Quibank app: now on the App Store — free, works offline. Get the app ↗

50 Tonometry, Visual Fields & Biometry Practice Questions & Answers

Every Tonometry, Visual Fields & Biometry practice question from the Ophthalmic Assistant (COA) Practice Test, with the correct answer and a short explanation.

Start practice test →

Practice tests

  1. 1. Goldmann applanation tonometry estimates intraocular pressure from the force needed to flatten a standardized area of cornea. Why is that particular flattened area used?

    • A.At that area, the cornea no longer refracts any light from the slit beam
    • B.At that area, corneal rigidity and tear-film surface tension cancel out✓ Answer
    • C.At that area, the fluorescein ring is too small to touch the lid margin

    The Imbert-Fick relation treats pressure as force divided by flattened area, but the cornea is not an ideal membrane: its own rigidity resists flattening while the tear film's surface tension pulls the prism toward the eye. The standard applanation diameter is the size at which those two opposing forces cancel, so the force read off the dial reflects intraocular pressure alone.

    Ref QB-JMG5NSSource: Imbert-Fick principle as applied in Goldmann applanation tonometry; IJCAHPO Core Criteria Handbook, tonometry content areaReport a problem with this question

  2. 2. While measuring with the applanation tonometer, the two fluorescein semicircles look very wide and thick and overlap broadly. What has most likely happened, and how does it affect the reading?

    • A.Too little fluorescein thinned out the mires, biasing the reading high
    • B.Excess fluorescein or a wet lid thickened the mires, so it reads high✓ Answer
    • C.The cobalt filter is out of the beam, so the reading is not affected

    Each mire should be about a tenth of its own diameter in width. Excess fluorescein, tearing or a wet lid margin fattens the semicircles so their inner edges meet with more force applied than the eye truly requires, which reads high; too little fluorescein gives faint, thin mires and reads low. The fix is to dry the lids, reapply a minimal amount of fluorescein and remeasure.

    Ref QB-25BY2SSource: Goldmann applanation technique, mire width and fluorescein quantity; IJCAHPO Core Criteria Handbook, tonometry content areaReport a problem with this question

  3. 3. A patient has had laser refractive surgery and now has a cornea thinner than average. Compared with the true intraocular pressure, how does applanation tend to read?

    • A.Unchanged, because thickness does not affect applanation
    • B.Falsely high, because a thin cornea resists flattening more
    • C.Falsely low, because a thin cornea resists flattening less✓ Answer

    Applanation assumes a cornea of average thickness and rigidity. A thinner cornea is easier to flatten, so less force is needed to reach the endpoint and the instrument reports a pressure below the true one; a thicker cornea over-reads for the same reason reversed. What matters clinically is the direction of the bias, which the physician weighs against the measured thickness.

    Ref QB-HM5TFRSource: Effect of central corneal thickness on applanation tonometry; IJCAHPO Core Criteria Handbook, tonometry content areaReport a problem with this question

  4. 4. What is the correct endpoint for the fluorescein mires during applanation, and what should be done if they pulsate with the ocular pulse?

    • A.Inner edges just separated; if pulsating, press the lids to damp the pulse
    • B.Inner edges just touching; if pulsating, read the middle of the swing✓ Answer
    • C.Outer edges just touching; if pulsating, read the widest point of the swing

    The dial is turned until the inner edges of the upper and lower semicircles just meet, because that position corresponds to the standardized flattened area the instrument is built around. A pulsating mire reflects the ocular pulse, and the average pressure is represented by the midpoint of the excursion; pressing on the lids or globe to stop the movement would itself raise the pressure.

    Ref QB-55DYJVSource: Goldmann applanation endpoint and ocular pulse; IJCAHPO Core Criteria Handbook, tonometry content areaReport a problem with this question

  5. 5. A patient's pressure reads 31 mmHg in the right eye today, compared with 14 mmHg at the last visit. What should the assistant do next?

    • A.Repeat after checking posture, collar and lid handling, then report it✓ Answer
    • B.Instill more anesthetic, remeasure and chart the average of the two
    • C.Tell the patient that this level means glaucoma and book a field test

    A large unexplained jump is often artifactual: breath holding, a tight collar, squeezing the lids, or the examiner's fingers pressing the globe all raise the reading. The assistant confirms the value with a correctly performed repeat and then reports it, with the method, the eye and the time, to the ophthalmologist; telling the patient what the number means is a diagnosis and outside the assistant's role.

    Ref QB-MM7TYUSource: Artifactual elevation of applanation readings and reporting of findings; IJCAHPO Core Criteria Handbook, tonometry content areaReport a problem with this question

  6. 6. A patient arrives with a painful red eye and a corneal epithelial defect that stains with fluorescein. How should the pressure check be handled?

    • A.Proceed, but touch the prism lightly and take one reading only
    • B.Defer contact tonometry and let the physician decide how to measure✓ Answer
    • C.Switch to indentation tonometry, which does not touch the cornea at all

    An eye with an epithelial defect, a possible infectious keratitis or recent surgery should not have a prism placed on it: contact can enlarge the defect, seed infection, and readings over an irregular cornea are unreliable anyway. The assistant reports the finding and lets the physician decide whether and how to measure. Indentation tonometry also rests on the cornea, so it is no safer.

    Ref QB-452NDUSource: Contraindications to contact tonometry; IJCAHPO Core Criteria Handbook, tonometry content areaReport a problem with this question

  7. 7. After a reusable tonometer tip is disinfected, why must it be rinsed with water and dried before it touches the next patient's eye?

