28 Ophthalmic Imaging & Photography Practice Questions & Answers
Every Ophthalmic Imaging & Photography practice question from the Ophthalmic Assistant (COA) Practice Test, with the correct answer and a short explanation.
Start practice test →1. During intravenous fluorescein angiography, which sequence correctly describes the filling phases as the dye reaches the eye?
- A.Arterial, choroidal flush, venous, arteriovenous, then late
- B.Choroidal flush, venous, arteriovenous, arterial, then late
- C.Choroidal flush, arterial, arteriovenous, venous, then late✓ Answer
The dye reaches the choroid first through the short posterior ciliary arteries, producing the patchy choroidal flush, then fills the retinal arterioles, then the capillary bed with laminar flow along the vein walls, then the veins completely; leakage and staining are judged on the late frames after recirculation.
Ref QB-S7SPMVSource: Fluorescein angiography technique, IJCAHPO Core Criteria Handbook ophthalmic imaging content areaReport a problem with this question
2. In a fundus camera set up for fluorescein angiography, what is the function of the barrier filter?
- A.Transmits the yellow-green light emitted by the dye and blocks reflected blue✓ Answer
- B.Corrects the photographer's own refractive error at the camera eyepiece
- C.Delivers the blue excitation light into the eye before the dye is injected
Fluorescein is excited by blue light and then emits longer-wavelength yellow-green light; the exciter filter supplies the blue illumination and the barrier filter in front of the sensor passes only the emitted yellow-green fluorescence, so reflected blue light does not wash out the image.
Ref QB-9QKH2VSource: Optics of fluorescein angiography filters, ophthalmic photography practice referencesReport a problem with this question
3. On an angiogram, an area of hyperfluorescence is small and faint early, then becomes larger and brighter with blurred margins on the late frames. Which mechanism does this describe?
- A.Late staining of an old chorioretinal scar
- B.Leakage of dye from abnormally permeable vessels✓ Answer
- C.A window defect from pigment epithelial atrophy
Leakage is the only cause of hyperfluorescence that both enlarges and intensifies as the study progresses, because dye keeps escaping into tissue or fluid; a window defect is bright early and fades with the dye without changing size, and staining brightens late but stays within fixed borders.
Ref QB-R7ECFUSource: Interpretation vocabulary of fluorescein angiography, ophthalmic imaging practice referencesReport a problem with this question
4. A patient telephones the morning after a fluorescein angiogram, alarmed that her urine is bright yellow. What is the best response?
- A.Arrange kidney function testing before any further angiography is scheduled
- B.Reassure her that the kidneys clear the dye and this color is normal✓ Answer
- C.Send her to the emergency department for a suspected dye reaction
Discolored urine and transient yellowing of the skin are common, temporary effects of the dye being excreted, not an allergic reaction, so the correct action is reassurance; patients should be told about this before the injection so the call is not frightening.
Ref QB-CQNHJYSource: FDA Fluorescite label, section 6.1 (skin and urine discoloration)Report a problem with this question
5. During the injection the patient reports burning at the venipuncture site and a swelling is rising under the skin. What should be done immediately?
- A.Elevate the arm and restart in the same vein higher up
- B.Massage the swollen area and continue the early frames
- C.Stop the injection, tell the physician and apply cold✓ Answer
Burning with local swelling means the dye is extravasating into the tissue rather than entering the vein; fluorescein outside the vessel is irritating and can cause local tissue damage, so the infusion is stopped at once, the physician is informed and cold is applied. Re-injecting the same vein or massaging the site spreads more dye into the tissue.
Ref QB-U7GCYXSource: Management of intravenous site complications during angiography, ophthalmic assisting practiceReport a problem with this question
6. Shortly after the injection the patient develops widespread hives, wheezing and difficulty speaking. What is the assistant's most appropriate action?
- A.Send the patient home and telephone them in the morning
- B.Call the physician at once, start emergency protocol, give oxygen✓ Answer
- C.Continue late-phase photography and report the rash later
Urticaria combined with bronchospasm and airway compromise is a true hypersensitivity reaction rather than the common transient nausea, so it is a medical emergency: the assistant stays with the patient, summons the physician, activates the practice emergency plan and supports oxygenation while the emergency cart is brought.
