OSHA exposure limits applies to ethylene oxide?
- A.PEL 5 ppm (8-h TWA); excursion 1 ppm over 15 min
- B.PEL 0.5 ppm (8-h TWA); excursion 10 ppm over 15 min
- C.PEL 1 ppm (8-h TWA); excursion 10 ppm over 30 min
- D.PEL 1 ppm (8-h TWA); excursion 5 ppm over 15 min✓ Answer
The OSHA ethylene oxide standard sets the permissible exposure limit at 1 ppm as an 8-hour time-weighted average. It also sets an excursion limit of 5 ppm averaged over a 15-minute sampling period, which catches short spikes such as opening a chamber or aerator. The 0.5 ppm figure is the action level, which triggers monitoring and other duties; it is not the PEL. Badges track each worker's exposure against both limits.
Ref QB-P825QVSource: OSHA 29 CFR 1910.1047(c)(1) permissible exposure limit 1 ppm 8-h TWA and 1910.1047(c)(2) excursion limit 5 ppm averaged over 15 minutesReport a problem with this question
25. A heat- and moisture-sensitive camera head will be processed in a hydrogen peroxide gas plasma sterilizer, and the device IFU allows that cycle. Which packaging should the technician choose?
- A.A paper-plastic peel pouch, heat-sealed
- B.Polypropylene wrap or a polyolefin pouch✓ Answer
- C.A double layer of cotton muslin (linen) wrap
- D.Crepe paper wrap secured with indicator tape
Hydrogen peroxide gas plasma cannot process cellulose (paper), linens or liquids, and it requires synthetic packaging such as polypropylene wrap or polyolefin pouches. Cellulose absorbs the hydrogen peroxide, which can abort the cycle and keep the sterilant from reaching the device. A paper-plastic pouch, muslin and crepe paper all contain cellulose, so each is acceptable for steam but not for this modality.
Ref QB-6RUX4VSource: CDC Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, updated 2024), Table 6 (p. 118), hydrogen peroxide gas plasma: cellulose (paper), linens and liquids cannot be processed; requires synthetic packaging (polypropylene wraps, polyolefin pouches)Report a problem with this question
26. This morning's routine steam BI reads positive, and the unprocessed control from the same lot grew as expected. The load's printout and chemical indicators looked normal. What should the technician do first?
- A.Keep using the sterilizer and recheck it with tomorrow's routine BI
- B.Release the load, since its printout and chemical indicators passed
- C.Stop using the sterilizer and notify the supervisor and infection control✓ Answer
- D.Recall every item run on that sterilizer during the last 30 days
Because the control grew, the positive result is valid, and the CDC protocol for a positive steam BI starts with taking the sterilizer out of service and notifying the supervisor and infection control. The BI is then repeated in three consecutive cycles, and loads are recalled if the sterilizer or procedure proves defective; implant loads are handled more strictly. Normal printouts and chemical indicators do not override a valid positive BI. A blanket 30-day recall is not the first step.
Ref QB-Q8DFVWSource: CDC Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, updated 2024), Figure 2 protocol for management of positive biological indicator in a steam sterilizer (p. 124), step 1: take the sterilizer out of service, notify area supervisor and infection controlReport a problem with this question
27. A heat-sensitive rigid scope was processed unwrapped in a liquid peracetic acid immersion system. What should happen to it after the cycle?
- A.Wrap it and place it on the sterile storage shelf
- B.Take it to the point of use and use it immediately✓ Answer
- C.Air-dry it overnight, then package it for storage
- D.Store it in a closed bin under event-related shelf life
CDC states that critical items sterilized by the peracetic acid immersion process must be used immediately. They come out wet and unwrapped, so they are not fully protected from contamination, and long-term storage is unacceptable. Wrapping or drying them afterward cannot restore a barrier that was never there during sterilization. Event-related shelf life applies only to items packaged before they are sterilized.
Ref QB-7R7AAWSource: CDC Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, updated 2024), Recommendation (Sterilization) g, p. 93: items sterilized by the peracetic acid immersion process must be used immediately; long-term storage unacceptable (Category II)Report a problem with this question
28. While reviewing an ethylene oxide cycle record, a lead technician confirms that the process's four essential parameters stayed within range. Which set names those four parameters?
- A.Gas concentration, temperature, chamber pressure, and drying time
- B.Steam quality, temperature, relative humidity, and exposure time
- C.Gas concentration, vacuum depth, chamber pressure, and exposure time
- D.Gas concentration, temperature, relative humidity, and exposure time✓ Answer
CDC lists the four essential operating parameters of ethylene oxide sterilization as gas concentration (450 to 1200 mg/L), temperature (37 to 63°C), relative humidity (40 to 80%) and exposure time (1 to 6 hours). Humidity is essential because water molecules carry EO to its reactive sites on the organism. Steam quality, vacuum depth and drying time are steam-cycle concepts and are not among EO's essential parameters.
Ref QB-G8W4PNSource: CDC Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, updated 2024), Ethylene Oxide Gas Sterilization overview (p. 63): four essential parameters are gas concentration, temperature, relative humidity and exposure timeReport a problem with this question
29. A technician is labeling wrapped trays as they are placed on the sterilizer cart. According to CDC guidance, what must the load control label identify?