    • A.Water removes fluorescein that would otherwise thicken the mires
    • B.Disinfectant left on the tip can injure the corneal epithelium✓ Answer
    • C.Drying is the step that actually inactivates virus on the prism

    The agents used on tonometer tips are epithelial toxins: any film of bleach, peroxide or alcohol carried onto the cornea can produce a chemical keratitis and a painful abrasion. Disinfection kills the organisms, and the rinse-and-dry step removes the agent itself; disposable tips or covers avoid the problem altogether.

    Ref QB-KMUFTTSource: Tonometer tip disinfection and rinsing; IJCAHPO Core Criteria Handbook, tonometry and infection-control contentReport a problem with this question

  8. 8. An automated field shows a dense arcuate defect in the superior hemifield that stops sharply at the horizontal midline. What does respecting the horizontal midline point to?

    • A.Compression of the crossing fibers at the optic chiasm
    • B.A lesion of the optic radiations in the temporal lobe
    • C.Damage to nerve fiber bundles at the optic nerve head✓ Answer

    Retinal nerve fibers arc around the macula and meet along the temporal raphe, which lies on the horizontal meridian, so bundle damage at the disc produces arcuate defects and nasal steps that stop at the horizontal midline. Lesions at or behind the chiasm follow the vertical midline instead, because that is where the nasal and temporal halves of the retina divide.

    Ref QB-754NWXSource: Retinal nerve fibre anatomy and glaucomatous field defects; IJCAHPO Core Criteria Handbook, visual field testing content areaReport a problem with this question

  9. 9. A patient's fields show loss of the temporal half of each eye's field, with a sharp edge at the vertical midline. Where does this pattern localize?

    • A.The optic nerves on both sides, in front of the chiasm
    • B.The optic tract on one side, just behind the chiasm
    • C.The optic chiasm, where nasal retinal fibers cross✓ Answer

    Fibers from the nasal half of each retina carry the temporal half of that eye's field and cross in the chiasm, so a lesion pressing on the chiasm knocks out both temporal fields with an edge at the vertical midline. A defect behind the chiasm would be homonymous, affecting the same side of the field in both eyes rather than opposite sides.

    Ref QB-PMWFBXSource: Visual pathway anatomy and chiasmal field defects; IJCAHPO Core Criteria Handbook, visual field testing content areaReport a problem with this question

  10. 10. Why does automated static perimetry use the patient's distance correction together with an age-appropriate near add?

    • A.The add offsets the general depression caused by a cataract
    • B.Blur does not change thresholds, so the lens is for comfort
    • C.The bowl is at a near distance, so the eye must focus there✓ Answer

    The stimuli are projected on a bowl only a short distance from the eye, so a presbyopic patient cannot focus on that surface with a distance correction alone. Defocus blurs the stimulus and lowers measured sensitivity, producing a generalized depression that mimics disease, which is why the lens is chosen for the bowl distance, kept close to the eye and used in a full-aperture holder.

    Ref QB-3M264HSource: Refractive correction for automated perimetry; IJCAHPO Core Criteria Handbook, visual field testing content areaReport a problem with this question

  11. 11. A field printout shows a dense ring of absolute loss around the outermost tested points, with a normal central field. What is the most likely cause?

    • A.A very small pupil depressed sensitivity across the field
    • B.The rim of the trial lens blocked the peripheral stimuli✓ Answer
    • C.A drooping upper lid covered the pupil during the test

    A trial lens that is decentered or held too far from the eye casts the shadow of its rim on the peripheral test points, producing an absolute ring of loss with a normal center. Keeping a full-aperture lens centered on the pupil and as close to the lashes as possible prevents it; a ptotic lid gives superior depression, and a small pupil depresses the whole field.

    Ref QB-K48BPHSource: Trial lens rim artifact in automated perimetry; IJCAHPO Core Criteria Handbook, visual field testing content areaReport a problem with this question

  12. 12. A visual field returns a high false-positive rate. How does that distort the result?

    • A.Only the blind spot is affected, so the rest of the field stands
    • B.Sensitivities read abnormally high and real defects may be hidden✓ Answer
    • C.Sensitivities read abnormally low and mild defects look severe

    False positives are responses given when no stimulus was presented, so a trigger-happy patient is credited with seeing stimuli dimmer than they can truly detect. Thresholds are pushed upward, unrealistically bright islands can appear inside a damaged area, and the summary indices look better than the eye is, so the test should be repeated after reinstruction.

    Ref QB-2MPTWFSource: Reliability indices in automated perimetry; IJCAHPO Core Criteria Handbook, visual field testing content areaReport a problem with this question

  13. 13. A patient viewing an Amsler grid says the lines near the center look wavy and bent. What is being described, and how must the test be run?

    • A.Metamorphopsia; test one eye at a time with the near correction on✓ Answer
    • B.A central scotoma; test both eyes together at distance correction
    • C.Metamorphopsia; test both eyes together after instilling dilating drops

    Wavy, bowed or bulging lines are metamorphopsia, distortion that arises when macular photoreceptors are displaced by fluid or membrane. The grid maps only the central field, so it is held at the patient's reading distance with the near correction, one eye at a time with the fellow eye covered, while the patient keeps fixation on the central spot and reports missing or distorted areas.

    Ref QB-J4VYGGSource: Amsler grid technique and metamorphopsia; IJCAHPO Core Criteria Handbook, visual field testing content areaReport a problem with this question

  14. 14. Compared with immersion ultrasound, how does contact (applanation) A-scan biometry bias the measurement and the lens power calculated from it?