Ref QB-BQA7DXSource: Adverse dye reaction recognition and emergency response, IJCAHPO Core Criteria Handbook ophthalmic imaging content areaReport a problem with this question
7. When the history is taken before a planned angiogram, which item most directly affects whether and how the study proceeds?
- A.The date of the patient's most recent spectacle prescription
- B.A family history of age-related macular degeneration
- C.A previous reaction to an injected dye or contrast agent✓ Answer
A documented prior reaction to injected dye predicts a higher risk of a repeat reaction and must be reported to the physician before the study, along with asthma, other allergies and relevant liver or kidney disease; consent and a checked emergency cart are part of the same preparation.
Ref QB-T752TWSource: Pre-angiography history, screening and consent, ophthalmic assisting practiceReport a problem with this question
8. Why are color and red-free (green-filtered) fundus photographs taken before the fluorescein is injected?
- A.They give a control image on which blood and pigment look dark✓ Answer
- B.They confirm that dye has already reached the choroidal vessels
- C.They remove the need to refocus the camera after the injection
Green-filtered light is absorbed by blood, so hemorrhages and pigment render dark and superficial retinal detail stands out; taken before any dye is present, these frames are the baseline against which the angiogram is compared, while a separate pre-injection frame shot through the angiography filters shows whether later brightness is real fluorescence or autofluorescence and pseudofluorescence.
Ref QB-6QHHHRSource: Baseline colour and red-free photography in angiography protocol, ophthalmic photography practice referencesReport a problem with this question
9. Compared with fluorescein angiography, what does indocyanine green angiography demonstrate better?
- A.The structures lying behind the iris in an eye with cloudy media
- B.The retinal nerve fiber layer thickness around the optic disc
- C.The choroidal vessels beneath the retinal pigment epithelium✓ Answer
Indocyanine green absorbs and emits in the near-infrared, which penetrates the retinal pigment epithelium, macular pigment, thin hemorrhage and hazy media, and it is far more strongly protein-bound, so less dye escapes the fenestrated choriocapillaris and the choroidal vasculature is imaged cleanly.
Ref QB-N7CCXQSource: Indocyanine green angiography principles, ophthalmic imaging practice referencesReport a problem with this question
10. A patient asks how an optical coherence tomography scan produces its image. Which explanation is accurate?
- A.An injected dye outlines the retinal vessels as it circulates
- B.Reflected light builds a cross-sectional image of the retina✓ Answer
- C.Sound waves reflected from the retina are timed through a gel
Optical coherence tomography uses interference between light reflected from tissue and a reference beam to measure echo delay, giving a cross-section of the retinal layers; it is the optical analogue of B-scan ultrasound but needs no dye, no injection and no contact with the eye.
Ref QB-MBQWKXSource: Principles of optical coherence tomography, IJCAHPO Core Criteria Handbook scanning laser imaging criteriaReport a problem with this question
11. A glaucoma suspect's nerve fiber layer scan returns with low signal strength and one sector coded red; the patient has marked dry eye and blinked during acquisition. What should the assistant do?
- A.Tell the patient the red sector confirms glaucoma damage
- B.Save the scan and report the nerve fiber layer as thin
- C.Lubricate, ask for a full blink and repeat the scan✓ Answer
An unstable tear film, blinking or media opacity scatters the beam, lowers signal strength and causes segmentation errors that mimic thinning, so a low-quality scan is repeated after smoothing the tear film rather than saved; interpreting the color code for the patient is the physician's role.
Ref QB-5UV37YSource: Scan quality and artifact recognition in scanning laser imaging, ophthalmic assisting practiceReport a problem with this question
12. On a Placido-disc corneal topography map using the standard color scale, what does an area shown in warm red-orange indicate?
- A.It is steeper there, with higher dioptric power✓ Answer
- B.It is flatter there, with lower dioptric power
- C.It is thinner there, with reduced pachymetry
The conventional scale codes curvature, not thickness or clarity: warm colors mark steeper radii and therefore higher dioptric power, cool colors mark flatter, lower-power areas, which is why a Placido map alone cannot report corneal thickness.
