← Infection Control

Chapter 2 · Lesson 3 of 4

Instrument processing and sterilization monitoring

Varies by state. Your state sets this exam: the number of questions, time limit, pass mark and any state-law section can differ from this practice. Check your state's official handbook or licensing board before you test.

1 min read

The Spaulding classification sets the minimum processing for each item by what it touches. Critical items penetrate soft tissue or bone and must be sterilized. Semicritical items touch mucous membranes without penetrating them, and any that can withstand heat are heat sterilized. Noncritical items touch only intact skin.

Example: a scaler is critical, a mouth mirror is semicritical and goes in the sterilizer, and a blood pressure cuff is noncritical. Liquid chemical sterilants are a last resort for heat-sensitive items, because soaked items cannot be packaged, biologically monitored or stored sterile.

  • Ultrasonic cleaner: open the hinges, submerge the instruments completely, and run it with the lid closed to contain aerosol.
  • Handpieces, low-speed motors and reusable prophy angles are cleaned, lubricated per the instructions and heat sterilized between patients, since fluids reach their internal channels.
  • Steam kills faster than dry heat because moist heat denatures protein quickly, while dry air carries heat poorly and needs a higher temperature and longer time.

Monitoring has three layers. Mechanical readings show time, temperature and pressure. Chemical indicators change color when exposed to conditions in the cycle, and an external one only proves the package went through the process.

Biological indicators, spore tests run at least weekly, are the only proof of a kill. After a positive spore test and a correction, the sterilizer must pass spore tests in three consecutive empty-chamber cycles before it is used again.

Check yourself

1. A pouch leaves the sterilizer with the printed stripe on its outside fully darkened, and a new employee says this shows the contents are sterile. What does that external marking actually establish?

2. Impression trays and mouth mirrors are heat tolerant and contact mucous membranes without penetrating tissue. The office asks whether soaking them in a liquid chemical sterilant is an acceptable shortcut. What is correct?

Facts checked 2026-11-15 · Sources: CDC: Guidelines for Infection Control in Dental Health-Care Settings, 2003, OSHA: Bloodborne Pathogens Standard, 29 CFR 1910.1030, DANB: Infection Control (ICE) exam