    • A.It shortens the axial length, so the calculated lens power comes out too strong✓ Answer
    • B.It lengthens the axial length, so the calculated power comes out too strong
    • C.It shortens the axial length, so the calculated lens power comes out too weak

    Resting the probe on the cornea indents it, so the sound path measured is shorter than the true axial length. In every lens power formula a shorter eye calls for a stronger lens, so an artificially short measurement yields an implant with too much power and a myopic result. Immersion technique keeps a fluid column between probe and cornea and avoids the compression altogether.

    Ref QB-ZKBGRESource: Contact versus immersion A-scan technique and axial length error; IJCAHPO Core Criteria Handbook, biometry content areaReport a problem with this question

  15. 15. Axial length measures 23.4 mm in one eye and 24.6 mm in the fellow eye, and nothing in the refraction explains the difference. What should the assistant do?

    • A.Pass the readings on for the surgeon to adjust
    • B.Average the two lengths and use that one figure
    • C.Remeasure both eyes before the figures are used✓ Answer

    Axial length is a major source of error in a lens power calculation, and an error of about 1 mm shifts the result by roughly 2.5 to 3 diopters. A large difference between two eyes that the refraction does not support usually means the probe was off the visual axis, fixation was poor, or the cornea was compressed, so both eyes are remeasured until the readings repeat consistently.

    Ref QB-H4HNBFSource: Axial length error and interocular comparison in biometry; IJCAHPO Core Criteria Handbook, biometry content areaReport a problem with this question

  16. 16. Before an A-scan the assistant selects the eye-type setting: phakic, aphakic, pseudophakic or silicone-filled. Why does this setting matter?

    • A.It supplies the lens constant given by the manufacturer
    • B.Sound travels at a different speed in each of those eyes✓ Answer
    • C.It tells the machine which lens power formula it should apply

    Ultrasound biometry times an echo and converts that time into a distance using an assumed velocity, and sound moves at different speeds through a natural lens, aqueous and vitreous alone, an implant material, or silicone oil. Choosing the wrong eye type applies the wrong velocity and returns an axial length that is systematically wrong, no matter how clean the echoes look. The lens constant and the power formula belong to the later implant calculation and are chosen for the implant being planned, not by the eye-type setting.

    Ref QB-YKY5KDSource: Sound velocity settings in A-scan biometry; IJCAHPO Core Criteria Handbook, biometry content areaReport a problem with this question

  17. 17. Optical biometry cannot capture an axial length in an eye with a very dense cataract. What is the appropriate next step?

    • A.Use the fellow eye's length in the calculation
    • B.Dilate and repeat, since pupil size is the limit
    • C.Measure that eye with immersion ultrasound✓ Answer

    Optical biometry works by sending light to the retina and detecting its return, so an opacity dense enough to block that light returns no reading. Ultrasound uses sound, which passes through the opacity, and the immersion technique is preferred because it measures without compressing the cornea; the fellow eye can never substitute for the eye being operated on.

    Ref QB-G46B6ESource: Optical versus ultrasound biometry and their limitations; IJCAHPO Core Criteria Handbook, biometry content areaReport a problem with this question

  18. 18. When screening color vision with pseudoisochromatic plates, how is the test given and what does it mainly detect?

    • A.One eye at a time in good light; it screens red-green defects✓ Answer
    • B.Both eyes together in good light; it screens blue-yellow loss
    • C.One eye at a time in a dim room; it screens blue-yellow loss

    The plates hide their figures in colors that lie along the red-green confusion axis, so that is the deficiency they reveal; blue-yellow and acquired defects need other plate sets or arrangement tests. Testing is monocular under good daylight-balanced lighting with the near correction, because an inherited defect is symmetric while an acquired one may affect one eye only and would be masked binocularly.

    Ref QB-Q4NNCPSource: Pseudoisochromatic plate colour vision screening; IJCAHPO Core Criteria Handbook, supplemental testing content areaReport a problem with this question

  19. 19. A patient reads the acuity chart well in the exam lane but cannot drive in bright sun. Which supplemental test documents this complaint best?

    • A.Color plates screened one eye at a time with near correction
    • B.Best-corrected acuity remeasured with a bright light in view✓ Answer
    • C.Best-corrected acuity remeasured through a pinhole aperture

    A dim examination lane hides the problem: light scattered by a cataract or a clouded posterior capsule reduces contrast only when a bright source is in the field of view. Glare testing repeats best-corrected acuity with that source present, so the drop in lines is documented objectively. A pinhole tests for refractive causes, and color testing answers an entirely different question.

    Ref QB-WFBMJYSource: Glare testing in cataract evaluation; IJCAHPO Core Criteria Handbook, supplemental testing content areaReport a problem with this question

  20. 20. During applanation, a technician holds the patient's upper lid open by pressing it back against the eyeball, and the patient is also squeezing. What effect does this most likely have on the reading?

    • A.It has no effect, since the prism measures only central cornea
    • B.It reads falsely high, since pressure on the globe adds force✓ Answer
    • C.It reads falsely low, since lid traction flattens the cornea

    Any outside pressure on the globe passes to the eye's contents. That includes fingers on the lid and the patient squeezing or squinting. The tonometer then needs more force to flatten the cornea, so the reading is artificially high. Hold the lids against the bony orbital rim, not the eye, and ask the patient to relax and breathe normally.

    Ref QB-DY5FYXSource: StatPearls, Tonometry (NBK493225), Technique Considerations: avoid artificially raising IOP by pressing on the eyelid; squinting can raise IOPReport a problem with this question

  21. 21. A patient declines all eye drops but still needs an intraocular pressure check. Which method can be done without topical anesthetic or fluorescein?