Ref QB-KBCKEWSource: Corneal topography map conventions, IJCAHPO Core Criteria Handbook corneal topography criteriaReport a problem with this question
13. A first topography map shows broken, irregular rings across the surface. The patient wears soft contact lenses and has an unstable tear film. What is the most likely explanation?
- A.Excess dilating drops instilled just before the measurement
- B.Lens warpage and an unstable tear film distort the reflected rings✓ Answer
- C.A miscalibrated eyepiece reticle inside the instrument optics
Placido topography measures the mirror image of the rings reflected off the tear film, so anything that spoils that smooth optical surface, including contact-lens warpage and a dry or broken tear film, produces false irregularity; the fix is a full blink and re-measurement, with lenses left out beforehand.
Ref QB-4UHRZWSource: Patient preparation for corneal topography, ophthalmic assisting practiceReport a problem with this question
14. Before taking fundus photographs the photographer focuses the crosshair reticle in the camera eyepiece. What does this adjustment accomplish?
- A.It centers the illumination ring inside the patient's pupil
- B.It offsets the photographer's own refractive error at the eyepiece✓ Answer
- C.It corrects the patient's refractive error before capture
The reticle sits in the same plane as the film or sensor, so bringing it sharp with accommodation relaxed, turning from plus toward minus and repeating the setting, cancels the photographer's own refraction; only then does a fundus that looks sharp in the eyepiece record sharp.
Ref QB-JBZ89VSource: Fundus camera eyepiece focusing technique, ophthalmic photography practice referencesReport a problem with this question
15. With the camera at the correct working distance, a fundus photograph shows a bright crescent of light along one edge of the frame with the opposite side underexposed. What is the most likely cause?
- A.A film of dust has settled on the objective lens
- B.An upper eyelash has drifted across the light path
- C.Part of the illumination ring is blocked by the iris✓ Answer
The illumination enters as a ring through the pupil separately from the imaging path, so if the camera is decentered or the pupil is too small the iris clips one side of that ring and the light spills as a crescent, leaving the opposite side dark; recentering or better dilation fixes it.
Ref QB-3U6EUVSource: Fundus photography artifact recognition and correction, ophthalmic photography practice referencesReport a problem with this question
16. A patient has a dense vitreous hemorrhage and no view of the fundus. Which study best assesses the retina behind it?
- A.Fluorescein angiography, imaging the dye in late frames
- B.Wide-angle color fundus photography of the posterior pole
- C.B-scan ultrasonography of the posterior segment✓ Answer
Every optical method, including photography, angiography and optical coherence tomography, needs light to reach the retina and return, so all of them fail behind blood or a dense cataract; ultrasound uses sound waves transmitted through the opaque media and still shows retinal detachment or a mass.
Ref QB-HBNW3USource: Indications for diagnostic ophthalmic ultrasound, IJCAHPO Core Criteria Handbook diagnostic ultrasound content areaReport a problem with this question
17. Which slit-lamp illumination technique is best for photographing posterior capsular opacification after cataract surgery?
- A.Sclerotic scatter with the beam placed at the limbus
- B.Specular reflection off the corneal endothelial surface
- C.Retroillumination, with light returning from the fundus✓ Answer
Retroillumination silhouettes an opacity against light reflected back out of the eye, so the capsule wrinkles and deposits appear as dark shapes on a bright orange background; specular reflection is reserved for the endothelial mosaic and sclerotic scatter for subtle corneal haze.
Ref QB-2UT3PUSource: Slit-lamp illumination techniques for anterior segment photography, IJCAHPO Core Criteria Handbook photography criteriaReport a problem with this question
18. When specular microscopy of the cornea is performed, which layer is imaged and what is assessed?
- A.The epithelium, assessing punctate surface staining
- B.The tear film, timing the break-up after a blink
- C.The endothelium, assessing cell density and shape✓ Answer
Specular microscopy captures light mirrored at the interface between the endothelium and the aqueous, showing the hexagonal cell mosaic; the report describes cell density along with variation in cell size and shape, and dark drop-like gaps where guttae interrupt the mosaic.