- A.The sterilizer used, the cycle or load number, and the date of sterilization✓ Answer
- B.The patient's name, the surgeon, and the date of the scheduled procedure
- C.The packer's initials, the tray weight, and the storage shelf location
- D.The BI lot number, the incubation result, and the date the load was released
CDC recommends labeling sterilized items with a load number that shows the sterilizer used, the cycle or load number and the date of sterilization, plus an expiration date if one applies. Those three facts let the department find and recall exactly the items from one load if a monitor later fails, and trace them to where they were used. BI lot numbers and results belong in the load record, not on each package.
Ref QB-YQQX3NSource: CDC Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, updated 2024), Recommendation (Storage of Sterile Items) c, p. 94: label sterilized items with a load number indicating the sterilizer used, the cycle or load number, the date of sterilization, and, if applicable, the expiration date (Category IB)Report a problem with this question
30. A tech runs the daily Bowie-Dick test in an empty prevacuum steam sterilizer. The test sheet shows darker edges with a pale, unchanged area in the center. What does this result indicate?
- A.Exposure time was too long
- B.A normal passing result
- C.Inadequate air removal✓ Answer
- D.Spores survived the cycle
The Bowie-Dick test detects air leaks and inadequate air removal: trapped air keeps steam from reaching the center of the test sheet, so a pass requires a uniform color change across the whole sheet. A pale center is a failure, and the sterilizer must not be used until it is inspected and passes the test. The Bowie-Dick test does not measure spore kill; that is the job of a biological indicator.
Ref QB-H6BAUPSource: CDC Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, updated June 2024), Steam Sterilization section, p. 60: Bowie-Dick test detects air leaks and inadequate air removal; do not use the sterilizer until it passesReport a problem with this question
31. A heat-sensitive device is staged for a hydrogen peroxide gas plasma cycle, but it is packaged in a paper-plastic peel pouch. What should the tech do?
- A.Repackage it in a Tyvek pouch✓ Answer
- B.Run it on an extended cycle
- C.Add an extra internal indicator
- D.Place it on the bottom shelf
Cellulose materials such as paper and linen cannot be processed in hydrogen peroxide gas plasma because they absorb the peroxide and can abort the cycle or prevent sterilization. A longer cycle, an extra indicator or a different shelf does not fix incompatible packaging; the item must be repackaged in a compatible material such as a Tyvek-plastic pouch or polypropylene wrap.
Ref QB-2PGFEQSource: CDC Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, updated June 2024), p. 75 (linens and paper unacceptable packaging for hydrogen peroxide gas plasma) and Table 6, p. 118 (cellulose/paper, linens and liquids cannot be processed)Report a problem with this question
32. A steam load containing an orthopedic implant has finished, and its biological indicator is still incubating. There is no documented emergency. What should happen to the implant?
- A.Release it once the internal CI passes
- B.Release it once the printout is verified
- C.Release it after it cools for 30 minutes
- D.Release it once the BI reads negative✓ Answer
Every load containing implantable items must be monitored with a biological indicator, and the implant should not be used until the spore test is known to be negative. Physical monitors and chemical indicators only show that cycle conditions were met; only the BI directly shows that the process killed resistant spores, so neither can substitute for it outside a documented emergency release.
Ref QB-G6YXNNSource: CDC Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, updated June 2024), Monitoring section, p. 78: each load with implantable objects should be monitored; implantable items should not be used until spore tests are known to be negativeReport a problem with this question
33. A prevacuum steam sterilizer has just had a major repair. Before it returns to routine use, which testing is required in an empty chamber?
- A.One BI cycle and one Bowie-Dick test
- B.Three BI cycles and three Bowie-Dick tests✓ Answer
- C.Three Bowie-Dick tests and no BI cycles
- D.Three BI cycles and no Bowie-Dick tests
After installation, relocation, major repair or a malfunction, the sterilizer is qualified with three consecutive cycles containing biological indicators, and a prevacuum unit also runs three consecutive empty cycles with a Bowie-Dick test. The two tests check different things: BIs check lethality and Bowie-Dick checks air removal, so a prevac unit needs both before it goes back into service.
Ref QB-ZN559PSource: CDC Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, updated June 2024), Sterilizer qualification testing, p. 74: three consecutive BI cycles; in a prevacuum steam sterilizer three consecutive empty cycles with a Bowie-Dick testReport a problem with this question
34. Badge monitoring is used for staff who work near ethylene oxide (EO) sterilizers. What is the OSHA short-term excursion limit for EO exposure?
- A.1 ppm averaged over 15 minutes
- B.0.5 ppm averaged over 15 minutes
- C.5 ppm averaged over 15 minutes✓ Answer
- D.5 ppm averaged over 8 hours
OSHA sets two EO limits on different time bases: the permissible exposure limit is 1 ppm as an 8-hour time-weighted average, and the excursion limit caps short spikes at 5 ppm averaged over 15 minutes. The 0.5 ppm figure is the action level that triggers monitoring, not an exposure limit, so the distractors mix up these three numbers.
Ref QB-F6MMGMSource: OSHA 29 CFR 1910.1047(c)(2) Ethylene Oxide, Excursion limit: no employee exposed above 5 ppm averaged over a 15-minute sampling periodReport a problem with this question
Other topics in this exam
Practice questions based on the HSPA CRCST Certification Exam Content Outline, ANSI/AAMI ST79 sterility-assurance practice, and CDC/HICPAC disinfection and sterilization guidance. CRCST is a mark of the Healthcare Sterile Processing Association; this site is not affiliated with or endorsed by HSPA. Always follow the device manufacturer's written instructions for use, your facility's policies, and the current edition of the applicable standards. About the CRCST exam →