    • A.Goldmann slit-lamp applanation
    • B.Non-contact air-puff tonometry✓ Answer
    • C.Hand-held Tono-Pen tonometry

    A non-contact tonometer flattens the cornea with a jet of air. Nothing touches the eye and no prism mires are read, so no anesthetic or fluorescein is needed. Goldmann applanation needs anesthetic plus fluorescein to see the mires. The Tono-Pen touches the cornea and needs topical anesthetic. Rebound (iCare) tonometry is the other drop-free option.

    Ref QB-VFY9DXSource: StatPearls, Applanation Tonometry (NBK582132): NCT instruments generate force by air and do not require fluorescein and local topical anesthesia; Goldmann and Perkins require bothReport a problem with this question

  22. 22. A bedridden patient cannot be brought to the slit lamp, and the doctor wants a reading taken by the same applanation principle as Goldmann. Which instrument fits?

    • A.Schiotz weighted indentation tonometer
    • B.Slit-lamp Goldmann with the chair reclined
    • C.Perkins hand-held applanation tonometer✓ Answer

    The Perkins tonometer uses the same split-prism applanation system as Goldmann but is portable, so it can be used with the patient lying down. Slit-lamp Goldmann needs a seated, upright patient. Schiotz is portable but works by indentation, a different principle.

    Ref QB-CYS4TWSource: StatPearls, Applanation Tonometry (NBK582132): GAT and Perkins use the same system, differing only in that GAT requires a slit lamp while Perkins is portable and can be used supineReport a problem with this question

  23. 23. An assistant measures pressure with a Tono-Pen. Which routine matches how this instrument is meant to be used?

    • A.Anesthetize, use a new tip cover, average several taps✓ Answer
    • B.Skip drops, and reuse the cover after an alcohol wipe
    • C.Add fluorescein, align the mires, read the dial times 10

    The Tono-Pen touches the cornea, so it needs topical anesthetic. It uses a disposable cover for each patient, and the XL is calibration-checked daily. It averages several readings into one result (4 on the XL, 10 on the Avia). Fluorescein mires and the dial x10 belong to Goldmann.

    Ref QB-YF2AW2Source: StatPearls, Tonometry (NBK493225), Tono-Pen section: averages multiple readings (XL 4, Avia 10); requires daily recalibration and topical anesthesia and uses disposable coversReport a problem with this question

  24. 24. A Schiotz tonometer can give a misleading reading in some eyes even when the technique is flawless. Which property of the eye mainly causes this?

    • A.Amount of fluorescein in the tear film
    • B.Width of the anterior chamber angle
    • C.Rigidity of the sclera and eye wall✓ Answer

    Schiotz is an indentation tonometer. A weighted plunger sinks into the cornea, and the conversion table assumes an average eye wall. An unusually stiff or unusually stretchy eye changes how far the plunger sinks at the same pressure, so ocular rigidity skews the result. Fluorescein matters for Goldmann mires, not for Schiotz.

    Ref QB-FYU4B2Source: StatPearls, Tonometry (NBK493225), Schiotz Tonometer: weighted plunger indentation; measurements subject to errors including influences due to ocular rigidityReport a problem with this question

  25. 25. Which range is conventionally cited as normal adult intraocular pressure, with readings above its upper limit commonly described as ocular hypertension?

    • A.5 to 15 mmHg
    • B.10 to 21 mmHg✓ Answer
    • C.15 to 30 mmHg

    The usual reference range is 10 to 21 mmHg, and pressures above 21 mmHg are commonly labeled ocular hypertension. This is a statistical convention, not a diagnosis. Glaucoma can occur with pressures inside the range, and a high reading alone is not glaucoma. The assistant records the reading, the method, the eye and the time.

    Ref QB-XFPXQZSource: StatPearls, Tonometry (NBK493225), Introduction: the normal range of intraocular pressure is 10 to 21 mmHg; ocular hypertension, in which IOP tends to be higher than 21 mmHgReport a problem with this question

  26. 26. Using the classic Heijl-Krakau method, how does an automated perimeter check for fixation losses during the test?

    • A.It flashes stimuli into the mapped blind spot✓ Answer
    • B.It pauses and presents no stimulus at all
    • C.It re-tests seen points with a brighter light

    The perimeter locates the physiologic blind spot at the start, then flashes stimuli there at random times. An eye that holds steady on the fixation target cannot see them, so a response means the gaze has moved. The manufacturer flags more than 20% fixation losses as unreliable. Blank presentations measure false positives, and brighter retests at seen points measure false negatives.

    Ref QB-EYGS5ZSource: StatPearls, Visual Field Testing / Perimetry (NBK585112), reliability indices: fixation losses calculated by the Heijl-Krakau method (random stimuli to the located blind spot); over 20% indicates poor reliability per the manufacturerReport a problem with this question

  27. 27. An automated field printout shows a high false-negative rate. What does this index record, and which way does it bias the field?

    • A.Presses with no stimulus shown; the field looks better
    • B.Missed bright repeats at seen points; field looks better
    • C.Missed bright repeats at seen points; field looks worse✓ Answer

    A false negative is recorded when the patient fails to respond to a brighter stimulus at a spot where a dimmer one was already seen. That usually means fatigue or inattention. These misses add depressions that are not real, so the field looks worse than it is. False positives, pressing with no stimulus, do the opposite and make the field look better.

    Ref QB-2GRX85Source: StatPearls, Visual Field Testing / Perimetry (NBK585112), reliability indices: false negatives = no response to a higher-intensity stimulus where a lower one was seen; more than 33% suggests poor reliability; excess false positives mask defectsReport a problem with this question

  28. 28. On a Goldmann kinetic field, one isopter is labeled V4e. What do the Roman numeral V and the Arabic numeral 4 indicate?