Ref QB-GBAKWSSource: Specular microscopy of the corneal endothelium, ophthalmic imaging practice referencesReport a problem with this question
19. A child has nystagmus that changes with gaze direction and head posture. Why is video preferred over still photographs for documenting it?
- A.It records the movement over time, as stills cannot✓ Answer
- B.It substitutes for the physician's motility examination
- C.It removes the need to note the gaze position used
A still frame freezes one instant, so it cannot show the waveform, the null position or how the movement changes with gaze; video captures the dynamic finding, and the recording is still labelled with the gaze position and head posture and does not replace the physician's examination.
Ref QB-ZTFRGTSource: Videography for dynamic ocular findings, IJCAHPO Core Criteria Handbook photography and videography content areaReport a problem with this question
20. Reviewing images before they go to the chart, the assistant finds fundus photographs of the left eye stored under the label for the right eye. What should be done?
- A.Send them on and add a note at the next visit
- B.Delete the study and re-photograph the fellow eye
- C.Correct the laterality before the study is filed✓ Answer
Laterality is a patient-safety identifier, not a clerical detail, because a mislabelled study can direct treatment or surgery to the wrong eye; the image set is corrected and verified against the patient and encounter before it enters the record.
Ref QB-H9QR3USource: Image labelling and record documentation requirements, ophthalmic assisting practiceReport a problem with this question
21. The practice wants to use an identifiable external photograph of a patient in a lecture and on its social media. What is required?
- A.A verbal agreement noted in the chart at that visit
- B.Nothing more, as the image is already on file
- C.Separate written authorization from the patient✓ Answer
Clinical photographs are protected health information, and consent to be photographed for care does not extend to publication, outside lectures or marketing; identifiable images used beyond the patient's own treatment need a separate signed authorization that states the intended use.
Ref QB-ZRHKGTSource: Confidentiality and authorization for use of patient images, ophthalmic assisting ethics and records practiceReport a problem with this question
22. Before a Schirmer test, the assistant instills topical anesthetic and gently dries the lower fornix before placing the strips. What does the strip reading then mainly reflect?
- A.Reflex secretion only, basal tearing blocked
- B.Basal secretion only, reflex tearing blocked✓ Answer
- C.Basal plus reflex secretion, both measured
Anesthetic removes the irritation that drives reflex tearing, so the Schirmer with anesthesia (basic secretion test) measures basal tear secretion. The Schirmer I test, done without anesthetic, measures basal plus reflex tearing, and Schirmer II adds nasal stimulation to measure reflex tearing.
Ref QB-G9CEWSSource: EyeWiki, Dry Eye Syndrome, 'Schirmer test' section (Schirmer I without anesthesia measures basic and reflex tearing; Schirmer with anesthesia measures basal secretion)Report a problem with this question
23. After fluorescein is instilled, the patient blinks a few times and then holds the eyes open while the assistant watches the cornea under cobalt blue light. What is recorded as the tear break-up time?
- A.Seconds from dye instillation to the first dry spot
- B.Seconds from the last blink to the first dry spot✓ Answer
- C.Seconds from the last blink to the next reflex blink
Tear break-up time measures tear film stability: the clock starts at the last complete blink and stops when the first dry spot or hole appears in the fluorescein-stained film. About 10 seconds is the usual cutoff, and shorter times are seen in dry eye, mucin deficiency and meibomian gland disease.
Ref QB-YR69BSSource: EyeWiki, Dry Eye Syndrome, 'Tear break up time (TBUT)' section (time from last blink to first dry spot under cobalt blue; 10 seconds typical cutoff)Report a problem with this question
24. At the slit lamp, a narrow beam angled about 60 degrees at the temporal limbus shows a peripheral anterior chamber gap less than one-quarter of the corneal thickness. How is this van Herick finding interpreted?
- A.Grade 3, an angle at which closure is improbable
- B.Grade 0, an angle that is already fully closed
- C.Grade 1, an angle at which closure is likely✓ Answer
Van Herick compares the peripheral chamber depth with corneal thickness: 1:1 or more is grade 4, 1:1/2 is grade 3, 1:1/4 is grade 2, less than 1:1/4 is grade 1 (closure likely), and no visible gap is grade 0. Higher grades mean wider angles, not narrower ones. A shallow result should be reported to the doctor before dilation, and the estimate does not replace gonioscopy.