    • A.V is the intensity; 4 is the target size
    • B.V is the test distance; 4 is the speed
    • C.V is the target size; 4 is the intensity✓ Answer

    In Goldmann notation the Roman numeral (I to V) is the target size, and each step is 4 times the area. The Arabic numeral (1 to 4) is the relative intensity, and each step is about 3.15 times brighter. The lowercase letter (a to e) is a minor filter in 0.1 log-unit steps, with e the brightest. V4e is therefore the largest, brightest standard target and plots the widest isopter.

    Ref QB-HYJSN4Source: Ophthalmology Secrets, Visual Fields chapter, Q7 (entokey.com/visual-fields-3): Roman numeral = target size (each step x4), Arabic numeral = relative intensity (each step 3.15x brighter), lowercase letter = minor filter in 0.1 log-unit stepsReport a problem with this question

  29. 29. For confrontation fields, the patient covers the left eye with the left hand. How should the examiner set up so the two fields can be compared?

    • A.Close own left eye; patient looks at examiner's left eye
    • B.Close own right eye; patient looks at examiner's left eye✓ Answer
    • C.Close own right eye; patient watches the moving fingers

    The patient's open right eye fixates the examiner's left eye, and the examiner closes the right eye. The two open eyes then face each other and their fields overlap, so the examiner's own normal field is the reference for each quadrant. The patient must keep looking at the examiner's eye rather than chase the fingers, or the test no longer measures peripheral vision.

    Ref QB-TDBF2VSource: StatPearls, Hemianopsia (NBK562262), Evaluation, Confrontation steps: patient covers left eye with left hand, fixates right eye on the examiner's left eye; the examiner closes their right eyeReport a problem with this question

  30. 30. A standard Amsler grid is a 10 cm square held about 33 cm from the eye. How much of the visual field does it test?

    • A.About 10 degrees on each side of fixation✓ Answer
    • B.About 30 degrees on each side of fixation
    • C.About 5 degrees on each side of fixation

    At 33 cm the 10 cm grid covers a 20-degree field, 10 degrees above, below, left and right of fixation. That makes it a macular test for distortion and central scotomas. The physiologic blind spot, about 15 degrees from fixation, falls outside the grid, so it is not seen. Glaucoma defects may not show until they come close to fixation.

    Ref QB-BWGMMVSource: StatPearls, Amsler Grid (NBK538141), Equipment: grid measures 10 cm x 10 cm; at 33 cm it checks a 20-degree field, 10 degrees on each side of fixation; physiologic blind spot not appreciatedReport a problem with this question

  31. 31. A patient has swollen optic discs from raised intracranial pressure (papilledema). Which visual field finding is most commonly seen?

    • A.A dense central scotoma at fixation
    • B.A superior arcuate defect from the disc
    • C.An enlarged physiologic blind spot✓ Answer

    The physiologic blind spot is the projection of the optic disc. When the disc swells, the area of non-seeing retina around it grows, so the blind spot enlarges. Generalized constriction appears later if optic atrophy develops. A central scotoma suggests optic neuritis or macular disease, and an arcuate defect suggests glaucoma.

    Ref QB-UDPS7WSource: StatPearls, Visual Field Testing / Perimetry (NBK585112), Clinical Significance, Papilledema: enlargement of the blind spot is the most commonly seen visual field defect; constriction in late cases with optic atrophyReport a problem with this question

  32. 32. A technician checks IOP with a handheld rebound tonometer (for example, iCare) loaded with a new single-use probe. How should the device be set up and positioned?

    • A.Anesthetic drops; probe tip resting directly on the central cornea
    • B.No drops; probe aimed at the temporal limbus, held 15-20 mm away
    • C.No drops; probe perpendicular to the central cornea, held 4-8 mm away✓ Answer

    A rebound tonometer fires a light probe that touches the cornea only for an instant and measures how it decelerates and rebounds, so no topical anesthetic is needed. For the readings to be valid the probe must hit the center of the cornea at a right angle, with the tip about 4-8 mm from the cornea on standard models; aiming at the limbus or resting the tip on the eye defeats the rebound measurement.

    Ref QB-CWUXSWSource: NICE Medtech Innovation Briefing MIB57, 'Icare rebound tonometer to measure intraocular pressure', Description: brief corneal contact, usable without local anaesthesia; probe perpendicular to the centre of the cornea, 4-8 mm from the cornea (TA01i, ic100, HOME)Report a problem with this question

  33. 33. A glaucoma patient's air-puff (non-contact) tonometer reading is 32 mmHg. Compared with Goldmann applanation, how do air-puff tonometers tend to err across the pressure range?

    • A.Read high at low pressures and low at high pressures✓ Answer
    • B.Read low at low pressures and high at high pressures
    • C.Read high by a fixed amount at every pressure level

    Studies comparing air-puff tonometry with Goldmann show it overestimates IOP in the lower range and underestimates it in the higher range, so the error is not a fixed offset. A high air-puff reading such as 32 mmHg may therefore understate the true pressure, which is why elevated non-contact readings are confirmed with Goldmann applanation, the reference standard.

    Ref QB-VD25DXSource: AAO EyeWiki, 'The Reliability of Intraocular Pressure Measurements', Non-contact tonometry: air-puff tonometry overestimates IOP in the lower range and underestimates it in the higher range compared with GAT; Goldmann and Perkins are the gold standardReport a problem with this question

  34. 34. A patient with healthy-looking discs has Goldmann readings of 24 mmHg in both eyes, and pachymetry shows a central corneal thickness of about 610 µm. How does the thick cornea most likely affect the reading?