Ref QB-K9EEEWSource: EyeWiki, Gonioscopy, 'Van Herick' section and Table 5 (Van Herick classification system)Report a problem with this question
25. A retina specialist needs to know whether a patient's choroidal neovascular membrane is actively leaking. The patient asks whether the dye-free OCT angiography scan can answer that instead. What is accurate?
- A.Yes; OCT-A shows leakage as spreading flow on the scan
- B.No; OCT-A maps flow but cannot show dye leakage over time✓ Answer
- C.No; OCT-A needs an iodine dye, so it carries the same risk
OCT-A builds vessel maps from the motion of red blood cells on repeated scans, so it needs no injected dye, but for the same reason it cannot show dye escaping from vessels. Leakage is seen on fluorescein angiography as hyperfluorescence that grows larger and brighter on late frames.
Ref QB-3S89UVSource: EyeWiki, Optical Coherence Tomography Angiography (OCT-A uses reflectance of moving red blood cells, eliminating intravascular dyes; 'OCT-A does not detect leakage')Report a problem with this question
26. A young patient's refraction keeps changing and the doctor suspects keratoconus. Which pattern on the axial corneal topography map most supports that suspicion?
- A.A symmetric bowtie with a straight vertical axis
- B.Inferior steepening with an asymmetric, skewed bowtie✓ Answer
- C.Central flattening ringed by a steeper midperiphery
Keratoconus thins and bulges the cornea, most often inferiorly, so topography shows a warm-colored zone of inferior steepening with an asymmetric bowtie and a skewed axis. A symmetric bowtie on a straight axis is ordinary regular astigmatism, and central flattening with a steeper ring is the pattern left by myopic refractive surgery.
Ref QB-J9349VSource: EyeWiki, Keratoconus, 'Diagnosis' and signs list (inferior steepening, skewed axis on computerized corneal topography)Report a problem with this question
27. After a corneal laceration repair, the doctor paints concentrated fluorescein over the wound and examines it under cobalt blue light. Which appearance indicates a positive Seidel test?
- A.A bright green patch of dye pooling over the wound
- B.Dark spots scattered in the tear film after a blink
- C.A dark stream of diluted dye flowing from the wound✓ Answer
Concentrated fluorescein looks green under cobalt blue light; aqueous escaping through a full-thickness defect dilutes it, so the leak appears as a dark stream flowing from the wound. A bright green patch is dye staining damaged epithelium, and dry spots after a blink are tear break-up, not a wound leak. The rate of flow should be noted.
Ref QB-2SVWNUSource: EyeWiki, Seidel Test (aqueous leakage dilutes fluorescein and appears as a dark blue stream under cobalt blue light = positive Seidel test)Report a problem with this question
28. The doctor wants fundus photographs that make a diabetic patient's small hemorrhages and fine retinal vessels stand out with the best contrast. Which filter should the photographer use?
- A.A green filter that blocks the red wavelengths✓ Answer
- B.A cobalt blue filter used for corneal staining
- C.A yellow barrier filter used for angiography
Red-free photography uses a green filter (about 540-570 nm) that removes red light, so blood absorbs the remaining light and vessels and hemorrhages appear dark against a brighter background. Cobalt blue is the excitation light for fluorescein staining, and the yellow barrier filter is used only when fluorescein dye has been injected for angiography.
Ref QB-D9AEDPSource: Wikipedia, Fundus photography, 'Modes' section (red-free photography uses a ~540-570 nm green filter to block red wavelengths for better contrast of vessels and hemorrhages)Report a problem with this question
Other topics in this exam
Practice questions based on the IJCAHPO core certification content areas for the Certified Ophthalmic Assistant and on standard ophthalmic clinical practice references. COA, COT, and COMT are marks of the International Joint Commission on Allied Health Personnel in Ophthalmology; this site is not affiliated with or endorsed by IJCAHPO. Ophthalmic assistants work under the delegation of a supervising ophthalmologist and permitted tasks vary by state and by practice — confirm your own scope, and check the current exam requirements and content outline with IJCAHPO before testing. About IJCAHPO certification →