    • A.The true pressure is probably higher than the reading
    • B.The true pressure is probably lower than the reading✓ Answer
    • C.Corneal thickness has no effect on applanation readings

    Goldmann applanation was designed for a cornea of average thickness. A physiologically thick cornea resists flattening, so more force is needed and the tonometer overestimates the true IOP; a substantial share of 'high' pressures in glaucoma suspects with thick corneas are measurement artifacts. That is why pachymetry is recorded alongside IOP, and why the reading here likely overstates the real pressure.

    Ref QB-DW7AYXSource: AAO EyeWiki, 'The Reliability of Intraocular Pressure Measurements', Applanation section: applanation designed for an average corneal thickness; physiologically thick corneas overestimate the actual IOP; 30-57% of elevated IOPs in glaucoma suspects are measurement artifactsReport a problem with this question

  35. 35. Goldmann perimetry is kinetic, while Humphrey threshold testing is static. How is a point on a kinetic isopter found?

    • A.A stationary target is dimmed at each point until it is missed
    • B.A target is moved from an unseen area inward until it is seen✓ Answer
    • C.A target is moved from fixation outward until it disappears

    In kinetic perimetry a target of fixed size and brightness is moved at a standardized speed (about 2-4 degrees per second) from a non-seeing area toward a seeing area, and the spot where the patient first reports it is plotted; joining those spots for one target forms an isopter. Dimming a stationary target at fixed locations describes static threshold testing, and moving from seeing to non-seeing reverses the standard direction.

    Ref QB-WDDFJYSource: AAO EyeWiki, 'Standard Automated Perimetry', Static vs. dynamic perimetry: in kinetic perimetry a stimulus is moved from a non-seeing (subthreshold) area to a seeing area at 2-4 degrees per second; in static perimetry stationary stimuli are presented at defined pointsReport a problem with this question

  36. 36. During an A-scan, the axial length is mismeasured by 1 mm. Roughly how large an error does this create in the calculated IOL power?

    • A.About 0.25 to 0.5 diopter
    • B.About 8 to 10 diopters
    • C.About 2.5 to 3 diopters✓ Answer

    Axial length is the most influential input in modern IOL formulas: each millimeter shifts the required IOL power by roughly 2.5 to 3 D (about 0.27 D for every 0.1 mm), with larger effects in short eyes. That is why even small biometry errors, such as corneal compression during contact A-scan, can leave a patient with a clear refractive surprise.

    Ref QB-EWJM5ZSource: AAO EyeWiki, 'Biometry for Intra-Ocular Lens (IOL) Power Calculation', Axial length section: axial length changes the IOL power by nearly 2.5 to 3 times, more so in short eyes; Optometry Times, 'Retinal considerations crucial for accurate biometry': a 0.1 mm error equates to about 0.27 DReport a problem with this question

  37. 37. For a cataract patient, the average K entered into the IOL formula is off by 1.00 D from the true corneal power. About how much does this shift the calculated IOL power?

    • A.About 1 diopter✓ Answer
    • B.About 0.1 diopter
    • C.About 3 diopters

    Corneal power is the second most important input to IOL calculation, and changes in K alter the IOL power in a ratio of nearly 1:1. A 1.00 D keratometry error therefore produces about a 1 D error in the lens chosen, which is why K readings must be accurate and repeatable before they are used. The roughly 3 D figure belongs to a 1 mm axial length error, not to K.

    Ref QB-XDRSQZSource: AAO EyeWiki, 'Biometry for Intra-Ocular Lens (IOL) Power Calculation', Keratometry section: corneal power is the second most important factor for IOL power determination; changes in K alter the IOL power in a ratio of nearly 1:1Report a problem with this question

  38. 38. A patient has a dense vitreous hemorrhage, and the doctor cannot see the retina. Which test can show whether the retina is detached behind the blood?

    • A.Optical partial-coherence biometry
    • B.Two-dimensional B-scan ultrasonography✓ Answer
    • C.Automated static threshold perimetry

    Sound passes through blood and other opaque media that block light, and B-scan displays the echoes as a two-dimensional cross-section of the globe. When a hemorrhage hides the retina, B-scan can reveal retinal tears, retinal detachment, posterior vitreous detachment, tumors or foreign bodies. Optical biometry relies on light and fails in dense opacities, and perimetry measures function rather than imaging the retina.

    Ref QB-FWWXA2Source: AAO EyeWiki, 'Vitreous Hemorrhage', Diagnosis: if the hemorrhage obscures a complete view of the retina, B-scan ultrasonography can detect PVD, retinal tears, retinal detachment, tractional membranes, intraocular tumors and foreign bodiesReport a problem with this question

  39. 39. Before dilating a patient, an assistant performs a van Herick estimate at the slit lamp. The peripheral anterior chamber gap looks less than one-quarter of the adjacent corneal thickness. What should be reported?

    • A.A wide angle, so dilation can safely proceed as planned
    • B.A narrow angle, so the doctor should check before dilating✓ Answer
    • C.A thin cornea, so the pressure reading will run falsely low

    The van Herick technique aims a thin slit beam at about 60 degrees onto the peripheral cornea near the limbus and compares the depth of the peripheral anterior chamber with the corneal thickness. A ratio below 1/4 is grade 1, where angle closure is likely. It is most useful before dilation and cannot replace gonioscopy, so the assistant documents the finding and has the doctor assess the angle before drops are given.

    Ref QB-Y75CX2Source: EyeWiki (AAO), Gonioscopy, 'Van Herick' section and Table 5: slit beam about 60 degrees near the limbus, corneal thickness compared with peripheral AC depth; grade 1 = ratio <1/4, angle closure likely; most helpful before dilation; should not replace gonioscopyReport a problem with this question

  40. 40. A Schirmer strip is placed in the lateral third of the lower lid with no topical anesthetic, and the wetting is measured after 5 minutes. What does this version of the test measure?

    • A.Basal tear secretion alone, without reflex
    • B.Reflex tearing alone, without basal flow
    • C.Basic tearing plus reflex tearing combined✓ Answer

    Without anesthetic, the strip itself irritates the eye, so the Schirmer I test captures both basic and reflex tearing. Adding topical anesthetic blocks the reflex component, which turns it into the basic secretion test of basal tears only. Schirmer II, done after nasal stimulation, is the version aimed at reflex tearing.

    Ref QB-FQX6C2Source: EyeWiki (AAO), Dry Eye Syndrome, 'Schirmer test': strip in the lateral third of the lower eyelid, wetting measured after 5 minutes; Schirmer I without anesthesia measures basic and reflex tearing; Schirmer with anesthesia measures basal secretion; Schirmer II follows nasal stimulationReport a problem with this question

  41. 41. For tear break-up time, fluorescein is applied and the patient blinks, then stares straight ahead without blinking while the assistant watches under cobalt blue light. What interval is recorded?

    • A.From the last blink to the first dry spot in the tear film✓ Answer
    • B.From dye instillation to the first dry spot in the tear film
    • C.From the last blink to complete fading of the fluorescein

    TBUT measures tear-film stability, so the clock starts when the last blink spreads a fresh, even film and stops when the first dry spot or hole appears in it. The strip should be wet with non-preserved saline, because the preservative benzalkonium chloride speeds up break-up. About 10 seconds is the typical cutoff between normal and abnormal.

    Ref QB-Z7GP44Source: EyeWiki (AAO), Dry Eye Syndrome, 'Tear break up time (TBUT)': fluorescein strip wet with non-preserved saline; time between the last blink and the first dry spot under cobalt blue light; 10 seconds typical cutoffReport a problem with this question

  42. 42. A patient with ocular hypertension is sent for pachymetry. Why is central corneal thickness important in this workup?

    • A.Thick corneas raise glaucoma risk and can make IOP read low
    • B.Thin corneas raise glaucoma risk and can make IOP read low✓ Answer
    • C.Thin corneas lower glaucoma risk and can make IOP read high

    In the large ocular hypertension trials, a thin central cornea was one of the strongest predictors of converting to glaucoma. A thin cornea also offers less resistance to applanation, so the measured pressure can underestimate the true IOP. Pachymetry therefore informs both risk and how the doctor reads the tonometry number.

    Ref QB-GQAHH3Source: EyeWiki (AAO), Ocular Hypertension, Risk Factors: thin corneas (CCT <555) are among the strongest predictors and influence true IOP estimation; EyeWiki, Primary Open-Angle Glaucoma: a thinner cornea may lead to measurements that underestimate true IOPReport a problem with this question

  43. 43. When setting up a Humphrey threshold test, the assistant types in the wrong birth date, making the patient 20 years younger than they are. Why does this matter?

    • A.The perimeter shrinks the stimulus to suit a younger patient
    • B.The machine skips blind-spot checks for younger patients
    • C.Results are compared with normal values for the wrong age group✓ Answer

    Each threshold value is compared with an age-adjusted normal database. Because normal sensitivity declines with age, entering a younger age compares an older eye against higher normals, which can make the deviation plots show loss that is not really there. The stimulus size and fixation monitoring are set by the program, not by the age entered.

    Ref QB-W7EPKYSource: EyeWiki (AAO), Standard Automated Perimetry, 'Common perimetric errors': incorrect patient age leads to comparison with the wrong set of age-adjusted normal valuesReport a problem with this question

  44. 44. A glaucoma patient using pilocarpine has very small pupils. The new automated field shows a general, overall drop in sensitivity compared with last year. What should the assistant note for the doctor?

    • A.The small pupils enlarge the blind spot and cause fixation losses
    • B.The small pupils let in less light and can depress the whole field✓ Answer
    • C.The small pupils block the edge stimuli like a lens rim artifact

    A miotic pupil, like a cataract, lets less light reach the retina, which causes diffuse depression of the field and can falsely suggest glaucoma progression. Recording the pupil size lets the doctor weigh this, often using the pattern deviation plot, which adjusts for generalized depression.

    Ref QB-DQ8HZXSource: EyeWiki (AAO), Standard Automated Perimetry, 'Common perimetric errors' - Miotic pupils or cataracts: less light reaches the retina, causing diffuse depression and a false impression of progression; pattern deviation plot adjusts for generalized depressionReport a problem with this question

  45. 45. A field printout shows fixation losses of 8/20 and false negatives of 40%, and the plot is full of scattered deep defects. What is the appropriate assistant action?

    • A.Flag the test as unreliable, re-instruct the patient and retest✓ Answer
    • B.File the test as valid, since the defects prove real field loss
    • C.Rerun it on the 10-2 program, which ignores reliability indices

    Fixation losses, false positives and false negatives measure how well the patient cooperated. High rates mean the defects may be artifacts and the comparison with normals may not be valid; clinical trials such as OHTS excluded fields when an index exceeded 33%. The assistant stays in role by flagging the test, re-explaining the task and retesting, rather than reporting the defects as real loss. Every threshold program, including 10-2, reports these indices.

    Ref QB-X7SZRZSource: EyeWiki (AAO), Standard Automated Perimetry, 'Establishing reliability of results': fixation losses, false positives and false negatives; fields with rates over 33% excluded as unreliable (OHTS); high rates indicate poor cooperation and inaccurate resultsReport a problem with this question

  46. 46. A patient taking hydroxychloroquine is scheduled for an annual toxicity screening field, and the doctor orders a 10-2 program. Why this program?

    • A.It samples the far periphery densely, where toxicity tends to appear
    • B.It maps the blind spot densely, where toxicity tends to appear
    • C.It samples the central macula densely, where toxicity tends to appear✓ Answer

    The 10-2 field has high resolution within the macula, and early hydroxychloroquine toxicity typically shows as a paracentral scotoma about 2 to 6 degrees from fixation in non-Asian patients. Patients of Asian descent may need a wider 24-2 or 30-2 pattern because damage can extend beyond the central field. Annual screening pairs the field with SD-OCT.

    Ref QB-EQKU6ZSource: EyeWiki (AAO), Hydroxychloroquine Toxicity: annual screening with automated fields and SD-OCT; the 10-2 field has high resolution within the macula; area of risk typically 2 to 6 degrees from center in non-Asian patients; wider 24-2/30-2 for patients of Asian descentReport a problem with this question

  47. 47. A patient with known early glaucoma has full fields to finger-counting confrontation testing. How should this result be understood?

    • A.It rules out field loss, so automated perimetry can be skipped
    • B.It is a gross screen that often misses early glaucoma loss✓ Answer
    • C.It is more sensitive than automated perimetry for early loss

    Confrontation compares the patient's field with the examiner's own using finger counting or hand motion. It is quick, but its ability to detect early glaucomatous loss is poor and it mainly reveals gross defects. A normal confrontation result therefore does not replace the automated perimetry the doctor ordered.

    Ref QB-48JPM7Source: EyeWiki (AAO), Standard Automated Perimetry, 'Alternatives When Automated Perimetry Is Unavailable': confrontation compares the patient's field with the examiner's using finger counting or hand motion; poor ability to detect early glaucomatous loss, may reveal gross defectsReport a problem with this question

  48. 48. On a contact A-scan, the retinal spike rises tall and steep, but it is not followed by the usual train of smaller orbital echoes. What does this mean?

    • A.The beam is on the macula, so this reading is the best one
    • B.The cornea is compressed, so lower the gain and keep it
    • C.The beam is on the optic nerve, so discard this reading✓ Answer

    In a good scan, orbital fat produces medium-to-low echoes right after the retinal spike. When those echoes are missing, the beam is passing down the optic nerve rather than hitting the macula, so the reading should be discarded. Corneal compression shows differently, as a shortened anterior chamber depth and axial length.

    Ref QB-KQCH27Source: EyeWiki (AAO), Biometry for Intra-Ocular Lens (IOL) Power Calculation, 'Qualities of a Good A-Scan Graph': orbital fat produces medium-to-low echoes; if the retinal spike is not followed by multiple small spikes, you are hitting the optic nerve, discard that reading; reduced ACD and AL indicate corneal compressionReport a problem with this question

  49. 49. A cataract patient needs keratometry, applanation tonometry and a contact A-scan at the same visit. Which order protects the accuracy of the K readings?

    • A.Applanation first, since the anesthetic smooths the cornea for Ks
    • B.Keratometry first, before any contact test touches the cornea✓ Answer
    • C.Contact A-scan first, since probe contact steadies the tear film

    Keratometry reads the curvature of the front corneal surface from reflected mires, so anything that disturbs the epithelium or tear film first, such as a tonometer tip, an A-scan probe or the drops used for them, can distort the result. Biometry guidance is to avoid other contact procedures before keratometry and to average several readings.

    Ref QB-V557EXSource: EyeWiki (AAO), Biometry for Intra-Ocular Lens (IOL) Power Calculation, 'Tips for K Measurement': do not perform other contact procedures before keratometry; take an average of 3 readingsReport a problem with this question

  50. 50. Automated keratometry gives an average K of 43.00 D in the right eye and 44.75 D in the left. The patient has no history of corneal surgery or disease. What should the assistant do?

    • A.Recheck the readings, since the eyes differ by more than 1.0 D✓ Answer
    • B.Accept them, since each value falls in the normal K range
    • C.Average both eyes and enter that single value for each eye

    The two corneas are usually similar, so a commonly used biometry rule is to repeat keratometry when the eyes differ by more than 1.0 D. Here the gap is 1.75 D. Being inside the normal range does not make each value right, and averaging would put a wrong K into both IOL calculations. The assistant repeats, has a second person cross-check if needed, and compares with the refraction.

    Ref QB-DNACZXSource: EyeWiki (AAO), Biometry for Intra-Ocular Lens (IOL) Power Calculation, 'Tips for K Measurement': repeat if the difference is more than 1.0 D between the two eyes; ask a second person to cross-check; compare the refraction with the K valuesReport a problem with this question

Other topics in this exam

Practice questions based on the IJCAHPO core certification content areas for the Certified Ophthalmic Assistant and on standard ophthalmic clinical practice references. COA, COT, and COMT are marks of the International Joint Commission on Allied Health Personnel in Ophthalmology; this site is not affiliated with or endorsed by IJCAHPO. Ophthalmic assistants work under the delegation of a supervising ophthalmologist and permitted tasks vary by state and by practice — confirm your own scope, and check the current exam requirements and content outline with IJCAHPO before testing. About IJCAHPO certification →